Application of a micro-droplet generator for an ICP-sector field mass spectrometer - optimization and analytical characterization

被引:68
作者
Shigeta, Kaori [1 ,2 ]
Traub, Heike [1 ]
Panne, Ulrich [1 ]
Okino, Akitoshi [2 ]
Rottmann, Lothar [3 ]
Jakubowski, Norbert [1 ]
机构
[1] BAM Fed Inst Mat Res & Testing, D-12489 Berlin, Germany
[2] Tokyo Inst Technol, Dept Energy Sci, Midori Ku, Yokohama, Kanagawa 2268502, Japan
[3] Thermo Fisher Sci, D-28199 Bremen, Germany
关键词
INDUCTIVELY-COUPLED PLASMA; DRIED MICROPARTICULATE INJECTOR; SINGLE CELLS; MS; NANOPARTICLES; PARTICLES; TOOL;
D O I
10.1039/c2ja30207a
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A micro-droplet generator (mDG) sample introduction system was coupled to a sector field ICP-MS instrument to investigate the analytical figures of merit with respect to single cell analysis. The sector field instrument was operated for the first time in a fast scanning mode (E-scan) with the shortest time resolution of 100 mu s to measure the single droplet time resolved and using the original detector in a pulse counting mode without modification of the existing electronics. For reduction of the droplet diameter a triple pulse mode of the droplet generator was applied and a droplet diameter down to 23 mm has been achieved for this investigation with a 100% transport efficiency of droplets. Signal duration times of single droplets of less than 500 ms have been measured. Overall detection efficiencies in the range of 10(-3) counts per atom have been achieved and absolute limits of detection range between 120 ag for Fe and 1.1 ag for Mg as a mean value from 1000 droplet events.
引用
收藏
页码:646 / 656
页数:11
相关论文
共 22 条
[1]  
BORCHERT UK, 1989, J AEROSOL SCI, V20, P1525
[2]   SPECTROTHERMAL EMISSION AEROSOL PARTICLE ANALYZER (SEAPA) [J].
CRIDER, WL .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1968, 39 (02) :212-&
[3]   Quantitative Imaging of Gold and Silver Nanoparticles in Single Eukaryotic Cells by Laser Ablation ICP-MS [J].
Drescher, Daniela ;
Giesen, Charlotte ;
Traub, Heike ;
Panne, Ulrich ;
Kneipp, Janina ;
Jakubowski, Norbert .
ANALYTICAL CHEMISTRY, 2012, 84 (22) :9684-9688
[4]   Single particle inductively coupled plasma mass spectrometry: evaluation of three different pneumatic and piezo-based sample introduction systems for the characterization of silver nanoparticles [J].
Franze, Bastian ;
Strenge, Ingo ;
Engelhard, Carsten .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2012, 27 (07) :1074-1083
[5]   MONODISPERSE DRIED MICROPARTICULATE INJECTOR FOR ANALYTICAL INSTRUMENTATION [J].
FRENCH, JB ;
ETKIN, B ;
JONG, R .
ANALYTICAL CHEMISTRY, 1994, 66 (05) :685-691
[6]   Influence of design and operating parameters of pneumatic concentric nebulizer on micro-flow aerosol characteristics and ICP-MS analytical performances [J].
Geertsen, Valerie ;
Lemaitre, Pascal ;
Tabarant, Michel ;
Chartier, Frederic .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2012, 27 (01) :146-158
[7]   Iodine as an elemental marker for imaging of single cells and tissue sections by laser ablation inductively coupled plasma mass spectrometry [J].
Giesen, Charlotte ;
Waentig, Larissa ;
Mairinger, Thomas ;
Drescher, Daniela ;
Kneipp, Janina ;
Roos, Peter H. ;
Panne, Ulrich ;
Jakubowski, Norbert .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2011, 26 (11) :2160-2165
[8]   Multiplexed Immunohistochemical Detection of Tumor Markers in Breast Cancer Tissue Using Laser Ablation Inductively Coupled Plasma Mass Spectrometry [J].
Giesen, Charlotte ;
Mairinger, Thomas ;
Khoury, Lina ;
Waentig, Larissa ;
Jakubowski, Norbert ;
Panne, Ulrich .
ANALYTICAL CHEMISTRY, 2011, 83 (21) :8177-8183
[9]   Local effects of atomizing analyte droplets on the plasma parameters of the inductively coupled plasma [J].
Groh, Sebastian ;
Garcia, Carmen C. ;
Murtazin, Ayrat ;
Horvatic, Vlasta ;
Niemax, Kay .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 2009, 64 (03) :247-254
[10]   Capabilities of inductively coupled plasma mass spectrometry for the detection of nanoparticles carried by monodisperse microdroplets [J].
Gschwind, Sabrina ;
Flamigni, Luca ;
Koch, Joachim ;
Borovinskaya, Olga ;
Groh, Sebastian ;
Niemax, Kay ;
Guenther, Detlef .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2011, 26 (06) :1166-1174