Cyclic fatigue of Reciproc and WaveOne reciprocating instruments

被引:162
|
作者
Plotino, G. [1 ]
Grande, N. M. [1 ]
Testarelli, L. [1 ]
Gambarini, G. [1 ]
机构
[1] Univ Roma La Sapienza, Dept Endodont, I-00197 Rome, Italy
关键词
cyclic fatigue; nickel-titanium; reciprocation; NITI ROTARY INSTRUMENTS; NICKEL-TITANIUM FILES; SHAPING ABILITY; RESISTANCE; FRACTURE; CANALS; K3;
D O I
10.1111/j.1365-2591.2012.02015.x
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Plotino G, Grande NM, Testarelli L, Gambarini G. Cyclic fatigue of Reciproc and WaveOne reciprocating instruments. International Endodontic Journal, 45, 614618, 2012. Abstract Aim To evaluate the cyclic fatigue resistance of Reciproc (R) and WaveOne (R) instruments in simulated root canals. Methodology Two groups of 15 NiTi endodontic instruments of identical tip size of 0.25 mm were tested, group A; Reciproc (R) R25 and group B: WaveOne (R) primary. Cyclic fatigue testing was performed in a stainless steel artificial canal manufactured by reproducing the instruments size and taper. A simulated root canal with a 60 degrees angle of curvature and 5-mm radius of curvature was constructed for both the instruments tested. The centre of the curvature was 5 mm from the tip of the instrument and the curved segment of the canal was approximately 5 mm in length. The Reciproc (R) instruments were activated using the preset programme specific for the Reciproc (R) instruments, whilst the WaveOne (R) instruments were activated using the preset programme specific for the WaveOne (R) instruments. All instruments were rotated until fracture occurred and the time to fracture (TtF) and the length of the fractured tip were recorded and registered. Means and standard deviations of TtF and fragment length were calculated for each system and data were subjected to Students t-test (P < 0.05). Results A statistically significant difference (P < 0.05) was noted between Reciproc (R) and WaveOne (R) instruments. Reciproc (R) R25 instruments were associated with a significant increase in the mean time to fracture when compared with primary WaveOne (R) instruments (130.8 +/- 18.4 vs. 97.8 +/- 15.9 s). There was no significant difference (P > 0.05) in the mean length of the fractured fragments between the instruments. Conclusions Reciproc (R) instruments were associated with a significantly higher cyclic fatigue resistance than WaveOne (R) instruments.
引用
收藏
页码:614 / 618
页数:5
相关论文
共 50 条
  • [1] Cyclic Fatigue Resistance of Reciproc Blue, Reciproc, and WaveOne Gold Reciprocating Instruments
    Keskin, Cangul
    Ivan, Ugur
    Demiral, Murat
    Keles, Ali
    JOURNAL OF ENDODONTICS, 2017, 43 (08) : 1360 - 1363
  • [2] Bending Resistance and Cyclic Fatigue Life of Reciproc, Unicone, and WaveOne Reciprocating Instruments
    Silva, Emmanuel Joao Nogueira Leal
    Villarino, Laise Silva
    Vieira, Victor Talarico Leal
    Accorsi-Mendonca, Thais
    Antunes, Henrique dos Santos
    De-Deus, Gustavo
    Lopes, Helio Pereira
    JOURNAL OF ENDODONTICS, 2016, 42 (12) : 1789 - 1793
  • [3] Effect of sodium hypochlorite and saline on cyclic fatigue resistance of WaveOne Gold and Reciproc reciprocating instruments
    Elnaghy, A. M.
    Elsaka, S. E.
    INTERNATIONAL ENDODONTIC JOURNAL, 2017, 50 (10) : 991 - 998
  • [4] Cutting Efficiency of Reciproc and WaveOne Reciprocating Instruments
    Plotino, Gianluca
    Rubini, Alessio Giansiracusa
    Grande, Nicola M.
    Testarelli, Luca
    Gambarini, Gianluca
    JOURNAL OF ENDODONTICS, 2014, 40 (08) : 1228 - 1230
  • [5] Cyclic Fatigue Resistance of Reciproc, WaveOne, and WaveOne Gold Nickel-Titanium Instruments
    Ozyurek, Taha
    JOURNAL OF ENDODONTICS, 2016, 42 (10) : 1536 - 1539
  • [6] Bending Resistance and Dynamic and Static Cyclic Fatigue Life of Reciproc and WaveOne Large Instruments
    De-Deus, Gustavo
    Leal Vieira, Victor Talarico
    Nogueira da Silva, Emmanuel Joao
    Lopes, Helio
    Elias, Carlos Nelson
    Moreira, Edson Jorge
    JOURNAL OF ENDODONTICS, 2014, 40 (04) : 575 - 579
  • [7] Bending resistance and cyclic fatigue resistance of WaveOne Gold, Reciproc Blue, and HyFlex EDM instruments
    Oh, Soram
    Kum, Kee-Yeon
    Kim, Hyun-Jung
    Moon, Su-Young
    Kim, Hyeon-Cheol
    Chaniotis, Antonis
    Perinpanayagam, Hiran
    Pedulla, Eugenio
    Chang, Seok Woo
    JOURNAL OF DENTAL SCIENCES, 2020, 15 (04) : 472 - 478
  • [8] Cyclic Fatigue Resistance of OneShape, HyFlex EDM, WaveOne Gold, and Reciproc Blue Nickel-titanium Instruments
    Gundogar, Mustafa
    Ozyurek, Taha
    JOURNAL OF ENDODONTICS, 2017, 43 (07) : 1192 - 1196
  • [9] Ex vivo shaping ability of reciprocating instruments operated by new users: Reciproc versus WaveOne
    Mary S. H. Lam
    Jeffrey W. W. Chang
    Gary S. P. Cheung
    Clinical Oral Investigations, 2021, 25 : 2791 - 2799
  • [10] Cyclic fatigue resistance of new reciprocating files (Reciproc Blue, WaveOne Gold, and SmartTrack) in two different curved canals
    Topcuoglu, Hueseyin Sinan
    Demirbuga, Sezer
    Duzgun, Salih
    Topcuoglu, Gamze
    JOURNAL OF INVESTIGATIVE AND CLINICAL DENTISTRY, 2018, 9 (03) : e12344