Reliable measurement of the Seebeck coefficient of organic and inorganic materials between 260 K and 460 K

被引:27
作者
Beretta, D. [1 ,2 ]
Bruno, P. [1 ]
Lanzani, G. [1 ,2 ]
Caironi, M. [1 ]
机构
[1] Ist Italiano Tecnol, PoliMi, Ctr Nano Sci & Technol, I-20133 Milan, MI, Italy
[2] Politecn Milan, Dipartimento Fis, I-20133 Milan, MI, Italy
关键词
TRANSPORT; RESISTIVITY; THERMOPOWER;
D O I
10.1063/1.4926885
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
A new experimental setup for reliable measurement of the in-plane Seebeck coefficient of organic and inorganic thin films and bulk materials is reported. The system is based on the "Quasi-Static" approach and can measure the thermopower in the range of temperature between 260 K and 460 K. The system has been tested on a pure nickel bulk sample and on a thin film of commercially available PEDOT:PSS deposited by spin coating on glass. Repeatability within 1.5% for the nickel sample is demonstrated, while accuracy in the measurement of both organic and inorganic samples is guaranteed by time interpolation of data and by operating with a temperature difference over the sample of less than 1 K. (C) 2015 AIP Publishing LLC.
引用
收藏
页数:9
相关论文
共 16 条
[1]  
[Anonymous], Introduction to Thermoelectricity', DOI DOI 10.1007/978-3-642-00716-3
[2]  
BEYLIN VM, 1975, PHYS MET METALLOGR, V39, P63
[3]   Experimental setup for the measurement of the electrical resistivity and thermopower of thin films and bulk materials -: art. no. 053907 [J].
Boffoué, O ;
Jacquot, A ;
Dauscher, A ;
Lenoir, B ;
Stölzer, M .
REVIEW OF SCIENTIFIC INSTRUMENTS, 2005, 76 (05)
[4]   RETRACTED: Towards polymer-based organic thermoelectric generators (Retracted Article) [J].
Bubnova, Olga ;
Crispin, Xavier .
ENERGY & ENVIRONMENTAL SCIENCE, 2012, 5 (11) :9345-9362
[5]   Experimental set-up for thermopower and resistivity measurements at 100-1300 K [J].
Burkov, AT ;
Heinrich, A ;
Konstantinov, PP ;
Nakama, T ;
Yagasaki, K .
MEASUREMENT SCIENCE AND TECHNOLOGY, 2001, 12 (03) :264-272
[6]   Apparatus for measuring the Seebeck coefficients of highly resistive organic semiconducting materials [J].
Cai, H. Y. ;
Cui, D. F. ;
Li, Y. T. ;
Chen, X. ;
Zhang, L. L. ;
Sun, J. H. .
REVIEW OF SCIENTIFIC INSTRUMENTS, 2013, 84 (04)
[7]   Data analysis for Seebeck coefficient measurements [J].
de Boor, J. ;
Mueller, E. .
REVIEW OF SCIENTIFIC INSTRUMENTS, 2013, 84 (06)
[8]   An experimental apparatus for simultaneously measuring Seebeck coefficient and electrical resistivity from 100 K to 600 K [J].
Guan, Aiqiang ;
Wang, Hanfu ;
Jin, Hao ;
Chu, Weiguo ;
Guo, Yanjun ;
Lu, Guiwu .
REVIEW OF SCIENTIFIC INSTRUMENTS, 2013, 84 (04)
[9]   A high temperature apparatus for measurement of the Seebeck coefficient [J].
Iwanaga, Shiho ;
Toberer, Eric S. ;
LaLonde, Aaron ;
Snyder, G. Jeffrey .
REVIEW OF SCIENTIFIC INSTRUMENTS, 2011, 82 (06)
[10]  
Jacquot A, 2006, ICT'06: XXV INTERNATIONAL CONFERENCE ON THERMOELECTRICS, PROCEEDINGS, P184