Proton radiography and proton computed tomography based on time-resolved dose measurements

被引:60
作者
Testa, Mauro [1 ]
Verburg, Joost M. [1 ,2 ]
Rose, Mark [3 ]
Min, Chul Hee [1 ]
Tang, Shikui [1 ]
Bentefour, El Hassane [1 ]
Paganetti, Harald [1 ]
Lu, Hsiao-Ming [1 ]
机构
[1] Harvard Univ, Massachusetts Gen Hosp, Sch Med, Dept Radiat Oncol, Boston, MA 02114 USA
[2] Eindhoven Univ Technol, Sch Med Phys & Engn, NL-5600 MB Eindhoven, Netherlands
[3] Sun Nucl Corp, Melbourne, FL 32940 USA
关键词
VIVO RANGE VERIFICATION; SPATIAL-RESOLUTION; BEAM RADIOGRAPHY; THERAPY; SYSTEM; RADIOTHERAPY; DENSITY; GEANT4;
D O I
10.1088/0031-9155/58/22/8215
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
We present a proof of principle study of proton radiography and proton computed tomography (pCT) based on time-resolved dose measurements. We used a prototype, two-dimensional, diode-array detector capable of fast dose rate measurements, to acquire proton radiographic images expressed directly in water equivalent path length (WEPL). The technique is based on the time dependence of the dose distribution delivered by a proton beam traversing a range modulator wheel in passive scattering proton therapy systems. The dose rate produced in the medium by such a system is periodic and has a unique pattern in time at each point along the beam path and thus encodes the WEPL. By measuring the time dose pattern at the point of interest, the WEPL to this point can be decoded. If one measures the time-dose patterns at points on a plane behind the patient for a beam with sufficient energy to penetrate the patient, the obtained 2D distribution of the WEPL forms an image. The technique requires only a 2D dosimeter array and it uses only the clinical beam for a fraction of second with negligible dose to patient. We first evaluated the accuracy of the technique in determining the WEPL for static phantoms aiming at beam range verification of the brain fields of medulloblastoma patients. Accurate beam ranges for these fields can significantly reduce the dose to the cranial skin of the patient and thus the risk of permanent alopecia. Second, we investigated the potential features of the technique for real-time imaging of a moving phantom. Real-time tumor tracking by proton radiography could provide more accurate validations of tumor motion models due to the more sensitive dependence of proton beam on tissue density compared to x-rays. Our radiographic technique is rapid (similar to 100 ms) and simultaneous over the whole field, it can image mobile tumors without the problem of interplay effect inherently challenging for methods based on pencil beams. Third, we present the reconstructed pCT images of a cylindrical phantom containing inserts of different materials. As for all conventional pCT systems, the method illustrated in this work produces tomographic images that are potentially more accurate than x-ray CT in providing maps of proton relative stopping power (RSP) in the patient without the need for converting x-ray Hounsfield units to proton RSP. All phantom tests produced reasonable results, given the currently limited spatial and time resolution of the prototype detector. The dose required to produce one radiographic image, with the current settings, is similar to 0.7 cGy. Finally, we discuss a series of techniques to improve the resolution and accuracy of radiographic and tomographic images for the future development of a full-scale detector.
引用
收藏
页码:8215 / 8233
页数:19
相关论文
共 50 条
  • [21] A maximum likelihood proton path formalism for application in proton computed tomography
    Schulte, R. W.
    Penfold, S. N.
    Tafas, J. T.
    Schubert, K. E.
    MEDICAL PHYSICS, 2008, 35 (11) : 4849 - 4856
  • [22] Comparative accuracy and resolution assessment of two prototype proton computed tomography scanners
    Dedes, George
    Drosten, Hubertus
    Goetz, Stefanie
    Dickmann, Jannis
    Sarosiek, Christina
    Pankuch, Mark
    Krah, Nils
    Rit, Simon
    Bashkirov, Vladimir
    Schulte, Reinhard W.
    Johnson, Robert P.
    Parodi, Katia
    DeJongh, Ethan
    Landry, Guillaume
    MEDICAL PHYSICS, 2022, 49 (07) : 4671 - 4681
  • [23] PROGRESS TOWARD A PROTON COMPUTED TOMOGRAPHY APPARATUS
    Menichelli, D.
    Valentini, S.
    Bruzzi, M.
    Bucciolini, M.
    Capineri, L.
    Marrazzo, L.
    Talamonti, C.
    Tesi, M.
    Brianzi, M.
    Civinini, C.
    Lo Presti, D.
    Candiano, G.
    Cirrone, G. A. P.
    Cuttone, G.
    Randazzo, N.
    Russo, M.
    Fucile, A.
    Sipala, V.
    PROCEEDINGS OF THE 13TH ITALIAN CONFERENCE ON SENSORS AND MICROSYSTEMS, 2009, : 271 - +
  • [24] A pencil beam approach to proton computed tomography
    Rescigno, Regina
    Bopp, Cecile
    Rousseau, Marc
    Brasse, David
    MEDICAL PHYSICS, 2015, 42 (11) : 6610 - 6624
  • [25] An empirical artifact correction for proton computed tomography
    Dickmann, Jannis
    Sarosiek, Christina
    Goetz, Stefanie
    Pankuch, Mark
    Coutrakon, George
    Johnson, Robert P.
    Schulte, Reinhard W.
    Parodi, Katia
    Landry, Guillaume
    Dedes, George
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2021, 86 : 57 - 65
  • [26] Electron density uncertainties in proton computed tomography
    Erdelyi, B.
    PHYSICS IN MEDICINE AND BIOLOGY, 2010, 55 (23) : 7121 - 7134
  • [27] Proton computed tomography images with algebraic reconstruction
    Bruzzi, M.
    Civinini, C.
    Scaringella, M.
    Bonanno, D.
    Brianzi, M.
    Carpinelli, M.
    Cirrone, G. A. P.
    Cuttone, G.
    Lo Presti, D.
    Maccioni, G.
    Pallotta, S.
    Randazzo, N.
    Romano, F.
    Sipala, V.
    Talamonti, C.
    Vanzi, E.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2017, 845 : 652 - 655
  • [28] Expected proton signal sizes in the PRaVDA Range Telescope for proton Computed Tomography
    Price, T.
    Esposito, M.
    Poludniowski, G.
    Taylor, J.
    Waltham, C.
    Parker, D. J.
    Green, S.
    Manolopoulos, S.
    Allinson, N. M.
    Anaxagoras, T.
    Evans, P.
    Nieto-Camero, J.
    JOURNAL OF INSTRUMENTATION, 2015, 10
  • [29] The Effect of Tissue Inhomogeneities on the Accuracy of Proton Path Reconstruction for Proton Computed Tomography
    Wong, Kent
    Erdelyi, Bela
    Schulte, Reinhard
    Bashkirov, Vladimir
    Coutrakon, George
    Sadrozinski, Hartmut
    Penfold, Scott
    Rosenfeld, Anatoly
    APPLICATION OF ACCELERATORS IN RESEARCH AND INDUSTRY, 2009, 1099 : 476 - +
  • [30] Development of a proton Computed Tomography detector system
    Naimuddin, Md.
    Coutrakon, G.
    Blazey, G.
    Boi, S.
    Dyshkant, A.
    Erdelyi, B.
    Hedin, D.
    Johnson, E.
    Krider, J.
    Rukalin, V.
    Uzunyan, S. A.
    Zutshi, V.
    Fordt, R.
    Sellberg, G.
    Rauch, J. E.
    Roman, M.
    Rubinov, P.
    Wilson, P.
    JOURNAL OF INSTRUMENTATION, 2016, 11