Auto-masked 2D/3D image registration and its validation with clinical cone-beam computed tomography

被引:15
作者
Steininger, P. [1 ]
Neuner, M. [1 ]
Weichenberger, H. [1 ,2 ]
Sharp, G. C. [3 ]
Winey, B. [3 ]
Kametriser, G. [2 ]
Sedlmayer, F. [1 ,2 ]
Deutschmann, H. [1 ,2 ]
机构
[1] Paracelsus Med Univ, Inst Res & Dev Adv Radiat Technol radART, A-5020 Salzburg, Austria
[2] Paracelsus Med Univ, Univ Clin Radiotherapy & Radiooncol, A-5020 Salzburg, Austria
[3] Massachusetts Gen Hosp, Dept Radiat Oncol, Boston, MA 02114 USA
关键词
GUIDED RADIATION-THERAPY; 2D-3D REGISTRATION; DIGITAL TOMOSYNTHESIS; CT; RADIOTHERAPY; RECONSTRUCTION; IMPLEMENTATION; GUIDANCE; SYSTEMS;
D O I
10.1088/0031-9155/57/13/4277
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Image-guided alignment procedures in radiotherapy aim at minimizing discrepancies between the planned and the real patient setup. For that purpose, we developed a 2D/3D approach which rigidly registers a computed tomography (CT) with two x-rays by maximizing the agreement in pixel intensity between the x-rays and the corresponding reconstructed radiographs from the CT. Moreover, the algorithm selects regions of interest (masks) in the x-rays based on 3D segmentations from the pre-planning stage. For validation, orthogonal x-ray pairs from different viewing directions of 80 pelvic cone-beam CT (CBCT) raw data sets were used. The 2D/3D results were compared to corresponding standard 3D/3D CBCT-to-CT alignments. Outcome over 8400 2D/3D experiments showed that parametric errors in root mean square were <0.18 degrees (rotations) and <0.73 mm (translations), respectively, using rank correlation as intensity metric. This corresponds to a mean target registration error, related to the voxels of the lesser pelvis, of <2mm in 94.1% of the cases. From the results we conclude that 2D/3D registration based on sequentially acquired orthogonal x-rays of the pelvis is a viable alternative to CBCT-based approaches if rigid alignment on bony anatomy is sufficient, no volumetric intra-interventional data set is required and the expected error range fits the individual treatment prescription.
引用
收藏
页码:4277 / 4292
页数:16
相关论文
共 50 条
[21]   An Innovative Approach to Reduce Streaking Artifacts in FDK Based 3D Cone-beam Tomography [J].
Bhowmik, Ujjal Kumar ;
Sudeepth, Araveti Venkata ;
Adhami, Reza R. .
IEEE SOUTHEASTCON 2011: BUILDING GLOBAL ENGINEERS, 2011, :359-364
[22]   Evaluation of similarity measures for 3D/2D image registration [J].
Skerl, Darko ;
Likar, Bostjan ;
Pernus, Franjo .
MEDICAL IMAGING 2006: IMAGE PROCESSING, PTS 1-3, 2006, 6144
[23]   2D radiographs, cone-beam computed tomography and 3D CBCT-based planning software in access cavity preparation: A single blinded randomised controlled in vitro study [J].
Christofzik, David W. ;
Glandorf, Patrick ;
Conrad, Jonas ;
El-Sayed, Karim M. Fawzy ;
Groessner-Schreiber, Birte ;
Doerfer, Christof E. .
AUSTRALIAN ENDODONTIC JOURNAL, 2022, 48 (02) :283-296
[24]   A novel evaluation model of image registration for cone-beam computed tomography guided lung cancer radiotherapy [J].
Liu, Yimei ;
Chen, Meining ;
Fang, Jianlan ;
Xiao, Liangjie ;
Liu, Songran ;
Li, Qiwen ;
Qiu, Bo ;
Huang, Runda ;
Zhang, Jun ;
Peng, Yinglin .
THORACIC CANCER, 2024, 15 (17) :1333-1342
[25]   Twin robotic x-ray system for 2D radiographic and 3D cone-beam CT imaging [J].
Fieselmann, Andreas ;
Steinbrener, Jan ;
Jerebko, Anna K. ;
Voigt, Johannes M. ;
Scholz, Rosemarie ;
Ritschl, Ludwig ;
Mertelmeier, Thomas .
MEDICAL IMAGING 2016: PHYSICS OF MEDICAL IMAGING, 2016, 9783
[26]   Evaluation of Computed Tomography to Ultrasound 2D Image Registration for Atrial Fibrillation Treatment [J].
Sandoval, Zulma L. ;
Dillenseger, Jean-Louis .
2013 COMPUTING IN CARDIOLOGY CONFERENCE (CINC), 2013, 40 :245-248
[27]   5D respiratory motion model based image reconstruction algorithm for 4D cone-beam computed tomography [J].
Liu, Jiulong ;
Zhang, Xue ;
Zhang, Xiaoqun ;
Zhao, Hongkai ;
Gao, Yu ;
Thomas, David ;
Low, Daniel A. ;
Gao, Hao .
INVERSE PROBLEMS, 2015, 31 (11)
[28]   Motion-aware temporal regularization for improved 4D cone-beam computed tomography [J].
Mory, Cyril ;
Janssens, Guillaume ;
Rit, Simon .
PHYSICS IN MEDICINE AND BIOLOGY, 2016, 61 (18) :6856-6877
[29]   Influence of imaging source and panel position uncertainties on the accuracy of 2D/3D image registration of cranial images [J].
Warmerdam, Guy ;
Steininger, Philipp ;
Neuner, Markus ;
Sharp, Gregory ;
Winey, Brian .
MEDICAL PHYSICS, 2012, 39 (09) :5547-5556
[30]   Intraoperative 3D imaging with cone-beam computed tomography leads to revision of pedicle screws in dorsal instrumentation: a retrospective analysis [J].
Zimmermann, Felix ;
Kohl, Katharina ;
Privalov, Maxim ;
Franke, Jochen ;
Vetter, Sven Y. .
JOURNAL OF ORTHOPAEDIC SURGERY AND RESEARCH, 2021, 16 (01)