Influence of field-of-view and section thickness of diagnostic imaging on thermal neutron flux estimation in dose-planning for boron neutron capture therapy

被引:1
作者
Sato, Hiroyuki [1 ,2 ]
Takata, Takushi [3 ]
Sakurai, Yoshinori [3 ]
机构
[1] Kyoto Univ, Grad Sch Engn, Kyoto, Japan
[2] Tottori Univ Hosp, Dept Radiol, Tottori, Japan
[3] Kyoto Univ, Inst Integrated Radiat & Nucl Sci, Osaka, Japan
基金
日本学术振兴会;
关键词
Boron neutron capture therapy; Epi-thermal neutron flux; Field-of-view; Section thickness; Thermal neutron flux; Treatment planning; ELECTRON-DENSITY; DOSIMETRY; BNCT; PHANTOM; CANCER;
D O I
10.1007/s12194-018-00494-3
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Radiation treatment planning for boron neutron capture therapy (BNCT) often uses computed tomography (CT) images reconstructed utilizing various section thickness and field-of-view (FOV) settings. Based on these images, a geometrical model is created by setting material regions manually over the pixel space defined in the treatment planning system. Thus, a setting difference of several pixels inevitably occurs in creation of the model. The influence of different section thicknesses and FOVs on thermal neutron flux estimations using the BNCT planning system was studied here. A virtual phantom was created with six FOV sizes on the planning system. The position of the irradiated side of the phantom surface was shifted by 1-10 pixels along the beam direction or in the opposite direction to simulate the material setting on different pixels in the geometric model. The effect of a one-pixel-difference setting on thermal neutron flux increased with increasing FOV size. Next, a cylindrical and a spherical phantom were scanned, and each CT image was reconstructed with six FOV sizes and seven section thicknesses. The flux changes for all conditions were compared, with an allowable error rate of +/- 0.05, as in conventional X-ray radio therapy. The accuracy of neutron flux estimations was also evaluated by repeating the calculation procedures with CT scanning 5 or 10 times, and was found to be mostly within 0.03, except for the FOV-500 condition (0.074). These results suggested that a smaller FOV and section thickness with realistic conditions could improve evaluation accuracy of the thermal neutron flux for BNCT.
引用
收藏
页码:76 / 85
页数:10
相关论文
共 19 条
  • [1] [Anonymous], 1992, 46 ICRU
  • [2] Boron neutron capture therapy of cancer: Current status and future prospects
    Barth, RF
    Coderre, JA
    Vicente, MGH
    Blue, TE
    [J]. CLINICAL CANCER RESEARCH, 2005, 11 (11) : 3987 - 4002
  • [3] Boron neutron-capture therapy (BNCT) for glioblastoma multiforme (GBM) using the epithermal neutron beam at the Brookhaven National Laboratory
    Chadha, M
    Capala, J
    Coderre, JA
    Elowitz, EH
    Iwai, J
    Joel, DD
    Liu, HYB
    Wielopolski, L
    Chanana, AD
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1998, 40 (04): : 829 - 834
  • [4] The radiation biology of boron neutron capture therapy
    Coderre, JA
    Morris, GM
    [J]. RADIATION RESEARCH, 1999, 151 (01) : 1 - 18
  • [5] AN ELECTRON-DENSITY CALIBRATION PHANTOM FOR CT-BASED TREATMENT PLANNING COMPUTERS
    CONSTANTINOU, C
    HARRINGTON, JC
    DEWERD, LA
    [J]. MEDICAL PHYSICS, 1992, 19 (02) : 325 - 327
  • [6] BNCT: Status and dosimetry requirements
    Gahbauer, R
    Gupta, N
    Blue, T
    Goodman, J
    Grecula, J
    Soloway, AH
    Wambersie, A
    [J]. RADIATION PROTECTION DOSIMETRY, 1997, 70 (1-4) : 547 - 554
  • [7] Some biological effects of nuclear disintegration products on neoplastic tissue
    Kruger, PG
    [J]. PROCEEDINGS OF THE NATIONAL ACADEMY OF SCIENCES OF THE UNITED STATES OF AMERICA, 1940, 26 : 181 - 192
  • [8] Verification of the computational dosimetry system in JAERI (JCDS) for boron neutron capture therapy
    Kumada, H
    Yamamoto, K
    Matsumura, A
    Yamamoto, T
    Nakagawa, Y
    Nakai, K
    Kageji, T
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2004, 49 (15) : 3353 - 3365
  • [9] Improvement of dose calculation accuracy for BNCT dosimetry by the multi-voxel method in JCDS
    Kumada, H
    Yamamoto, K
    Yamamoto, T
    Nakai, K
    Nakagawa, Y
    Kageji, T
    Matsumura, A
    [J]. APPLIED RADIATION AND ISOTOPES, 2004, 61 (05) : 1045 - 1050
  • [10] Verification of the accuracy of BNCT treatment planning system THORplan
    Li, H. S.
    Liu, Y-W H.
    Lee, C. Y.
    Lin, T. Y.
    Hsu, F. Y.
    [J]. APPLIED RADIATION AND ISOTOPES, 2009, 67 (7-8) : S122 - S125