Assay in serum of exchangeable copper and total copper using inductively coupled plasma mass spectrometry (ICP-MS): development, optimisation and evaluation of a routine procedure

被引:9
作者
Daymond, Richard
Curtis, Sarah L.
Mishra, Vinita
Roberts, Norman B.
机构
[1] Royal Liverpool Univ Hosp, Dept Clin Biochem, Liverpool Clin Labs, Liverpool, Merseyside, England
[2] Broadgreen Univ Hosp, Dept Clin Biochem, Liverpool Clin Labs, Liverpool, Merseyside, England
关键词
Free copper; non-caeruloplasmin copper; exchangeable copper; total copper; relative exchangeable copper; ultrafiltration; ultrafiltrable copper; caeruloplasmin; ICP-MS; WILSONS-DISEASE; CHROMATOGRAPHY; CERULOPLASMIN; BINDING; BRAIN;
D O I
10.1080/00365513.2020.1821397
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
The assay in serum of non-caeruloplasmin copper, as exchangeable copper after complexation with EDTA (ExCu) and total copper has been evaluated and compared in patients with varying c-reactive protein(CRP). Measurement of ExCu and total copper, range 0.2-47.2 mu mol/L, was developed using ICP-MS. The chelating agents EDTA and TEPA were compared over 0.0-10 g/L after incubation with serum for 60 mins followed by ultrafiltration with Amicon 10 kDa filter. The assay for ExCu was optimised with EDTA 3 g/L (8.1 mmol/L) maintained at pH 7.0-8.0 before ultrafiltration. TEPA was not as selective in chelation of copper. Patientsn = 82 were studied in relation to changes in inflammatory marker CRP and a group of patientsn = 37 with normal CRP. The ExCu assay gave excellent recoveries (94-102 % but poor recovery for free uncomplexed copper), good repeatability, limit of quantitation 0.19 mu mol/l with a provisional reference range 0.48 to 1.63 mu mol/L (n = 37 patients). The range for relative exchangeable copper (exchangeable copper divided by total serum copper) was 2.49 to 9.96 %. ExCu was elevated in conditions with increased CRP greater than 100 mg/L suggesting an effect of inflammation on the free copper fraction. A reliable and reproducible assay for ExCu and total copper has been developed. The upregulated inflammatory state increases the ExCu suggesting excess free copper.
引用
收藏
页码:630 / 639
页数:10
相关论文
共 25 条
[1]   Copper binding components of blood plasma and organs, and their responses to influx of large doses of 65Cu, in the mouse [J].
Cabrera, Anthony ;
Alonzo, Erin ;
Sauble, Eric ;
Chu, Yu Ling ;
Nguyen, Dionne ;
Linder, Maria C. ;
Sato, Dee S. ;
Mason, Andrew Z. .
BIOMETALS, 2008, 21 (05) :525-543
[2]   Free copper in serum: An analytical challenge and its possible applications [J].
Catalani, Simona ;
Paganelli, Matteo ;
Gilberti, Maria Enrica ;
Rozzini, Luca ;
Lanfranchi, Francesco ;
Padovani, Alessandro ;
Apostoli, Pietro .
JOURNAL OF TRACE ELEMENTS IN MEDICINE AND BIOLOGY, 2018, 45 :176-180
[3]   Copper transport to the brain by the blood-brain barrier and blood-CSF barrier [J].
Choi, Byuny-Sun ;
Zheng, Wei .
BRAIN RESEARCH, 2009, 1248 :14-21
[4]  
CHUTKOW JG, 1978, P SOC EXP BIOL MED, V158, P113
[5]   Effect of different glycation agents on Cu(II) binding to human serum albumin, studied by liquid chromatography, nitrogen microwave-plasma atomic-emission spectrometry, inductively-coupled-plasma mass spectrometry, and high-resolution molecular-mass spectrometry [J].
Corrales Escobosa, Alma Rosa ;
Wrobel, Katarzyna ;
Yanez Barrientos, Eunice ;
Jaramillo Ortiz, Sarahi ;
Ramirez Segovia, Alejandra Sarahi ;
Wrobel, Kazimierz .
ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2015, 407 (04) :1149-1157
[6]   The significance of copper chelators in clinical and experimental application [J].
Ding, Xueqin ;
Xie, Huiqi ;
Kang, James .
JOURNAL OF NUTRITIONAL BIOCHEMISTRY, 2011, 22 (04) :301-310
[7]   Relative exchangeable copper: A new highly sensitive and highly specific biomarker for Wilson's disease diagnosis [J].
El Balkhi, Souleiman ;
Trocello, Jean-Marc ;
Poupon, Joel ;
Chappuis, Philippe ;
Massicot, France ;
Girardot-Tinant, Nadege ;
Woimant, France .
CLINICA CHIMICA ACTA, 2011, 412 (23-24) :2254-2260
[8]   Human Plasma Copper Proteins Speciation by Size Exclusion Chromatography Coupled to Inductively Coupled Plasma Mass Spectrometry. Solutions for Columns Calibration by Sulfur Detection [J].
El Balkhi, Souleiman ;
Poupon, Joel ;
Trocello, Jean-Marc ;
Massicot, France ;
Woimant, France ;
Laprevote, Olivier .
ANALYTICAL CHEMISTRY, 2010, 82 (16) :6904-6910
[9]   Determination of ultrafiltrable and exchangeable copper in plasma: stability and reference values in healthy subjects [J].
El Balkhi, Souleiman ;
Poupon, Joel ;
Trocello, Jean-Marc ;
Leyendecker, Angelique ;
Massicot, France ;
Galliot-Guilley, Martine ;
Woimant, France .
ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2009, 394 (05) :1477-1484
[10]   SIMPLE ASSAY OF SERUM COPPER FRACTIONS BY ULTRAFILTRATION AND FLAMELESS ATOMIC-ABSORPTION [J].
FAVIER, A ;
RUFFIEUX, D .
BIOLOGICAL TRACE ELEMENT RESEARCH, 1988, 18 :145-160