Tomographic spectral Imaging with multivariate statistical analysis: Comprehensive 3D microanalysis

被引:79
作者
Kotula, PG [1 ]
Keenan, MR [1 ]
Michael, JR [1 ]
机构
[1] Sandia Natl Labs, Mat Characterizat Dept, Albuquerque, NM 87185 USA
关键词
tomography; spectral imaging; multivariate statistical analysis; multivariate curve resolution; 3D chemical analysis; tomographic spectral imaging; serial sectioning; 3D microanalysis;
D O I
10.1017/S1431927606060193
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
A comprehensive three-dimensional (31)) microanalysis procedure using a combined scanning electron microscope (SEM)/focused ion beam (FIB) system equipped with an energy-dispersive X-ray spectrometer (EDS) has been developed. The FIB system was used First to prepare a site-specific region for X-ray microanalysis followed by the acquisition of an electron-beam generated X-ray spectral image. A small section of material was then removed by the FIB, followed by the acquisition of another X-ray spectral image. This serial sectioning procedure was repeated 10-12 times to sample a volume of material. The series of two-spatial-dimension spectral images were then concatenated into a single data set consisting of a series of volume elements or voxels each with an entire X-ray spectrum. This four-dimensional (three real space and one spectral dimension) spectral image was then comprehensively analyzed with Sandia's automated X-ray spectral image analysis software. This technique was applied to a simple Cu-Ag eutectic and a more complicated localized corrosion study where the powerful site-specific comprehensive analysis capability of tomographic spectral imaging (TSI) combined with multivariate statistical analysis is demonstrated.
引用
收藏
页码:36 / 48
页数:13
相关论文
共 32 条
[1]  
[Anonymous], MICROSCOPY MICROANAL
[2]   MICRO-ANALYSIS BY A FLYING-SPOT X-RAY METHOD [J].
COSSLETT, VE ;
DUNCUMB, P .
NATURE, 1956, 177 (4521) :1172-1173
[3]   Serial block-face scanning electron microscopy to reconstruct three-dimensional tissue nanostructure [J].
Denk, W ;
Horstmann, H .
PLOS BIOLOGY, 2004, 2 (11) :1900-1909
[4]   Quantitative three-dimensional reconstruction of geometrically complex structures with nanoscale resolution [J].
Dunn, DN ;
Shiflet, GJ ;
Hull, R .
REVIEW OF SCIENTIFIC INSTRUMENTS, 2002, 73 (02) :330-334
[5]   Reconstruction of three-dimensional chemistry and geometry using focused ion beam microscopy [J].
Dunn, DN ;
Hull, R .
APPLIED PHYSICS LETTERS, 1999, 75 (21) :3414-3416
[6]  
HULL R, 2001, MICROSC MICROANAL S2, V7, P934
[7]   Accounting for Poisson noise in the multivariate analysis of ToF-SIMS spectrum images [J].
Keenan, MR ;
Kotula, PG .
SURFACE AND INTERFACE ANALYSIS, 2004, 36 (03) :203-212
[8]  
KEENAN MR, 2004, Patent No. 6675106
[9]  
KEENAN MR, 2003, Patent No. 6584413
[10]  
KEENAN MR, 2004, MICROSC MICROANAL S2, V10, P874