Fabrication of ultra-low radioactivity detector holders for Edelweiss-II

被引:13
作者
Navick, XF [1 ]
Carty, M
Chapellier, M
Chardin, G
Goldbach, C
Granelli, R
Hervé, S
Karolak, M
Nollez, G
Nizery, F
Riccio, C
Starzynski, P
Villar, V
机构
[1] CEA, Ctr Etudes Nucl Saclay, DSM, DAPNIA, F-91191 Gif Sur Yvette, France
[2] CEA, Ctr Etudes Nucl Saclay, DSM, DRECAM, F-91191 Gif Sur Yvette, France
[3] CNRS, INSU, Inst Astrophys, F-75014 Paris, France
关键词
cryogenic detector; bolometer; holder; low radioactivity; wimp; dark matter;
D O I
10.1016/j.nima.2003.11.290
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
In its first stage, the EDELWEISS-II dark matter experiment will use 21 x 320 g ionization-heat bolometers with NTD thermal sensors. To improve the present sensitivity of EDELWEISS to WIMP interactions by a further factor 100, ultra-low radioactivity detector holders exclusively made of copper and Teflon have been designed. The new design is using the relative expansion coefficients of copper, Teflon and germanium to hopefully ensure a dissipation-free detector holding. In order to validate this new holder, we need to compare the vibration behavior of the old holder and the new one. The amplitude of the bolometer vibrations is measured using the variation of capacitance between electrodes sputtered on the Ge crystal and reference electrodes fixed above the detector holders. We present noise measurements and compare them to the equivalent measurements using the detector holders of the Edelweiss-I experiment, which gave very satisfying results. (C) 2003 Elsevier B.V. All rights reserved.
引用
收藏
页码:189 / 192
页数:4
相关论文
共 5 条
[1]   Improved exclusion limits from the EDELWEISS WIMP search [J].
Benoit, A ;
Bergé, L ;
Broniatowski, A ;
Chabert, L ;
Chambon, B ;
Chapellier, M ;
Chardin, G ;
Charvin, R ;
De Jésus, M ;
Di Stefano, P ;
Drain, D ;
Dumoulin, L ;
Gascon, J ;
Gerbier, G ;
Gerlic, E ;
Goldbach, C ;
Goyot, M ;
Gros, M ;
Hadjout, JP ;
Hervé, S ;
Juillard, A ;
de Lesquen, A ;
Loidl, M ;
Mallet, J ;
Marnieros, S ;
Martineau, O ;
Mirabolfathi, N ;
Mosca, L ;
Navick, XF ;
Nollez, G ;
Pari, P ;
Riccio, C ;
Sanglard, V ;
Stern, M ;
Vagneron, L .
PHYSICS LETTERS B, 2002, 545 (1-2) :43-49
[2]  
BENOIT A, ASTROPH0206271
[3]  
GASCON J, NUCL I MEHT A
[4]   320 g ionization-heat bolometers design for the EDELWEISS experiment [J].
Navick, XF ;
Chapellier, M ;
Déliot, F ;
Hervé, S ;
Miramonti, L .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2000, 444 (1-2) :361-363
[5]   A solution to the dead-layer problem in ionization and phonon-based dark matter detectors [J].
Shutt, T ;
Emes, J ;
Haller, EE ;
Hellmig, J ;
Sadoulet, B ;
Seitz, D ;
Young, BA ;
White, S .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2000, 444 (1-2) :340-344