Plausibility check for error compensation in electromagnetic navigation in endoscopic sinus surgery

被引:0
作者
Mucha, Dirk [1 ]
Kosmecki, Bartosz [2 ]
Bier, Juergen [1 ]
机构
[1] Charite, Dept Maxillofacial Surg Clin Nav & Robot, Berlin, Germany
[2] Fraunhofer Inst Prod Engn & Design Technol, Berlin, Germany
关键词
Electromagnetic tracking; Surgical navigation; Plausibility check; Distortion;
D O I
暂无
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
In this contribution we present a method for checking plausibility of electromagnetically measured position values in the vicinity of metallic surgical instruments and OR equipment Furthermore, the method is used in a simple and applicable error compensation algorithm. We present the results of a first analysis of the method for navigation in ENT surgery. The concept is based on redundant pose information of a tool. This is achieved by a rigid electromagnetic localizer consisting of two sensor elements each measuring the pose of the tool tip separately. The length of the difference vector of the measured positions is a measure for the plausibility. We applied the method in a navigation system for FESS and determined the correlation of the measuring error caused by metallic instruments. The experiment showed that the computed plausibility value shows sufficient correlation to relevant measuring errors.
引用
收藏
页码:316 / 318
页数:3
相关论文
共 8 条
  • [1] Systematic distortions in magnetic position digitizers
    Birkfellner, W
    Watzinger, F
    Wanschitz, F
    Enislidis, G
    Kollmann, C
    Rafolt, D
    Nowotny, R
    [J]. MEDICAL PHYSICS, 1998, 25 (11) : 2242 - 2248
  • [2] Birkfellner W, 1998, LECT NOTES COMPUT SC, V1496, P343, DOI 10.1007/BFb0056218
  • [3] Accuracy assessment protocols for electromagnetic tracking systems
    Frantz, DD
    Wiles, AD
    Leis, SE
    Kirsch, SR
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2003, 48 (14) : 2241 - 2251
  • [4] Evaluation of a miniature electromagnetic position tracker
    Hummel, J
    Figl, M
    Kollmann, C
    Bergmann, H
    Birkfellner, W
    [J]. MEDICAL PHYSICS, 2002, 29 (10) : 2205 - 2212
  • [5] New form factors for sensors and field generators of a magnetic tracking system
    Kirsch, SR
    Schilling, C
    Brunner, G
    [J]. MEDICAL IMAGING 2005: VISUALIZATION, IMAGE-GUIDED PROCEDURES, AND DISPLAY, PTS 1 AND 2, 2005, 5744 : 216 - 224
  • [6] Koele W, 2002, RHINOLOGY, V40, P1
  • [7] Mucha D, 2005, P COMP ASS RAD SURG
  • [8] Stability of miniature electromagnetic tracking systems
    Schicho, K
    Figl, M
    Donat, M
    Birkfellner, W
    Seemann, R
    Wagner, A
    Bergmann, H
    Ewers, R
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2005, 50 (09) : 2089 - 2098