The influence of MRI scan position on image registration accuracy, target delineation and calculated dose in prostatic radiotherapy

被引:36
作者
Hanvey, S. [1 ]
Sadozye, A. H. [2 ]
McJury, M. [1 ]
Glegg, M. [1 ]
Foster, J. [1 ]
机构
[1] Beatson W Scotland Canc Ctr, Dept Clin Phys & Bioengn, Glasgow G12 0YN, Lanark, Scotland
[2] Beatson W Scotland Canc Ctr, Dept Clin Oncol, Glasgow G12 0YN, Lanark, Scotland
关键词
VOLUME DELINEATION; RADIATION-THERAPY; CT; QUALITY; CANCER; TUMOR; BASE;
D O I
10.1259/bjr/26802977
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objective: To investigate the necessity of performing MRI in the radiotherapy position when using MRI for prostatic radiotherapy. Methods: 20 prostate patients received a CT, diagnostic MRI and an MRI scan in the radiotherapy position. The quality of registration between CT and MRI was compared between the two MRI set-ups. The prostate and seminal vesicles were contoured using all scans and intensity modulated radiotherapy (IMRT) plans were generated. Changes in the target volume and IMRT plans were investigated. Two-tailed paired Student's t-tests determined the statistical significance. Results: There was a decrease in the mean distance from the centre of the bony anatomy between CT and MRI (from 3.9 to 1.9 mm, p-value < 0.0001) when the MRI scan was acquired in the radiotherapy position. Assuming that registering CT with an MRI scan in the radiotherapy position is the gold standard for delineating the prostate and seminal vesicles, using a planning target volume delineated on the CT with a diagnostic MRI scan viewed separately, resulted in a mean conformation number of 0.80 instead of the expected 0.98 (p < 0.0001). Conclusion: By registering CT with an MRI scan in the radiotherapy position, there is a statistically significant improvement in the registration and IMRT quality. Advances in knowledge: To achieve an acceptable registration and IMRT quality in prostatic radiotherapy, neither CT with a separate diagnostic MRI nor CT registered to a diagnostic MRI will suffice. Instead, a CT registered with an MRI in the radiotherapy position should be used.
引用
收藏
页码:E1256 / E1262
页数:7
相关论文
共 26 条
[1]   The value of magnetic resonance imaging in target volume delineation of base of tongue tumours - A study using flexible surface coils [J].
Ahmed, Merina ;
Schmidt, Maria ;
Sohaib, Aslam ;
Kong, Christine ;
Burke, Kevin ;
Richardson, Cheryl ;
Usher, Marianne ;
Brennan, Sinead ;
Riddell, Angela ;
Davies, Mark ;
Newbold, Kate ;
Harrington, Kevin J. ;
Nutting, Christopher M. .
RADIOTHERAPY AND ONCOLOGY, 2010, 94 (02) :161-167
[2]  
American Association of Physicists in Medicine, 2010, ACC TEST QUAL ASS PR
[3]   The use of CT-MR image registration to define target volumes in pelvic radiotherapy in the presence of bilateral hip replacements [J].
Charnley, N ;
Morgan, A ;
Thomas, E ;
Wilson, S ;
Bacon, S ;
Wilson, D ;
Bottomley, D .
BRITISH JOURNAL OF RADIOLOGY, 2005, 78 (931) :634-636
[4]   Impact of magnetic resonance imaging versus CT on nasopharyngeal carcinoma: Primary tumor target delineation for radiotherapy [J].
Chung, NN ;
Ting, LL ;
Hsu, WC ;
Lui, LT ;
Wang, PM .
HEAD AND NECK-JOURNAL FOR THE SCIENCES AND SPECIALTIES OF THE HEAD AND NECK, 2004, 26 (03) :241-246
[5]   Conventional versus hypofractionated high-dose intensity-modulated radiotherapy for prostate cancer: preliminary safety results from the CHHiP randomised controlled trial [J].
Dearnaley, David ;
Syndikus, Isabel ;
Sumo, Georges ;
Bidmead, Margaret ;
Bloomfield, David ;
Clark, Catharine ;
Gao, Annie ;
Hassan, Shama ;
Horwich, Alan ;
Huddart, Robert ;
Khoo, Vincent ;
Kirkbride, Peter ;
Mayles, Helen ;
Mayles, Philip ;
Naismith, Olivia ;
Parker, Chris ;
Patterson, Helen ;
Russell, Martin ;
Scrase, Christopher ;
South, Chris ;
Staffurth, John ;
Hall, Emma .
LANCET ONCOLOGY, 2012, 13 (01) :43-54
[6]   A complete distortion correction for MR images: I. Gradient warp correction [J].
Doran, SJ ;
Charles-Edwards, L ;
Reinsberg, SA ;
Leach, MO .
PHYSICS IN MEDICINE AND BIOLOGY, 2005, 50 (07) :1343-1361
[7]   Aspects of MR image distortions in radiotherapy treatment planning [J].
Fransson, A ;
Andreo, P ;
Pötter, R .
STRAHLENTHERAPIE UND ONKOLOGIE, 2001, 177 (02) :59-73
[8]   Inter-observer variability in the delineation of pharyngo-laryngeal tumor, parotid glands and cervical spinal cord:: Comparison between CT-scan and MRI [J].
Geets, X ;
Daisne, JF ;
Arcangeli, S ;
Coche, E ;
De Poel, M ;
Duprez, T ;
Nardella, G ;
Grégoire, V .
RADIOTHERAPY AND ONCOLOGY, 2005, 77 (01) :25-31
[9]  
Institute of Physics in Engineering and Medicine, 1998, 80 IPEM
[10]   New developments in Mill for target volume delineation in radiotherapy [J].
Khoo, V. S. ;