An Infrared-heated Sample Introduction System for Inductively Coupled Plasma Mass Spectrometry

被引:5
作者
Teuma-Castelletti, Robert [1 ]
Al Hejami, Ahmed [1 ]
Beauchemin, Diane [1 ]
机构
[1] Queens Univ, Dept Chem, Kingston, ON K7L 3N6, Canada
基金
加拿大自然科学与工程研究理事会;
关键词
OPTICAL-EMISSION SPECTROMETRY; ANALYTICAL PERFORMANCE; TOTAL-CONSUMPTION; PREEVAPORATION; NEBULIZATION; WATER; ETV;
D O I
10.46770/AS.2022.233
中图分类号
O433 [光谱学];
学科分类号
0703 ; 070302 ;
摘要
This work involves heating the entire commercially available sample introduction system on a Varian 820MS instrument with off-the-shelf infrared heaters. This heating improves sensitivities by a factor of 2 to 16 times, with high mass elements receiving the greatest improvement. It also improves detection limits by a factor of 1.1 to 3.5 times for most analytes. At the price of doubled oxide formation, the formation of doubly charged species was decreased by a factor of 3. The Be-9(+)/Li-7(+) signal ratio, an indicator of plasma robustness, increased from 0.174 to 0.235 with infrared heating. The infrared-heated system enabled the accurate analysis of a drinking water standard reference material (SRM) (for Be, Mg, Al, Cr, Mn, Co, Cu, Zn and As) and of a waste water SRM (for Be, Mg, Cr, Mn, Co, Cu, Zn, Cd, and Pb) by simple external calibration.
引用
收藏
页码:371 / 377
页数:7
相关论文
共 26 条
[1]   How much aqueous sample can an inductively coupled plasma withstand? [J].
Al Hejami, Ahmed ;
Burgener, John ;
Burgener, Mirah J. ;
Beauchemin, Diane .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2020, 35 (07) :1300-1305
[2]   A total consumption (up to 75 μL min-1) infrared-heated sample introduction system for inductively coupled plasma optical emission spectrometry [J].
Al Hejami, Ahmed ;
Burgener, Mirah J. ;
Burgener, John ;
Beauchemin, Diane .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2020, 35 (06) :1125-1130
[3]   Infrared heating of commercially available spray chambers to improve the analytical performance of inductively coupled plasma optical emission spectrometry [J].
Al Hejami, Ahmed ;
Beauchemin, Diane .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2018, 33 (11) :2008-2014
[4]  
Al Hejami A, 2018, J ANAL ATOM SPECTROM, V33, P738, DOI [10.1039/c8ja00044a, 10.1039/C8JA00044A]
[5]   Improvements to the analytical performance of inductively coupled plasma optical emission spectrometry by coupling a multi-mode sample introduction system to an infrared heated pre-evaporation tube [J].
Anderlini, Tia K. ;
Beauchemin, Diane .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2018, 33 (06) :1068-1075
[6]   Improvement of sample introduction to inductively coupled plasma optical emission spectrometry using an ultrasonic nebulizer with an infrared heated pre-evaporation tube [J].
Anderlini, Tia K. ;
Beauchemin, Diane .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2018, 33 (01) :127-134
[7]   Ultrasonic nebulization with an infrared heated pre-evaporation tube for sample introduction in ICP-OES: application to geological and environmental samples [J].
Asfaw, Alemayehu ;
MacFarlane, William R. ;
Beauchemin, Diane .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2012, 27 (08) :1254-1263
[8]   Comparison of pneumatic and ultrasonic nebulizations in inductively coupled plasma atomic emission spectrometry - matrix effects and plasma parameters [J].
Borkowska-Burnecka, Jolanta ;
Lesniewicz, Anna ;
Zymicki, Wieslaw .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 2006, 61 (05) :579-587
[9]   SAMPLE INTRODUCTION - THE ACHILLES HEEL OF ATOMIC SPECTROSCOPY [J].
BROWNER, RF ;
BOORN, AW .
ANALYTICAL CHEMISTRY, 1984, 56 (07) :A786-&
[10]  
Dean JR, 2005, ANAL TECH SCI, P1, DOI 10.1002/047009351X