Proton radiography and proton computed tomography based on time-resolved dose measurements

被引:60
|
作者
Testa, Mauro [1 ]
Verburg, Joost M. [1 ,2 ]
Rose, Mark [3 ]
Min, Chul Hee [1 ]
Tang, Shikui [1 ]
Bentefour, El Hassane [1 ]
Paganetti, Harald [1 ]
Lu, Hsiao-Ming [1 ]
机构
[1] Harvard Univ, Massachusetts Gen Hosp, Sch Med, Dept Radiat Oncol, Boston, MA 02114 USA
[2] Eindhoven Univ Technol, Sch Med Phys & Engn, NL-5600 MB Eindhoven, Netherlands
[3] Sun Nucl Corp, Melbourne, FL 32940 USA
关键词
VIVO RANGE VERIFICATION; SPATIAL-RESOLUTION; BEAM RADIOGRAPHY; THERAPY; SYSTEM; RADIOTHERAPY; DENSITY; GEANT4;
D O I
10.1088/0031-9155/58/22/8215
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
We present a proof of principle study of proton radiography and proton computed tomography (pCT) based on time-resolved dose measurements. We used a prototype, two-dimensional, diode-array detector capable of fast dose rate measurements, to acquire proton radiographic images expressed directly in water equivalent path length (WEPL). The technique is based on the time dependence of the dose distribution delivered by a proton beam traversing a range modulator wheel in passive scattering proton therapy systems. The dose rate produced in the medium by such a system is periodic and has a unique pattern in time at each point along the beam path and thus encodes the WEPL. By measuring the time dose pattern at the point of interest, the WEPL to this point can be decoded. If one measures the time-dose patterns at points on a plane behind the patient for a beam with sufficient energy to penetrate the patient, the obtained 2D distribution of the WEPL forms an image. The technique requires only a 2D dosimeter array and it uses only the clinical beam for a fraction of second with negligible dose to patient. We first evaluated the accuracy of the technique in determining the WEPL for static phantoms aiming at beam range verification of the brain fields of medulloblastoma patients. Accurate beam ranges for these fields can significantly reduce the dose to the cranial skin of the patient and thus the risk of permanent alopecia. Second, we investigated the potential features of the technique for real-time imaging of a moving phantom. Real-time tumor tracking by proton radiography could provide more accurate validations of tumor motion models due to the more sensitive dependence of proton beam on tissue density compared to x-rays. Our radiographic technique is rapid (similar to 100 ms) and simultaneous over the whole field, it can image mobile tumors without the problem of interplay effect inherently challenging for methods based on pencil beams. Third, we present the reconstructed pCT images of a cylindrical phantom containing inserts of different materials. As for all conventional pCT systems, the method illustrated in this work produces tomographic images that are potentially more accurate than x-ray CT in providing maps of proton relative stopping power (RSP) in the patient without the need for converting x-ray Hounsfield units to proton RSP. All phantom tests produced reasonable results, given the currently limited spatial and time resolution of the prototype detector. The dose required to produce one radiographic image, with the current settings, is similar to 0.7 cGy. Finally, we discuss a series of techniques to improve the resolution and accuracy of radiographic and tomographic images for the future development of a full-scale detector.
引用
收藏
页码:8215 / 8233
页数:19
相关论文
共 50 条
  • [1] Improvement of single detector proton radiography by incorporating intensity of time-resolved dose rate functions
    Zhang, Rongxiao
    Jee, Kyung-Wook
    Cascio, Ethan
    Sharp, Gregory C.
    Flanz, Jacob B.
    Lu, Hsiao-Ming
    PHYSICS IN MEDICINE AND BIOLOGY, 2018, 63 (01)
  • [2] Concept of proton radiography using energy resolved dose measurement
    Bentefour, El H.
    Schnuerer, Roland
    Lu, Hsiao-Ming
    PHYSICS IN MEDICINE AND BIOLOGY, 2016, 61 (16) : N386 - N393
  • [3] Proton radiography and computed tomography with dynamic water range shifter
    Park, Seyjoon
    Jeong, Chiyoung
    Lee, Jiseoc
    Cho, Seungryong
    Shin, Jae-ik
    Jeong, Jong Hwi
    Kim, Haksoo
    Jo, Kwanghyun
    Kim, Yeon-Joo
    Park, Jeong-Hoon
    Shin, Dongho
    Lim, Young Kyung
    Lee, Se Byeong
    JOURNAL OF INSTRUMENTATION, 2017, 12
  • [4] Investigation of time-resolved proton radiography using x-ray flat-panel imaging system
    Jee, K-W
    Zhang, R.
    Bentefour, E. H.
    Doolan, P. J.
    Cascio, E.
    Sharp, G.
    Flanz, J.
    Lu, H-M
    PHYSICS IN MEDICINE AND BIOLOGY, 2017, 62 (05) : 1905 - 1919
  • [5] Proton radiography and tomography with application to proton therapy
    Poludniowski, G.
    Allinson, N. M.
    Evans, P. M.
    BRITISH JOURNAL OF RADIOLOGY, 2015, 88 (1053)
  • [6] Proton computed tomography
    Bucciantonio, Martina
    Sauli, Fabio
    MODERN PHYSICS LETTERS A, 2015, 30 (17)
  • [7] Higher order analysis of time-resolved proton radiographs
    Doolan, P. J.
    Bentefour, E. H.
    Testa, M.
    Cascio, E.
    Sharp, G.
    Royle, G.
    Lu, H-M
    BIOMEDICAL PHYSICS & ENGINEERING EXPRESS, 2019, 5 (05):
  • [8] Review of medical radiography and tomography with proton beams
    Johnson, Robert P.
    REPORTS ON PROGRESS IN PHYSICS, 2018, 81 (01)
  • [9] TOFpRad: a novel proton radiography prototype based on Time Of Flight measurements
    Anzalone, R.
    Battistoni, G.
    Ciarrocchi, E.
    Dong, Y.
    Francesconi, M.
    Galli, L.
    Goanta, A. M.
    Krah, N.
    Mereghetti, A.
    Muraro, S.
    Muscato, A.
    Pullia, M.
    Traini, G.
    Morrocchi, M.
    JOURNAL OF INSTRUMENTATION, 2025, 20 (03):
  • [10] A reconstruction approach for proton computed tomography by modeling the integral depth dose of the scanning proton pencil beam
    Chen, Xinyuan
    Medrano, Maria
    Sun, Baozhou
    Hao, Yao
    Reynoso, Francisco J.
    Darafsheh, Arash
    Yang, Deshan
    Zhang, Tiezhi
    Zhao, Tianyu
    MEDICAL PHYSICS, 2022, 49 (04) : 2602 - 2620