Reducing temperature elevation of robotic bone drilling

被引:47
作者
Feldmann, Arne [1 ]
Wandel, Jasmin [2 ]
Zysset, Philippe [1 ]
机构
[1] Univ Bern, Inst Surg Technol & Biomech, CH-3012 Bern, Switzerland
[2] Bern Univ Appl Sci, Inst Risks & Extremes, Burgdorf, Switzerland
基金
瑞士国家科学基金会;
关键词
Bone drilling; Temperature elevation of drilling; Cooling; Interval drilling; Drill bit design; Cumulative equivalent minutes; Tissue damage; Rake angle; Chisel edge; BOVINE CORTICAL BONE; IRRIGATION TEMPERATURE; THERMAL OSTEONECROSIS; COCHLEAR IMPLANTATION; MODELS; DAMAGE; FORCE; HEAT;
D O I
10.1016/j.medengphy.2016.10.001
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
This research work aims at reducing temperature elevation of bone drilling. An extensive experimental study was conducted which focused on the investigation of three main measures to reduce the temperature elevation as used in industry: irrigation, interval drilling and drill bit designs. Different external irrigation rates (0 ml/min, 15 ml/min, 30 ml/min), continuously drilled interval lengths (2 mm, 1 mm, 0.5 mm) as well as two drill bit designs were tested. A custom single flute drill bit was designed with a higher rake angle and smaller chisel edge to generate less heat compared to a standard surgical drill bit. A new experimental setup was developed to measure drilling forces and torques as well as the 2D temperature field at any depth using a high resolution thermal camera. The results show that external irrigation is a main factor to reduce temperature elevation due not primarily to its effect on cooling but rather due to the prevention of drill bit clogging. During drilling, the build up of bone material in the drill bit flutes result in excessive temperatures due to an increase in thrust forces and torques. Drilling in intervals allows the removal of bone chips and cleaning of flutes when the drill bit is extracted as well as cooling of the bone in-between intervals which limits the accumulation of heat. However, reducing the length of the drilled interval was found only to be beneficial for temperature reduction using the newly designed drill bit due to the improved cutting geometry. To evaluate possible tissue damage caused by the generated heat increase, cumulative equivalent minutes (CEM43) were calculated and it was found that the combination of small interval length (0.5 mm), high irrigation rate (30 ml/min) and the newly designed drill bit was the only parameter combination which allowed drilling below the time thermal threshold for tissue damage. In conclusion, an optimized drilling method has been found which might also enable drilling in more delicate procedures such as that performed during minimally invasive robotic cochlear implantation. (C) 2016 IPEM. Published by Elsevier Ltd. All rights reserved.
引用
收藏
页码:1495 / 1504
页数:10
相关论文
共 41 条
[1]   Investigation of the effect of movement and irrigation systems on temperature in the conventional drilling of cortical bone [J].
Alexandre Gehrke, Sergio ;
Neto, Henrique Loffredo ;
Mardegan, Fabio E. C. .
BRITISH JOURNAL OF ORAL & MAXILLOFACIAL SURGERY, 2013, 51 (08) :953-957
[2]  
[Anonymous], 2004, OPTIMIZATION DRILL D
[3]  
[Anonymous], 2006, THESIS
[4]  
Astakhov V.P., 2014, DRILLS SCI TECHNOLOG
[5]   Thermal osteonecrosis and bone drilling parameters revisited [J].
Augustin, Goran ;
Davila, Slavko ;
Mihoci, Kristijan ;
Udiljak, Toma ;
Vedrina, Denis Stjepan ;
Antabak, Anko .
ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 2008, 128 (01) :71-77
[6]   Temperature changes during cortical bone drilling with a newly designed step drill and an internally cooled drill [J].
Augustin, Goran ;
Davila, Slavko ;
Udilljak, Toma ;
Staroveski, Tomislav ;
Brezak, Danko ;
Babic, Slaven .
INTERNATIONAL ORTHOPAEDICS, 2012, 36 (07) :1449-1456
[7]   Cortical bone drilling and thermal osteonecrosis [J].
Augustin, Goran ;
Zigman, Tomislav ;
Davila, Slavko ;
Udilljak, Toma ;
Staroveski, Tomislav ;
Brezak, Danko ;
Babic, Slaven .
CLINICAL BIOMECHANICS, 2012, 27 (04) :313-325
[8]   Determination of spatial distribution of increase in bone temperature during drilling by infrared thermography: preliminary report [J].
Augustin, Goran ;
Davila, Slavko ;
Udiljak, Toma ;
Vedrina, Denis Stjepan ;
Bagatin, Dinko .
ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 2009, 129 (05) :703-709
[9]   A self-developed and constructed robot for minimally invasive cochlear implantation [J].
Bell, Brett ;
Stieger, Christof ;
Gerber, Nicolas ;
Arnold, Andreas ;
Nauer, Claude ;
Hamacher, Volkmar ;
Kompis, Martin ;
Nolte, Lutz ;
Caversaccio, Marco ;
Weber, Stefan .
ACTA OTO-LARYNGOLOGICA, 2012, 132 (04) :355-360
[10]   3-fluted orthopaedic drills exhibit superior bending stiffness to their 2-fluted rivals: Clinical implications for targeting ability and the incidence of drill-bit failure [J].
Bertollo, N. ;
Gothelf, T. K. ;
Walsh, W. R. .
INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2008, 39 (07) :734-741