Study of radiation exposure dependence on the physical parameters of medical diagnostic X ray machines

被引:1
作者
Hamed, AA [1 ]
Elshirbiny, N
Nassef, MH
机构
[1] Atom Energy Author, Natl Ctr Nucl Safety & Radiat Control, Cairo, Egypt
[2] Cairo Univ, Kasr Elaini Ctr Radiat Oncol, Cairo, Egypt
关键词
D O I
10.1093/oxfordjournals.rpd.a032636
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
In this study the physical parameters of 14 X ray machines located at 14 different hospitals in Egypt were subjected to investigation. The studied parameters were tube voltage, tube current, exposure time, filtration, and exposure to the patient. The radiation exposure to the patient was measured for different classes of X ray examinations in the light of the nominal physical parameters. Selected values for the physical parameters for every class of examination have been predicted. The predicted values of the physical parameters in this study could be used for achieving a reduction in radiation exposure to patients undergoing medical X ray diagnosis in Egypt.
引用
收藏
页码:277 / 283
页数:7
相关论文
共 7 条
[1]  
AALBERS AHL, 1985, 18 BIR, P138
[2]  
*FINN CTR RAD NUCL, 1992, 33 ST FINN CTR RAD N
[3]  
*FINN CTR RAD NUCL, 1991, 35 ST FINN CTR RAD N
[4]  
HAMED AA, 1996, RAD DOSES SOME DIAGN
[5]   A SURVEY OF RADIATION-DOSES TO PATIENTS IN 5 COMMON DIAGNOSTIC EXAMINATIONS .1. PRINCIPLES AND TECHNIQUES OF DOSE EVALUATION BY INDIRECT MEASUREMENTS .2. SURVEY RESULTS [J].
HARRISON, RM ;
CLAYTON, CB ;
DAY, MJ ;
OWEN, JP ;
YORK, MF .
BRITISH JOURNAL OF RADIOLOGY, 1983, 56 (666) :383-395
[6]  
*IAEA, 1994, SAFETY SERIES IAEA, V115
[7]  
WALL BF, 1988, 55 IPSM