Metal ion levels in large-diameter total hip and resurfacing hip arthroplasty-Preliminary results of a prospective five year study after two years of follow-up

被引:21
作者
Maurer-Ertl, W. [1 ]
Friesenbichler, J. [1 ]
Sadoghi, P. [1 ]
Pechmann, M. [1 ]
Trennheuser, M. [1 ]
Leithner, A. [1 ]
机构
[1] Med Univ Graz, Dept Orthopaed Surg, Graz, Austria
来源
BMC MUSCULOSKELETAL DISORDERS | 2012年 / 13卷
关键词
Hip resurfacing; Total hip arthroplasty; Prosthetic wear; Cobalt; Chromium; JOHN CHARNLEY; REPLACEMENT; COMPONENT; COBALT; BLOOD; ORIENTATION; BEARINGS; CHROMIUM; RELEASE; TRIAL;
D O I
10.1186/1471-2474-13-56
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background: Metal-on-metal hip resurfacing is an alternative to metal-on-metal total hip arthroplasty, especially for young and physically active patients. However, wear which might be detected by increased serum ion levels is a matter of concern. Methods: The aims of this preliminary study were to determine the raise of metal ion levels at 2-years follow-up in a prospective setting and to evaluate differences between patients with either resurfacing or total hip arthroplasty. Furthermore we investigated if the inclination of the acetabular component and the arc of cover would influence these findings. Therefore, 36 patients were followed prospectively. Results: The results showed increments for Co and Cr in both implant groups. Patients treated with large-diameter total hip arthroplasty showed fourfold and threefold, respectively, higher levels for Co and Cr compared to the resurfacing group (Co: p < 0,001 and Cr: p = 0,005). Nevertheless, we observed no significant correlation between serum ion levels, inclination and arc of cover. Discussion: In order to clarify the biologic effects of ion dissemination and to identify risks concerning long-term toxicity of metals, the exposure should be monitored carefully. Therefore, long-term studies have to be done to determine adverse effects of Co and Cr following metal-on-metal hip replacement.
引用
收藏
页数:7
相关论文
共 28 条
  • [1] Metal ion levels in the blood of patients after hip resurfacing: A comparison between twenty-eight and thirty-six-millimeter-head metal-on-metal prostheses
    Antoniou, John
    Zukor, David J.
    Mwale, Fackson
    Minarik, William
    Petit, Alain
    Huk, Olga L.
    [J]. JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2008, 90A : 142 - 148
  • [2] How do serum cobalt and chromium levels change after metal-on-metal hip resurfacing?
    Back, DL
    Young, AA
    Shimmin, AJ
    [J]. CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2005, (438) : 177 - 181
  • [3] Blood and urine metal ion levels in young and active patients after Birmingham hip resurfacing arthroplasty - Four-year results of a prospective longitudinal study
    Daniel, J.
    Ziaee, H.
    Pradhan, C.
    Pynsent, P. B.
    McMinn, D. J. W.
    [J]. JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2007, 89B (02): : 169 - 173
  • [4] Six-year results of a prospective study of metal ion levels in young patients with metal-on-metal hip resurfacings
    Daniel, J.
    Ziaee, H.
    Pradhan, C.
    McMinn, D. J. W.
    [J]. JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2009, 91B (02): : 176 - 179
  • [5] Correlation between inclination of the acetabular component and metal ion levels in metal-on-metal hip resurfacing replacement
    De Haan, R.
    Pattyn, C.
    Gill, H. S.
    Murray, D. W.
    Campbell, P. A.
    De Smet, K.
    [J]. JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2008, 90B (10): : 1291 - 1297
  • [6] Revision of metal-on-metal resurfacing arthroplasty of the hip - The influence of malpositioning of the components
    De Haan, R.
    Campbell, P. A.
    Su, E. P.
    De Smet, K. A.
    [J]. JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2008, 90B (09): : 1158 - 1163
  • [7] Metal Ion Measurement as a Diagnostic Tool to Identify Problems with Metal-on-Metal Hip Resurfacing
    de Smet, K.
    de Haan, R.
    Calistri, A.
    Campbell, P. A.
    Ebramzadeh, E.
    Pattyn, C.
    Gill, H. S.
    [J]. JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2008, 90A : 202 - 208
  • [8] Metal ion levels following resurfacing arthroplasty of the hip SERIAL RESULTS OVER A TEN-YEAR PERIOD
    deSouza, R. M.
    Parsons, N. R.
    Oni, T.
    Dalton, P.
    Costa, M.
    Krikler, S.
    [J]. JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2010, 92B (12): : 1642 - 1647
  • [9] 2008 John Charnley Award: Metal Ion Levels After Metal-on-Metal Total Hip Arthroplasty: A Randomized Trial
    Engh, C. Anderson, Jr.
    MacDonald, Steven J.
    Sritulanondha, Supatra
    Thompson, Abigail
    Naudie, Douglas
    Engh, Charles A.
    [J]. CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2009, 467 (01) : 101 - 111
  • [10] The John Charnley Award Metal-on-Metal Hip Resurfacing versus Large-diameter Head Metal-on-Metal Total Hip Arthroplasty A Randomized Clinical Trial
    Garbuz, Donald S.
    Tanzer, Michael
    Greidanus, Nelson V.
    Masri, Bassam A.
    Duncan, Clive P.
    [J]. CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2010, 468 (02) : 318 - 325