Sub-second adsorption for the pico-molar monitoring of diltiazem in pharmaceutical preparations by fast Fourier transformation with the use of continuous cyclic voltammetry at an Au microelectrode in a flowing system

被引:31
作者
Norouzi, P. [1 ,2 ]
Ganjali, M. R. [1 ,2 ]
Larijani, B. [2 ]
Mirabi-Semnakolaii, A. [3 ]
Mirnaghi, F. S. [1 ]
Mohammadi, A. [4 ]
机构
[1] Univ Tehran, Dept Chem, Ctr Excellence Electrochem, Tehran, Iran
[2] Univ Tehran Med Sci, Endocrinol & Metab Res Ctr, Tehran, Iran
[3] Islamic Azad Univ, Dept Chem, Fac Sci, Sci & Res Branch, Tehran, Iran
[4] Med Sci Univ Tehran, Sch Pharm, Dept Drug & Food Control, Tehran, Iran
来源
PHARMAZIE | 2008年 / 63卷 / 09期
关键词
D O I
10.1691/ph.2008.7572
中图分类号
R914 [药物化学];
学科分类号
100701 ;
摘要
Diltiazem was determined at the sub-nanomolar level for the first time by a new technique, involving fast Fourier continuous cyclic voltammetry in a flow-injection system. The best performance was achieved with the basic parameters being set at pH value of 2.0, scan rate value of 35 V/s, accumulation potential of 300 mV and accumulation time of 0.8 s. This paper additionally introduces a special computer based numerical method for the calculation of the analyte signal and the noise reduction. Concerning the electrode response calculations were carried out according to the partial and total charge exchanges on the electrode surface after subtraction of background current from that of noise. Furthermore, to obtain a sensitive determination, the currents integration range included all potential scan ranges, even oxidation and reduction of the Au surface electrode, during the measurements. In general, the potential waveform includes the potential steps for cleaning, accumulation and the step of the potential ramp of the analyte. This potential waveform was applied to an Au disk microelectrode in a continuous way. Finally, the method was found to be linear for the concentration range of 1-41450 pg/ml (r = 0.9986), while showing a limit of detection and quantitation of 0.29 and 1 pg/ml, respectively.
引用
收藏
页码:633 / 637
页数:5
相关论文
共 35 条
  • [11] *ICH, 1996, TOP Q2 B VAL AN PROC
  • [12] Lipkowski J., 1992, ADSORPTION MOL METAL
  • [13] Miller J.C., 1984, STAT ANAL CHEM, P82
  • [14] A stability-indicating high performance liquid chromatographic (HPLC) assay for the simultaneous determination of atorvastatin and amlodipine in commercial tablets
    Mohammadi, A.
    Rezanour, N.
    Dogaheh, M. Ansari
    Bidkorbeh, F. Ghorbani
    Hashem, M.
    Walker, R. B.
    [J]. JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2007, 846 (1-2): : 215 - 221
  • [15] A stability-indicating high performance liquid chromatographic assay for the determination of orlistat in capsules
    Mohammadi, A.
    Haririan, I.
    Rezanour, N.
    Ghiasi, L.
    Walker, R. B.
    [J]. JOURNAL OF CHROMATOGRAPHY A, 2006, 1116 (1-2) : 153 - 157
  • [16] An LC-MS-MS method for the determination of cyclizine in human serum
    Mohammadi, A
    Kanfer, I
    Sewram, V
    Walker, RB
    [J]. JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2005, 824 (1-2): : 148 - 152
  • [17] Subsecond FFT-adsorptive voltammetric technique as a novel method for subnano level monitoring of piroxicam in its tablets and bulk form at Au microelectrode in flowing solutions
    Norouzi, P.
    Ganjali, M. R.
    Labbafi, S.
    Mohammadi, A.
    [J]. ANALYTICAL LETTERS, 2007, 40 (04) : 747 - 762
  • [18] Development of fast Fourier transform continuous cyclic voltammetry at Au microelectrode in flowing solutions as a novel method for sub-nanomolar monitoring of lidocaine in injection and biological fluids
    Norouzi, P.
    Ganjali, M. R.
    Daneshgar, P.
    Dinarvand, R.
    Moosavi-Movahedi, A. A.
    Saboury, A. A.
    [J]. ANALYTICA CHIMICA ACTA, 2007, 590 (01) : 74 - 80
  • [19] Fast fourier transform continuous cyclic voltammetry development as a highly sensitive detection system for ultra trace monitoring of thiamine
    Norouzi, P.
    Ganjali, M. R.
    Daneshgar, P.
    Mohammadi, A.
    [J]. ANALYTICAL LETTERS, 2007, 40 (03) : 547 - 559
  • [20] Nano-level detection of naltrexone hydrochloride in its pharmaceutical preparation at Au microelectrode in flowing solutions by fast Fourier transforms continuous cyclic voltammetry as a novel detector
    Norouzi, P.
    Ganjali, M. R.
    Zare, M.
    Mohammadi, A.
    [J]. JOURNAL OF PHARMACEUTICAL SCIENCES, 2007, 96 (08) : 2009 - 2017