Early Outcomes of Intuity Rapid Deployment Aortic Valve Replacement Compared With Conventional Biological Valves in Japanese Patients

被引:4
作者
Ono, Yoshikazu [1 ]
Yajima, Shin [1 ]
Kainuma, Satoshi [1 ]
Kawamoto, Naonori [1 ]
Tadokoro, Naoki [1 ]
Kakuta, Takashi [1 ]
Koga-Ikuta, Ayumi [1 ]
Fujita, Tomoyuki [1 ]
Fukushima, Satsuki [1 ]
机构
[1] Natl Cerebral & Cardiovasc Ctr, Dept Cardiac Surg, 6-1 Kishibe Shinmachi, Suita, Osaka 5648565, Japan
关键词
Aortic valve replacement; Intuity; Rapid deployment; PROSPECTIVE MULTICENTER; MORTALITY; BIOPROSTHESIS; PERFORMANCE; MANAGEMENT; MISMATCH; IMPACT;
D O I
10.1253/circj.CJ-21-0959
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: This study aimed to elucidate the short-term surgical outcomes and hemodynamics of the Intuity valve compared to the standard bioprosthesis in Japanese patients. Methods and Results: Among the 307 consecutive patients who underwent aortic valve replacement (AVR) between February 2019 and March 2021, the Intuity valve was implanted in 95 patients (Intuity group) and a conventional stented bioprosthesis was implanted in 193 patients (conventional group). After propensity score matching, there was no significant difference in in-hospital mortality between the Intuity (n=2, 3%) and conventional groups (n=0, P=0.490). Operation, cardiopulmonary bypass, and aortic cross-clamping times were significantly shorter in the Intuity group. Although the effective orifice area index, trans-prosthetic mean pressure gradient, and peak velocity were similar between the 2 groups at 1 week postoperatively, the Intuity group showed a better mean pressure gradient and peak velocity at 1 year postoperatively. Complete atrioventricular block requiring permanent pacemaker implantation developed in 2 patients (3%) in the Intuity group and none in the conventional group (P=0.476). Mild or greater paravalvular leakage was present in 8 patients (13%) in the Intuity group and 2 patients (3%) in the conventional group (P=0.095). Conclusions: AVR using the Intuity valve in Japanese patients is satisfactory, with a better valve performance and a low incidence of complete atrioventricular block at 1 year postoperatively.
引用
收藏
页码:1710 / +
页数:12
相关论文
共 29 条
  • [21] Otto CM, 2021, J AM COLL CARDIOL, V77, pE25
  • [22] Permanent Pacemaker Implantation After Rapid Deployment Aortic Valve Replacement
    Romano, Matthew A.
    Koeckert, Michael
    Mumtaz, Mubashir A.
    Slachman, Frank N.
    Patel, Himanshu J.
    Chitwood, W. Randolph, Jr.
    Barnhart, Glenn R.
    Grossi, Eugene A.
    [J]. ANNALS OF THORACIC SURGERY, 2018, 106 (03) : 685 - 690
  • [23] A mechanistic investigation of the EDWARDS INTUITY Elite valve's hemodynamic performance
    Sadri, Vahid
    Bloodworth, Charles H.
    Madukauwa-David, Immanuel David
    Midha, Prem A.
    Raghav, Vrishank
    Yoganathan, Ajit P.
    [J]. GENERAL THORACIC AND CARDIOVASCULAR SURGERY, 2020, 68 (01) : 9 - 17
  • [24] Evaluation and Management of Chronic Kidney Disease: Synopsis of the Kidney Disease: Improving Global Outcomes 2012 Clinical Practice Guideline
    Stevens, Paul E.
    Levin, Adeera
    [J]. ANNALS OF INTERNAL MEDICINE, 2013, 158 (11) : 825 - +
  • [25] Simple interrupted suturing increases valve performance after aortic valve replacement with a small supra-annular bioprosthesis
    Tabata, Minoru
    Shibayama, Kentaro
    Watanabe, Hiroyuki
    Sato, Yasunori
    Fukui, Toshihiro
    Takanashi, Shuichiro
    [J]. JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 2014, 147 (01) : 321 - 325
  • [26] Clinical outcome after surgical aortic valve replacement in low-risk Japanese patients with severe aortic stenosis
    Takeji, Yasuaki
    Taniguchi, Tomohiko
    Morimoto, Takeshi
    Saito, Naritatsu
    Ando, Kenji
    Shirai, Shinichi
    Kawase, Yuichi
    Kitai, Takeshi
    Shiomi, Hiroki
    Minamino-Muta, Eri
    Matsuda, Shintaro
    Yamazaki, Kazuhiro
    Miyake, Makoto
    Murata, Koichiro
    Kanamori, Norio
    Izumi, Chisato
    Mitsuoka, Hirokazu
    Kato, Masashi
    Hirano, Yutaka
    Inada, Tsukasa
    Nagao, Kazuya
    Mabuchi, Hiroshi
    Takeuchi, Yasuyo
    Yamane, Keiichiro
    Toyofuku, Mamoru
    Ishii, Mitsuru
    Inoko, Moriaki
    Ikeda, Tomoyuki
    Ishii, Katsuhisa
    Hotta, Kozo
    Jinnai, Toshikazu
    Higashitani, Nobuya
    Kato, Yoshihiro
    Inuzuka, Yasutaka
    Morikami, Yuko
    Minatoya, Kenji
    Kimura, Takeshi
    [J]. CARDIOVASCULAR INTERVENTION AND THERAPEUTICS, 2021, 36 (01) : 121 - 130
  • [27] Rapid Deployment of Aortic Bioprosthesis in Elderly Patients With Small Aortic Annulus
    Theron, Alexis
    Gariboldi, Vlad
    Grisoli, Dominique
    Jaussaud, Nicolas
    Morera, Pierre
    Lagier, David
    Leroux, Severine
    Amanatiou, Cecile
    Guidon, Catherine
    Riberi, Alberto
    Collart, Frederic
    [J]. ANNALS OF THORACIC SURGERY, 2016, 101 (04) : 1434 - 1442
  • [28] Contemporary Outcomes of Surgical Aortic Valve Replacement in Japan
    Tokuda, Yoshiyuki
    Yamamoto, Hiroyuki
    Miyata, Hiroaki
    Usui, Akihiko
    Motomura, Noboru
    [J]. CIRCULATION JOURNAL, 2020, 84 (02) : 277 - 282
  • [29] Sutureless Aortic Valve and Pacemaker Rate: From Surgical Tricks to Clinical Outcomes
    Vogt, Ferdinand
    Moscarelli, Marco
    Nicoletti, Anna
    Gregorini, Renato
    Pollari, Francesco
    Kalisnik, Jurji M.
    Pfeiffer, Steffen
    Fischlein, Theodor
    Santarpino, Giuseppe
    [J]. ANNALS OF THORACIC SURGERY, 2019, 108 (01) : 99 - 105