Free flap monitoring using simultaneous non-invasive laser Doppler flowmetry and tissue spectrophotometry

被引:170
作者
Hölzle, F [1 ]
Loeffelbein, DJ [1 ]
Nolte, D [1 ]
Wolff, KD [1 ]
机构
[1] Ruhr Univ Bochum, Dept Oral & Maxillofacial Plast Surg, Knappschaftskrankenhaus Bochum Langendreer, D-44892 Bochum, Germany
关键词
microvascular tissue transplantation; laser-Doppler flowmetry; tissue spectrophotometry;
D O I
10.1016/j.jcms.2005.07.010
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Background: Early identification of flap failure is an indispensable prerequisite for flap salvage. Although many methods of free flap monitoring are available, there is still no single reliable non-invasive technique for early recognition of flap failure and for differentiation between arterial occlusion and venous congestion. The aim of this study was to investigate the benefits of the tissue oxygen analysis system O2C for monitoring patients undergoing maxillofacial reconstruction with fasciocutaneous radial forearm flaps. Material and methods: In a prospective clinical study the microcirculatory parameters of blood flow, flow velocity, haemoglobin concentration (AU, Arbitrary Units) and oxygen saturation (%) were assessed by clinical means, by laser Doppler flowmetry and tissue spectrophotometry in 61 patients intraoperatively. Measurements were carried out before flap harvest, in the separated radial forearm flap, immediately after anastomoses and up to 14 days after reconstruction. Results: Following anastomosis, blood flow and flow velocity exceeded the level before flap elevation and reached significant differences by the third postoperative day (p < 0.05). Oxygen saturation decreased significantly by the third postoperative day and haemoglobin oxygenation showed stable values after performing anastomosis. Simultaneous, non-invasive laser-Doppler flowmetry and tissue spectrophotometry detected vascular complications in all cases with no false positive or false negative results and prior to clinical assessment. Conclusion: For the first time this new device allows reliable prediction of venous congestion by an increase of haemoglobin-concentration, and of arterial occlusion by a decrease in blood flow parameters and oxygen saturation. It can thus differentiate the mechanisms of flap failure before clinical assessment. (c) 2005 European Association for Cranio-Maxillofacial Surgery.
引用
收藏
页码:25 / 33
页数:9
相关论文
共 45 条
[1]  
ACLAND RD, 1981, PLAST RECONSTR SURG, V68, P554
[2]   SECONDARY ISCHEMIA TIME IN RODENTS - CONTRASTING COMPLETE PEDICLE INTERRUPTION WITH VENOUS OBSTRUCTION [J].
ANGEL, MF ;
MELLOW, CG ;
KNIGHT, KR ;
OBRIEN, BM .
PLASTIC AND RECONSTRUCTIVE SURGERY, 1990, 85 (05) :789-793
[3]   CORRELATION OF LASER DOPPLER WAVE PATTERNS WITH UNDERLYING MICROVASCULAR ANATOMY [J].
BRAVERMAN, IM ;
KEH, A ;
GOLDMINZ, D .
JOURNAL OF INVESTIGATIVE DERMATOLOGY, 1990, 95 (03) :283-286
[4]   Factors that influence the outcome of salvage in free tissue transfer [J].
Brown, JS ;
Devine, JC ;
Magennis, P ;
Sillifant, P ;
Rogers, SN ;
Vaughan, ED .
BRITISH JOURNAL OF ORAL & MAXILLOFACIAL SURGERY, 2003, 41 (01) :16-20
[5]  
CASPARY L, 1988, INT J MICROCIRC, V7, P367
[6]   ESTABLISHMENT OF NORMAL RANGES OF LASER DOPPLER BLOOD-FLOW IN AUTOLOGOUS TISSUE-TRANSPLANTS [J].
CLINTON, MS ;
SEPKA, RS ;
BRISTOL, D ;
PEDERSON, WC ;
BARWICK, WJ ;
SERAFIN, D ;
KLITZMAN, B .
PLASTIC AND RECONSTRUCTIVE SURGERY, 1991, 87 (02) :299-309
[7]  
Disa JJ, 1999, PLAST RECONSTR SURG, V104, P97, DOI 10.1097/00006534-199907000-00014
[8]   IMPEDANCE MONITORING FOR SUBCUTANEOUS FREE FLAP TRANSFERS [J].
HARRISON, DH ;
MOTT, G .
BRITISH JOURNAL OF PLASTIC SURGERY, 1989, 42 (03) :318-323
[9]   Laser Doppler flowmeter monitoring of free-tissue transfers: Blood flow in normal and complicated cases [J].
Heller, L ;
Levin, LS ;
Klitzman, B .
PLASTIC AND RECONSTRUCTIVE SURGERY, 2001, 107 (07) :1739-1745
[10]   REGIONAL VARIATIONS OF LASER DOPPLER BLOOD-FLOW IN ISCHEMIC SKIN FLAPS [J].
HICKERSON, WL ;
COLGIN, SL ;
PROCTOR, KG .
PLASTIC AND RECONSTRUCTIVE SURGERY, 1990, 86 (02) :319-326