Evaluation of GAFCHROMIC EBT2 dosimetry for the low dose range using a flat-bed scanner with the reflection mode

被引:15
作者
Gotanda, Tatsuhiro [1 ,2 ]
Katsuda, Toshizo [3 ]
Akagawa, Takuya [4 ]
Gotanda, Rumi [5 ,6 ]
Tabuchi, Akihiko [7 ]
Yamamoto, Kenyu [8 ]
Kuwano, Tadao [6 ]
Yatake, Hidetoshi [9 ]
Yabunaka, Koichi [10 ]
Takeda, Yoshihiro [6 ]
机构
[1] Oishi Hosp, Dept Radiol, Fukuyama, Hiroshima 7200822, Japan
[2] Himeji Dokkyo Univ, Fac Hlth Care Sci, Himeji, Hyogo, Japan
[3] Seikeikai Coll, Fac Radiol Technol, Osaka, Japan
[4] Tokushima Red Cross Hosp, Dept Radiol Technol, Tokushima, Japan
[5] Ibaraki Prefectural Univ Hlth Sci, Sch Hlth Sci, Inashiki, Ibaraki, Japan
[6] Okayama Univ, Grad Sch Hlth Sci, Okayama 7008530, Japan
[7] Kawasaki Med Sch Hosp, Okayama, Japan
[8] Osaka Canc Prevent & Detect Ctr, Dept Radiol, Osaka, Japan
[9] Kaizuka City Hosp, Breast Canc Ctr, Osaka, Japan
[10] Katsuragi Hosp, Dept Radiol, Osaka, Japan
关键词
Radiochromic film; Flat-bed scanner; Transmission mode; Reflection mode; Density resolution; RADIOCHROMIC FILM; ENERGY; PHANTOM;
D O I
10.1007/s13246-013-0187-z
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Recently developed radiochromic films can easily be used to measure absorbed doses because they do not need development processing and indicate a density change that depends on the absorbed dose. However, in GAFCHROMIC EBT2 dosimetry (GAF-EBT2) as a radiochromic film, the precision of the measurement was compromised, because of non-uniformity problems caused by image acquisition using a flat-bed scanner with a transmission mode. The purpose of this study was to improve the precision of the measurement using a flat-bed scanner with a reflection mode at the low absorbed dose dynamic range of GAF-EBT2. The calibration curves of the absorbed dose versus the film density for GAF-EBT2 were provided. X-rays were exposed in the range between similar to 0 and 120 mGy in increments of about 12 mGy. The results of the method using a flat-bed scanner with the transmission mode were compared with those of the method using the same scanner with the reflection mode. The results should that the determination coefficients (r (2) ) for the straight-line approximation of the calibration curve using the reflection mode were higher than 0.99, and the gradient using the reflection mode was about twice that of the one using the transmission mode. The non-uniformity error that is produced by a flat-bed scanner with the transmission mode setting could be almost eliminated by converting from the transmission mode to the reflection mode. In light of these findings, the method using a flat-bed scanner with the reflection mode (only using uniform white paper) improved the precision of the measurement for the low absorbed dose range.
引用
收藏
页码:59 / 63
页数:5
相关论文
共 13 条
[1]   Scanner uniformity improvements for radiochromic film analysis with matt reflectance backing [J].
Butson, Ethan ;
Alnawaf, Hani ;
Yu, Peter K. N. ;
Butson, Martin .
AUSTRALASIAN PHYSICAL & ENGINEERING SCIENCES IN MEDICINE, 2011, 34 (03) :401-407
[2]   Energy response of the new EBT2 radiochromic film to x-ray radiation [J].
Butson, Martin J. ;
Yu, Peter K. N. ;
Cheung, Tsang ;
Alnawaf, Hani .
RADIATION MEASUREMENTS, 2010, 45 (07) :836-839
[3]   Radiochromic film for medical radiation dosimetry [J].
Butson, MJ ;
Yu, PKN ;
Cheung, T ;
Metcalfe, P .
MATERIALS SCIENCE & ENGINEERING R-REPORTS, 2003, 41 (3-5) :61-120
[4]   Absorption spectra time evolution of EBT-2 model GAFCHROMIC™ film [J].
Devic, Slobodan ;
Aldelaijan, Saad ;
Mohammed, Huriyyah ;
Tomic, Nada ;
Liang, Li-Heng ;
DeBlois, Francois ;
Seuntjens, Jan .
MEDICAL PHYSICS, 2010, 37 (05) :2207-2214
[5]   Computed tomography phantom for radiochromic film dosimetry [J].
Gotanda, R. ;
Katsuda, T. ;
Gotanda, T. ;
Eguchi, M. ;
Takewa, S. ;
Tabuchi, A. ;
Yatake, H. .
AUSTRALASIAN PHYSICAL & ENGINEERING SCIENCES IN MEDICINE, 2007, 30 (03) :194-199
[6]   Dose distribution in pediatric CT head examination using a new phantom with radiochromic film [J].
Gotanda, R. ;
Katsuda, T. ;
Gotanda, T. ;
Tabuchi, A. ;
Yatake, H. ;
Takeda, Y. .
AUSTRALASIAN PHYSICAL & ENGINEERING SCIENCES IN MEDICINE, 2008, 31 (04) :339-344
[7]   Evaluation of effective energy using radiochromic film and a step-shaped aluminum filter [J].
Gotanda, T. ;
Katsuda, T. ;
Gotanda, R. ;
Tabuchi, A. ;
Yamamoto, K. ;
Kuwano, T. ;
Yatake, H. ;
Kashiyama, K. ;
Yabunaka, K. ;
Akagawa, T. ;
Takeda, Y. .
AUSTRALASIAN PHYSICAL & ENGINEERING SCIENCES IN MEDICINE, 2011, 34 (02) :213-222
[8]   Half-value layer measurement: simple process method using radiochromic film [J].
Gotanda, T. ;
Katsuda, T. ;
Gotanda, R. ;
Tabuchi, A. ;
Yamamoto, K. ;
Kuwano, T. ;
Yatake, H. ;
Kashiyama, K. ;
Takeda, Y. .
AUSTRALASIAN PHYSICAL & ENGINEERING SCIENCES IN MEDICINE, 2009, 32 (03) :150-158
[9]   Evaluation of effective energy for QA and QC: measurement of half-value layer using radiochromic film density [J].
Gotanda, T. ;
Katsuda, T. ;
Gotanda, R. ;
Tabuchi, A. ;
Yamamoto, K. ;
Kuwano, T. ;
Yatake, H. ;
Takeda, Y. .
AUSTRALASIAN PHYSICAL & ENGINEERING SCIENCES IN MEDICINE, 2009, 32 (01) :26-29
[10]   EBT2 radiochromic film for quality assurance of complex IMRT treatments of the prostate: micro-collimated IMRT, RapidArc, and TomoTherapy [J].
Kairn, T. ;
Hardcastle, N. ;
Kenny, J. ;
Meldrum, R. ;
Tome, W. A. ;
Aland, T. .
AUSTRALASIAN PHYSICAL & ENGINEERING SCIENCES IN MEDICINE, 2011, 34 (03) :333-343