Hip Squeaking A 10-Year Follow-Up Study

被引:26
作者
Chevillotte, Christophe [1 ,2 ]
Pibarot, Vincent [1 ]
Carret, Jean-Paul [1 ,2 ]
Bejui-Hugues, Jacques [2 ,3 ]
Guyen, Olivier [1 ,2 ]
机构
[1] Hop Edouard Herriot, Dept Orthoped Surg, F-69437 Lyon 03, France
[2] Univ Lyon 1, Lab Biomecan Chocs, LBMC INRETS URMT 9406, F-69365 Lyon, France
[3] ICOT, Latina, Italy
关键词
hip; ceramic; squeaking; total hip arthroplasty; surface bearing; ON-CERAMIC BEARINGS; ACETABULAR COMPONENT; TOTAL JOINT; IN-VITRO; ARTHROPLASTY; ALUMINA; REPLACEMENT; FAILURE; WEAR;
D O I
10.1016/j.arth.2011.11.024
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
The aim of the study was to analyze the incidence of squeaking with ceramic-on-ceramic total hip arthroplasty (THA) after 10 years of follow-up and the potential complications that could occur related to this phenomenon. One hundred THAs implanted between November 1999 and December 2000 were evaluated. Incidence of squeaking was investigated clinically with a questionnaire. Implant positioning was analyzed on x-rays and computer tomography. Of the 100 THAs, 5 patients presented with squeaking. All of them were active, sporty, and heavy men. Functional scores were comparable with nonsqueaking patients. There was no malpositioning on the x-ray analysis, no wear, and no loosening. We could not demonstrate any relation between squeaking and ceramic fracture. Squeaking noise appeared at a mean of 66 months postsurgery. It appears to be an isolated phenomenon without any consequences at 10-year follow-up.
引用
收藏
页码:1008 / 1013
页数:6
相关论文
共 38 条
  • [1] Is Ceramic-On-Ceramic Squeaking Phenomenon Reproducible In Vitro? A Long-Term Simulator Study Under Severe Conditions
    Affatato, S.
    Traina, F.
    Mazzega-Fabbro, C.
    Sergo, V.
    Viceconti, M.
    [J]. JOURNAL OF BIOMEDICAL MATERIALS RESEARCH PART B-APPLIED BIOMATERIALS, 2009, 91B (01) : 264 - 271
  • [2] TREATMENT OF PRIMARY OSTEOARTHRITIS OF THE HIP - A COMPARISON OF TOTAL JOINT AND SURFACE REPLACEMENT ARTHROPLASTY
    AMSTUTZ, HC
    THOMAS, BJ
    JINNAH, R
    KIM, W
    GROGAN, T
    YALE, C
    [J]. JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1984, 66A (02) : 228 - 241
  • [3] Hybrid alumina total hip arthroplasty using a press-fit metal-backed socket in patients younger than 55 years - A six- to 11-year evaluation
    Bizot, P
    Hannouche, D
    Nizard, R
    Witvoet, J
    Sedel, L
    [J]. JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2004, 86B (02): : 190 - 194
  • [4] BOUTIN P, 1972, REV CHIR ORTHOP, V58, P229
  • [5] Boutin P, 2000, CLIN ORTHOP RELAT RE, V3
  • [6] Ceramic-on-Ceramic Total Hip Arthroplasty: Update
    Capello, William N.
    D'Antonio, James A.
    Feinberg, Judy R.
    Manley, Michael T.
    Naughton, Marybetb
    [J]. JOURNAL OF ARTHROPLASTY, 2008, 23 (07) : 39 - 43
  • [7] CHARNLEY J, 1975, CLIN ORTHOP RELAT RE, V170
  • [8] The 2009 Frank Stinchfield Award "Hip Squeaking'' A Biomechanical Study of Ceramic-on-ceramic Bearing Surfaces
    Chevillotte, Christophe
    Trousdale, Robert T.
    Chen, Qingshan
    Guyen, Olivier
    An, Kai-Nan
    [J]. CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2010, 468 (02) : 345 - 350
  • [9] Incidence and Factors Associated with Squeaking in Alumina-on-Alumina THA
    Choi, Il-Yong
    Kim, Yee-Suk
    Hwang, Kyu-Tae
    Kim, Young-Ho
    [J]. CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2010, 468 (12) : 3234 - 3239
  • [10] FUNCTIONAL RESULTS OF HIP ARTHROPLASTY WITH ACRYLIC PROSTHESIS
    DAUBIGNE, RM
    POSTEL, M
    [J]. JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1954, 36-A (03) : 451 - 475