Analytical considerations of beam hardening in medical accelerator photon spectra

被引:16
作者
Kleinschmidt, C [1 ]
机构
[1] Radiol Klin Merheim, Funkt Bereich Med Phys, Cologne, Germany
关键词
accelerator spectra; beam hardening; attenuation coefficient; spreadsheet calculations;
D O I
10.1118/1.598704
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Beam hardening is a well-known phenomenon for therapeutic accelerator beams passing through matter in narrow beam geometry. This study assesses quantitatively the magnitude of beam hardening of therapeutic beams in water. A formal concept of beam hardening is proposed which is based on the decrease of the mean attenuation coefficient with depth. On the basis of this concept calculations of beam hardening effects are easily performed by means of a commercial spreadsheet program. Published accelerator spectra and the tabulated values of attenuation coefficients serve as input for these calculations. It is shown that the mean attenuation coefficient starts at depth zero with an almost linear decrease and then slowly levels off to a limit value. A similar behavior is found for the beam hardening coefficient. A physically reasonable, semianalytical model is given which fits the data better than previously published functions. The energy dependence of the initial attenuation coefficient is evaluated and shown. It fits well to published experimental data. The initial beam hardening coefficient, however, shows no energy dependence. Its mean value (<(eta(0))over bar> approximate to 0.006 cm(-1)) is also in close agreement to the measured data. (C) 1999 American Association of Physicists in Medicine. [S0094-2405 (99)02209-9].
引用
收藏
页码:1995 / 1999
页数:5
相关论文
共 20 条
[1]   X-RAY-SPECTRA VS ATTENUATION DATA - A THEORETICAL-ANALYSIS [J].
BAIRD, LC .
MEDICAL PHYSICS, 1981, 8 (03) :319-323
[2]   RECONSTRUCTION OF MEGAVOLTAGE PHOTON SPECTRA BY ATTENUATION ANALYSIS [J].
BAKER, CR ;
AMAEE, B ;
SPYROU, NM .
PHYSICS IN MEDICINE AND BIOLOGY, 1995, 40 (04) :529-542
[3]   Relations between scatter factor, quality index and attenuation for x-ray beams [J].
Bjarngard, BE ;
Vadash, P .
PHYSICS IN MEDICINE AND BIOLOGY, 1998, 43 (05) :1325-1330
[4]  
Bjarngard BE, 1997, MED PHYS, V24, P1441, DOI 10.1118/1.598032
[5]  
BJARNGARD BE, 1994, MED PHYS, V21, P1069, DOI 10.1118/1.597349
[6]   Extraction of the photon spectra from measured beam parameters [J].
Bloch, P ;
McDonough, J .
MEDICAL PHYSICS, 1998, 25 (05) :752-757
[7]   FORWARD-DIRECTED BREMSSTRAHLUNG OF 10-MEV TO 30-MEV ELECTRONS INCIDENT ON THICK TARGETS OF AL AND PB [J].
FADDEGON, BA ;
ROSS, CK ;
ROGERS, DWO .
MEDICAL PHYSICS, 1990, 17 (05) :773-785
[8]   SIMULATION STUDIES OF 4-MV X-RAY SPECTRAL RECONSTRUCTION BY NUMERICAL-ANALYSIS OF TRANSMISSION DATA [J].
HUANG, PH ;
KASE, KR ;
BJARNGARD, BE .
MEDICAL PHYSICS, 1982, 9 (05) :695-702
[9]   RECONSTRUCTION OF 4-MV BREMSSTRAHLUNG SPECTRA FROM MEASURED TRANSMISSION DATA [J].
HUANG, PH ;
KASE, KR ;
BJARNGARD, BE .
MEDICAL PHYSICS, 1983, 10 (06) :778-785
[10]   SPECTRAL CHARACTERIZATION OF 4 MV BREMSSTRAHLUNG BY ATTENUATION ANALYSIS [J].
HUANG, PH ;
KASE, KR ;
BJARNGARD, BE .
MEDICAL PHYSICS, 1981, 8 (03) :368-374