Accurate skin dose measurements using radiochromic film in clinical applications

被引:156
作者
Devic, S
Seuntjens, J
Abdel-Rahman, W
Evans, M
Olivares, M
Podgorsak, EB
Vuong, T
Soares, CG
机构
[1] McGill Univ, Ctr Hlth, Dept Phys Med, Montreal, PQ, Canada
[2] McGill Univ, Ctr Hlth, Dept Radiat Oncol, Montreal, PQ, Canada
[3] NIST, Div Ionizing Radiat, Gaithersburg, MD 20899 USA
关键词
clinical dosimetry; skin dose; surface dose; radiochromic film;
D O I
10.1118/1.2179169
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Megavoltage x-ray beams exhibit the well-known phenomena of dose buildup within the first few millimeters of the incident phantom surface, or the skin. Results of the surface dose measurements, however, depend vastly on the measurement technique employed. Our goal in this study was to determine a correction procedure in order to obtain an accurate skin dose estimate at the clinically relevant depth based on radiochromic film measurements. To illustrate this correction, we have used as a reference point a depth of 70 mu. We used the new GAFCHROMIC (R) dosimetry films (HS, XR-T, and EBT) that have effective points of measurement at depths slightly larger than 70 mu. In addition to films, we also used an Attix parallel-plate chamber and a home-built extrapolation chamber to cover tissue-equivalent depths in the range from 4 mu, to 1 mm of water-equivalent depth. Our measurements suggest that within the first millimeter of the skin region, the PDD for a 6 MV photon beam and field size of 10 X 10 cm 2 increases from 14% to 43%. For the three GAFCHROMIC (R) dosimetry film models, the 6 MV beam entrance skin dose measurement corrections due to their effective point of measurement are as follows: 15% for the EBT, 15% for the HS, and 16% for the XR-T model GAFCHROMICO films. The correction factors for the exit skin dose due to the build-down region are negligible. There is a small field size dependence for the entrance skin dose correction factor when using the EBT GAFCHROMICO film model. Finally, a procedure that uses EBT model GAFCHROMICO film for an accurate measurement of the skin dose in a parallel-opposed pair 6 MV photon beam arrangement is described. (c) 2006 American Association of Physicists in Medicine.
引用
收藏
页码:1116 / 1124
页数:9
相关论文
共 39 条
[1]   Validation of Monte Carlo calculated surface doses for megavoltage photon beams [J].
Abdel-Rahman, W ;
Seuntjens, JP ;
Verhaegen, F ;
Deblois, F ;
Podgorsak, EB .
MEDICAL PHYSICS, 2005, 32 (01) :286-298
[2]  
[Anonymous], ICRP PUBLICATION
[3]  
BJARNGARD BE, 1995, MED PHYS, V22, P465, DOI 10.1118/1.597477
[4]   Measurement of 6-MV x-ray surface dose when topical agents are applied prior to external beam irradiation [J].
Burch, SE ;
Parker, SA ;
Vann, AM ;
Arazie, JC .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1997, 38 (02) :447-451
[5]   Extrapolated surface dose measurements with radiochromic film [J].
Butson, MJ ;
Yu, PKN ;
Metcalfe, PE .
MEDICAL PHYSICS, 1999, 26 (03) :485-488
[6]   Radiochromic film as a radiotherapy surface-dose detector [J].
Butson, MJ ;
Mathur, JN ;
Metcalfe, PE .
PHYSICS IN MEDICINE AND BIOLOGY, 1996, 41 (06) :1073-1078
[7]   A new radiotherapy surface dose detector: The MOSFET [J].
Butson, MJ ;
Rozenfeld, A ;
Mathur, JN ;
Carolan, M ;
Wong, TPY ;
Metcalfe, PE .
MEDICAL PHYSICS, 1996, 23 (05) :655-658
[8]   Measurement of off-axis and peripheral skin dose using radiochromic film [J].
Butson, MJ ;
Yu, PKN ;
Metcalfe, PE .
PHYSICS IN MEDICINE AND BIOLOGY, 1998, 43 (09) :2647-2650
[9]   Radiochromic film for medical radiation dosimetry [J].
Butson, MJ ;
Yu, PKN ;
Cheung, T ;
Metcalfe, P .
MATERIALS SCIENCE & ENGINEERING R-REPORTS, 2003, 41 (3-5) :61-120
[10]   IMPLEMENTATION OF IRCP RECOMMENDATION ON SKIN DOSE MEASUREMENT USING THERMOLUMINESCENT DOSIMETERS [J].
CHARLES, MW ;
KHAN, ZU .
PHYSICS IN MEDICINE AND BIOLOGY, 1978, 23 (05) :972-975