Optimization of a tissue-equivalent CVD-diamond dosimeter for radiotherapy using the Monte Carlo code PENELOPE

被引:12
作者
Gorka, Bartosz [1 ,2 ]
Fernandez-Varea, Jose M. [1 ,2 ,3 ]
Panettieri, Vanessa [4 ]
Nilsson, Bo [1 ,2 ]
机构
[1] Karolinska Inst, Dept Med Radiat Phys, SE-17176 Stockholm, Sweden
[2] Stockholm Univ, SE-17176 Stockholm, Sweden
[3] Univ Barcelona, Fac Fis ECM, ES-08028 Barcelona, Spain
[4] Karolinska Univ Hosp, Dept Hosp Phys, SE-17176 Stockholm, Sweden
关键词
CVD-diamond detector; dosimetry; Monte Carlo simulation; PENELOPE code;
D O I
10.1016/j.nima.2008.05.044
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The design of a chemical vapour deposited (CVD) diamond detector is studied by means of the Monte Carlo (MC) code PENELOPE. The energy, field-size and directional dependences have been studied, together with its potential for measurements of depth-dose curves and lateral beam profiles in a water phantom. The CVD-diamond sample modelled in this investigation had a thickness of 50 mu m. It was covered on both sides with 0.2-mu m-thick silver electrodes and encapsulated in polystyrene. The detector was exposed to a reference Co-60 photon beam as well as to clinically relevant 6 and 18 MV X-ray beams, with field sizes of 2 x 2, 5 x 5 and 10 x 10 cm(2) at the phantom surface. For the considered photon-beam qualities the calculated detector response varied by 3.4%. The change of the absorbed dose ratio diamond to water with field size was around 1.3% for the three qualities. The device showed a directional dependence with maximal variation of 1.7% and 2.0% for the Co-60 and 6MV beams, respectively, but for 18 MV X-rays the dependence was below 1%. MC simulations for narrow beams (2 x 2 cm(2)) proved that the prototype has an excellent spatial resolution, comparable to the thickness of the diamond layer, and is therefore suitable for lateral beam profile measurements in the presence of high dose gradients as often met in e.g. intensity-modulated radiotherapy. A better construction of the CVD-diamond dosimeter was proposed, where geometrical optimization of the encapsulation and the replacement of the silver electrodes with graphite was shown by further MC calculations to reduce the energy and angular dependences to about 1%. (C) 2008 Elsevier B.V. All rights reserved.
引用
收藏
页码:578 / 587
页数:10
相关论文
共 17 条
  • [1] Design and characterization of a tissue-equivalent CVD-diamond detector for clinical dosimetry in high-energy photon beams
    Gorka, B.
    Nilsson, B.
    Svensson, R.
    Brahme, A.
    Ascarelli, P.
    Trucchi, D. M.
    Conte, G.
    Kalish, R.
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2008, 24 (03): : 159 - 168
  • [2] An optimized Monte Carlo (PENELOPE) code for the characterization of gel-layer detectors in radiotherapy
    Castellano, G.
    Brusa, D.
    Carrara, M.
    Gambarini, G.
    Valente, M.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2007, 580 (01) : 502 - 505
  • [3] Modeling energy deposition in trabecular spongiosa using the Monte Carlo code Penelope
    Gersh, Jacob A.
    Dingfelder, Michael
    Toburen, Larry H.
    HEALTH PHYSICS, 2007, 93 (01): : 47 - 59
  • [4] DETERMINATION OF DOSE DISTRIBUTION ON LUNG SIMULATOR OBJECT WITH MONTE CARLO SIMULATION USING CODE PENELOPE
    Apaza, Danny G.
    Chahuara, Ruben D.
    Sellerico, Pedro P.
    Vega, Jose L.
    MOMENTO-REVISTA DE FISICA, 2019, (59): : 1 - 10
  • [5] Coincidence summing corrections for volume samples using the PENELOPE/penEasy Monte Carlo code
    Vargas, A.
    Camp, A.
    Serrano, I.
    Duch, M. A.
    APPLIED RADIATION AND ISOTOPES, 2014, 87 : 376 - 379
  • [6] Comparison of X-Ray Absorption in Mandibular Tissues and Tissue-Equivalent Polymeric Materials Using PHITS Monte Carlo Simulations
    Gokcekuyu, Yasemin
    Ekinci, Fatih
    Buyuksungur, Arda
    Guzel, Mehmet Serdar
    Acici, Koray
    Asuroglu, Tunc
    APPLIED SCIENCES-BASEL, 2024, 14 (23):
  • [7] Dosimetric Evaluation of Tissue Equivalent Gel Dosimeter Using Saccharide in Radiotherapy System
    Cho, Y.
    Lee, D.
    Jung, H.
    Ji, Y.
    Kim, K.
    Chang, U.
    Kwon, S.
    MEDICAL PHYSICS, 2014, 41 (06) : 252 - 252
  • [8] Monte Carlo simulation using the PENELOPE code with an ant colony algorithm to study MOSFET detectors
    Carvajal, M. A.
    Garcia-Pareja, S.
    Guirado, D.
    Vilches, M.
    Anguiano, M.
    Palma, A. J.
    Lallena, A. M.
    PHYSICS IN MEDICINE AND BIOLOGY, 2009, 54 (20) : 6263 - 6276
  • [9] Monte Carlo simulation of X-ray emission using the general-purpose code PENELOPE
    Llovet, X
    Fernández-Varea, JM
    Sempau, J
    Salvat, F
    SURFACE AND INTERFACE ANALYSIS, 2005, 37 (11) : 1054 - 1058
  • [10] Monte Carlo simulation of MOSFET dosimeter for electron backscatter using the GEANT4 code
    Chow, James C. L.
    Leung, Michael K. K.
    MEDICAL PHYSICS, 2008, 35 (06) : 2383 - 2390