Whole-procedure clinical accuracy of Gamma Knife treatments of large lesions

被引:26
作者
Ma, Lijun [1 ]
Chuang, Cynthia [1 ]
Descovich, Martina [1 ]
Petti, Paula [1 ]
Smith, Vernon [1 ]
Verhey, Lynn [1 ]
机构
[1] Univ Calif San Francisco, Dept Radiat Oncol, San Francisco, CA 94143 USA
关键词
Gamma Knife; stereotactic radiosurgery; dosimetry; dose calculation;
D O I
10.1118/1.2987669
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The mechanical accuracy of Gamma Knife radiosurgery based on single-isocenter measurement has been established to within 0.3 mm. However, the full delivery accuracy for Gamma Knife treatments of large lesions has only been estimated via the quadrature-sum analysis. In this study, the authors directly measured the whole-procedure accuracy for Gamma Knife treatments of large lesions to examine the validity of such estimation. The measurements were conducted on a head-phantom simulating the whole treatment procedure that included frame placement, computed tomography imaging, treatment planning, and treatment delivery. The results of the measurements were compared with the dose calculations from the treatment planning system. Average agreements of 0.1-1.6 mm for the isodose lines ranging from 25% to 90% of the maximum dose were found despite potentially large contributing uncertainties such as 3-mm imaging resolution, 2-mm dose grid size, 1-mm frame registration, multi-isocenter deliveries, etc. The results of our measurements were found to be significantly smaller (>50%) than the calculated value based on the quadrature-sum analysis. In conclusion, Gamma Knife treatments of large lesions can be delivered much more accurately than predicted from the quadrature-sum analysis of major sources of uncertainties from each step of the delivery chain. (C) 2008 American Association of Physicists in Medicine. [DOI: 10.1118/1.2987669]
引用
收藏
页码:5110 / 5114
页数:5
相关论文
共 15 条
[1]   An analysis of the accuracy of the Cyberknife: A robotic nameless stereotactic radiosurgical system [J].
Chang, SD ;
Main, W ;
Martin, DP ;
Gibbs, IC ;
Heilbrun, MP .
NEUROSURGERY, 2003, 52 (01) :140-146
[2]   Comparative study of treatment dose plans after the refinement of Leksell Gamma Knife® single-beam dose profiles [J].
Cheung, Joel Y. C. ;
Ng, K. P. ;
Yu, C. P. ;
Ho, Robert T. K. .
MEDICAL PHYSICS, 2007, 34 (09) :3556-3561
[3]   Effects of residual target motion for image-tracked spine radiosurgery [J].
Chuang, Cynthia ;
Sahgal, Arjun ;
Lee, Letitia ;
Larson, David ;
Huang, Kim ;
Petti, Paula ;
Verhey, Lynn ;
Ma, Lijun .
MEDICAL PHYSICS, 2007, 34 (11) :4484-4490
[4]   A study of the accuracy of CyberKnife spinal radiosurgery using skeletal structure tracking [J].
Ho, Anthony K. ;
Fu, Dongshan ;
Cotrutz, Cristian ;
Hancock, Steven L. ;
Chang, Steven D. ;
Gibbs, Iris C. ;
Maurer, Calvin R., Jr. ;
Adler, John R., Jr. .
NEUROSURGERY, 2007, 60 (02) :147-156
[5]   Gamma knife radiosurgery for benign cavernous sinus tumors: Quantitative analysis of treatment outcomes [J].
Kuo, JS ;
Chen, JCT ;
Yu, C ;
Zelman, V ;
Giannotta, SL ;
Petrovich, Z ;
MacPherson, D ;
Apuzzo, MLJ .
NEUROSURGERY, 2004, 54 (06) :1385-1393
[6]   Phase II study of high central dose gamma knife radiosurgery and marimastat in patients with recurrent malignant glioma [J].
Larson, DA ;
Prados, M ;
Lamborn, KR ;
Smith, V ;
Sneed, PK ;
Chang, S ;
Nicholas, KM ;
Wara, WM ;
Devriendt, D ;
Kunwar, S ;
Berger, M ;
McDermott, MW .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2002, 54 (05) :1397-1404
[7]   STEREOTACTIC RADIOSURGERY - PRINCIPLES AND COMPARISON OF TREATMENT METHODS [J].
LUXTON, G ;
PETROVICH, Z ;
JOZSEF, G ;
NEDZI, LA ;
APUZZO, MLJ .
NEUROSURGERY, 1993, 32 (02) :241-259
[8]   Experimentally determined three-dimensional dose distributions in small complex targets [J].
Mack, A ;
Wolff, R ;
Weltz, D ;
Mack, G ;
Jess, A ;
Heck, B ;
Czempiel, H ;
Kreiner, HJ ;
Wowra, B ;
Böttcher, H ;
Seiffert, V .
JOURNAL OF NEUROSURGERY, 2002, 97 :551-555
[9]   Quality assurance in stereotactic space.: A system test for verifying the accuracy of aim in radiosurgery [J].
Mack, A ;
Czempiel, H ;
Kreiner, HJ ;
Dürr, G ;
Wowra, B .
MEDICAL PHYSICS, 2002, 29 (04) :561-568
[10]   The accuracy of dose localization for an image-guided frameless radiosurgery system [J].
Murphy, MJ ;
Cox, RS .
MEDICAL PHYSICS, 1996, 23 (12) :2043-2049