Ray-tracing analysis of intraocular lens power in situ

被引:29
作者
Olsen, Thomas [1 ]
Funding, Mikkel [1 ]
机构
[1] Aarhus Hosp NBG, Univ Eye Clin, Aarhus C, Denmark
关键词
PARTIAL COHERENCE INTERFEROMETRY; POSTERIOR CORNEAL SURFACE; BIOMETRY;
D O I
10.1016/j.jcrs.2011.10.035
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
PURPOSE: To describe a method for back-solving the power of an intraocular lens (IOL) in situ based on laser biometry and ray-tracing analysis of the pseudophakic eye. SETTING: University Eye Clinic, Aarhus Hospital, Aarhus, Denmark. DESIGN: Evaluation of diagnostic test or technology. METHODS.: This study comprised pseudophakic eyes with an IOL power ranging from -2.00 to +36.00 diopters (D). Preoperatively, the corneal radius was measured with conventional autokeratometry and the axial length (AL) with optical biometry. After surgery, the position of the IOL was recorded using laser interferometry. Based on the postoperative refraction and the biometric measurements, a ray-tracing analysis was performed back-solving for the power of the IOL in situ. The analysis was performed assuming pupil diameters from 0.0 to 8.0 mm with and without correction for the Stiles-Crawford effect. RESULTS: The study evaluated 767 pseudophakic eyes (583 patients). Assuming a 3.0 mm pupil, the mean prediction error between the labeled and the calculated IOL power (+/- 1 standard deviation [SD]) was -0.26 D +/- 0.65 (SD) (range -2.4 to + 1.8 D). The prediction error showed no bias with IOL power or with AL. The calculated IOL power depended on the assumed pupil size and the Stiles-Crawford effect. However, the latter had a modulatory effect on the prediction error for large pupil diameters (>5.0 mm) only. CONCLUSION: The optics of the pseudophakic eye can be accurately described using exact ray tracing and modern biometric techniques.
引用
收藏
页码:641 / 647
页数:7
相关论文
共 18 条
[11]  
International Organization for Standardization, 2006, 119792 ISO
[13]   Calibration of axial length measurements with the Zeiss IOLMaster [J].
Olsen, T ;
Thorwest, M .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2005, 31 (07) :1345-1350
[14]  
OLSEN T, 1993, ACTA OPHTHALMOL, V71, P85
[15]   ON THE CALCULATION OF POWER FROM CURVATURE OF THE CORNEA [J].
OLSEN, T .
BRITISH JOURNAL OF OPHTHALMOLOGY, 1986, 70 (02) :152-154
[16]   Calculation of intraocular lens power: a review [J].
Olsen, Thomas .
ACTA OPHTHALMOLOGICA SCANDINAVICA, 2007, 85 (05) :472-485
[17]   COMPARISON OF THE SRK/T FORMULA AND OTHER THEORETICAL AND REGRESSION FORMULAS [J].
SANDERS, DR ;
RETZLAFF, JA ;
KRAFF, MC ;
GIMBEL, HV ;
RAANAN, MG .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 1990, 16 (03) :341-346
[18]   The luminous efficiency of rays entering the eye pupil at different points. [J].
Stiles, WS ;
Crawford, BH .
PROCEEDINGS OF THE ROYAL SOCIETY OF LONDON SERIES B-CONTAINING PAPERS OF A BIOLOGICAL CHARACTER, 1933, 112 (778) :428-450