Electromagnetic interference of digital and analog cellular telephones with implantable cardioverter defibrillators: In vitro and in vivo studies

被引:35
作者
Barbaro, V
Bartolini, P
Bellocci, F
Caruso, F
Donato, A
Gabrielli, D
Militello, C
Montenero, AS
Zecchi, P
机构
[1] Ist Super Sanita, Lab Ingn Biomed, I-00161 Rome, Italy
[2] Policlin Univ Agostino Gemelli, Cardiol Inst, Rome, Italy
来源
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY | 1999年 / 22卷 / 04期
关键词
implantable defibrillator; electromagnetic interference; cellular telephones;
D O I
10.1111/j.1540-8159.1999.tb00504.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
The present study examines the potential electromagnetic interference effects induced by cellular telephones on ICDs. We developed ad hoc protocols to conduct both in vitro and in vivo trials on most of the implantable cardioverter defibrillators available on the international market. Trials were conducted with three cellular telephones: two GSM (Global System for Mobile Communication) and one TAGS (Total Access Communication System). A human trunk simulator was used to carry out in vitro observations on six ICDs from five manufacturers. In vivo tests were conducted on 13 informed patients with eight different ICD models. During the trials in air, GSM telephones induced interference effects on 4 out of the 6 cardioverter defibrillators tested. Specifically, pulse inhibition, reprogramming, false ventricular fibrillation, and ventricular tachycardia detections occurred, which would have entailed inappropriate therapy delivery had this been activated. Effects were circumscribed to the area closely surrounding the connectors. When the ICD was immersed in saline solution, no effects were observed. Three cases of just ventricular triggering with the interfering signal were observed in vivo.
引用
收藏
页码:626 / 634
页数:9
相关论文
共 18 条
  • [1] DO EUROPEAN GSM MOBILE CELLULAR PHONES POSE A POTENTIAL RISK TO PACEMAKER PATIENTS
    BARBARO, V
    BARTOLINI, P
    DONATO, A
    MILITELLO, C
    ALTAMURA, G
    AMMIRATI, F
    SANTINI, M
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1995, 18 (06): : 1218 - 1224
  • [2] Electromagnetic interference of analog cellular telephones with pacemakers
    Barbaro, V
    Bartolini, P
    Donato, A
    Militello, C
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1996, 19 (10): : 1410 - 1418
  • [3] BARBARO V, 1996, BEMS VICT CAN JUN, P55
  • [4] BASSEN HI, 1995, CIRCULATION, V8, P738
  • [5] BOSTROM U, 1991, CLIN ENG UPDATE, V10, P1
  • [6] DECHAUX C, 1993, ALCATEL PROSPETTIVE, P118
  • [7] EICHER B, 1994, BEMS COP DENM JUN, P67
  • [8] Electromagnetic interference from welding and motors on implantable cardioverter-defibrillators as tested in the electrically hostile work site
    Fetter, JG
    Benditt, DG
    Stanton, MS
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1996, 28 (02) : 423 - 427
  • [9] GIORDANI M, 1993, NOTIZIARIO TECNICO S, V1, P23
  • [10] GRIMALDI F, 1991, SCUOLA SUPERIORE, P6