An improved statistical representation for ECG electrode movement and muscular activity noises in the context of T-wave alternan estimation

被引:8
作者
Bakhshi, A. D. [1 ]
Bashir, S. [2 ]
Maud, M. A. [1 ]
机构
[1] Univ Engn & Technol, Dept Comp Sci & Engn, Lahore, Pakistan
[2] Iqra Univ, Islamabad, Pakistan
关键词
ECG noise; Biexponential; Electrocardiography; Goodness-of-fit; Laplacian; T-wave alternans; chi(2) test; SUSCEPTIBILITY; MECHANISM;
D O I
10.1016/j.bspc.2012.11.004
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Microvolt T-wave alternans (TWA) are recognized as markers for malignant ventricular arrhythmias, leading to sudden cardiac death. Its extraordinary pathological significance and life-critical application demand elaborate modeling approaches and efficient analysis schemes. Accurate statistical model encompassing the dynamics of physiological noises and other outliers is highly significant to detection and estimation of the microvolt signal. The anomalies in parametric values characterizing the distributions of the above random phenomena are apt to incur modeling errors. Recent TWA detection theoretic approaches assume Laplacian noise due to leptokurtic distribution of electrode movement (em) and muscular activity (ma) recordings. The presented statistical analysis shows that the practiced model compromises the asymmetric nature of the probability distributions for em and ma. An analytical model called Biexponential distribution is suggested to realize the leptokurtic as well as the asymmetric nature of the noise characteristics. Comparative analysis is presented using visual inspection method, chi(2) goodness-of-fit and Monte Carlo simulations. The proposed model achieves a best match of 99.14% and 98.13% for em and ma as compared to a Laplacian fit of 95.20% and 93.84%, respectively. Conversely, the worst fit values for em and ma are found to be 96.32% and 92.45% for Biexponential and 60.47% and 15.18% for Laplacian models, respectively. The augmented degree of freedom is likely to increase the complexity of the already challenging TWA detection problem; however, the proposed model achieves a more realistic representation of the real noise data by closely matching the statistical parameters. (c) 2012 Elsevier Ltd. All rights reserved.
引用
收藏
页码:297 / 301
页数:5
相关论文
共 18 条
[1]  
Adam D. R., 1982, Computers in Cardiology, P307
[2]   FLUCTUATIONS IN T-WAVE MORPHOLOGY AND SUSCEPTIBILITY TO VENTRICULAR-FIBRILLATION [J].
ADAM, DR ;
SMITH, JM ;
AKSELROD, S ;
NYBERG, S ;
POWELL, AO ;
COHEN, RJ .
JOURNAL OF ELECTROCARDIOLOGY, 1984, 17 (03) :209-218
[3]  
Camm AJ, 1996, EUR HEART J, V17, P354
[4]   PhysioBank, PhysioToolkit, and PhysioNet - Components of a new research resource for complex physiologic signals [J].
Goldberger, AL ;
Amaral, LAN ;
Glass, L ;
Hausdorff, JM ;
Ivanov, PC ;
Mark, RG ;
Mietus, JE ;
Moody, GB ;
Peng, CK ;
Stanley, HE .
CIRCULATION, 2000, 101 (23) :E215-E220
[5]  
Greenwood P., 1996, Guide to Chi-Squared Testing
[6]  
Hering H., 1909, Z Exper Med, V7, P363, DOI DOI 10.1007/BF02659994
[7]  
Lewis T, 1911, Q J MED, V4, P141
[8]   Characterization of repolarization alternans during ischemia:: Time-course and spatial analysis [J].
Martínez, JP ;
Olmos, S ;
Wagner, G ;
Laguna, P .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2006, 53 (04) :701-711
[9]   Methodological principles of T wave alternans analysis:: A unified framework [J].
Martínez, JP ;
Olmos, S .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2005, 52 (04) :599-613
[10]   A robust T wave alternans detector based on the GLRT for Laplacian noise distribution [J].
Martínez, JP ;
Olmos, S .
COMPUTERS IN CARDIOLOGY 2002, VOL 29, 2002, 29 :677-680