Determination of local micromotion at the stem-neck taper junction of a bi-modular total hip prosthesis design

被引:14
作者
Falkenberg, Adrian [1 ]
Drummen, Paul [1 ]
Morlock, Michael M. [1 ]
Huber, Gerd [1 ]
机构
[1] Hamburg Univ Technol TUHH, Inst Biomech, Denickestr 15, D-21073 Hamburg, Germany
关键词
Bi-modular prosthesis; Taper junction; Micromotion; Assembly force; Flexural rigidity; FRETTING-CORROSION; METAL; RETRIEVAL; ARTHROPLASTY; INTERFACE; BEHAVIOR; FAILURE; CREVICE; STRESS; SYSTEM;
D O I
10.1016/j.medengphy.2019.01.003
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
High rates of clinical complications with bi-modular hip prostheses are attributed to failure of the stem neck taper junction. Taper wear analyses have shown extensive material loss as a result of corrosion, potentially initiated by micromotion. The purpose of the study was to determine the amount of micro motion at this junction for different loading, assembly and material conditions. Micromotion between the neck adapter (CoCr29Mo6-alloy) and the stem (TiMo12Zr6Fe2-alloy; both Rejuvenate, Stryker) within the taper junction of a bi-modular hip stem were determined by image matching analysis of consecutively recorded images through windows in the stem component. A finite element model was used to determine the micromotion in the taper regions outside the windows and validated with the measured micromotion. With the model, the influence of the load amplitude, assembly force and component materials were then investigated. Determined micromotion (14-79 mu m) by far exceeded critical values (5 mu m) associated with the onset of fretting corrosion. Increasing assembly forces achieved a significant reduction in micromotion. The numerical model revealed insufficient assembly to cause the neck to perform rocking motions under load, repetitively changing taper contact in combination with gap opening, which facilitates fluid ingress into the junction. Changing the stem material to a stiffer Ti-alloy achieved a reduction of the micromotion of about 30%. This study emphasises the high importance of material selection, assembly force and loading on the susceptibility of bi-modular hip stems to fretting and crevice corrosion. These findings can serve to explain the increased rate of clinically reported problems with this particular prosthesis design. (C) 2019 IPEM. Published by Elsevier Ltd. All rights reserved.
引用
收藏
页码:31 / 38
页数:8
相关论文
共 30 条
  • [1] Gross Trunnion Failure After Primary Total Hip Arthroplasty
    Banerjee, Samik
    Cherian, Jeffrey J.
    Bono, James V.
    Kurtz, Steven M.
    Geesink, Rudolph
    Meneghini, R. Michael
    Delanois, Ronald E.
    Mont, Michael A.
    [J]. JOURNAL OF ARTHROPLASTY, 2015, 30 (04) : 641 - 648
  • [2] Hip contact forces and gait patterns from routine activities
    Bergmann, G
    Deuretzbacher, G
    Heller, M
    Graichen, F
    Rohlmann, A
    Strauss, J
    Duda, GN
    [J]. JOURNAL OF BIOMECHANICS, 2001, 34 (07) : 859 - 871
  • [3] Eighty-six Percent Failure Rate of a Modular-Neck Femoral Stem Design at 3 to 5 Years Lessons Learned
    Bernstein, Derek T.
    Meftah, Morteza
    Paranilam, Jaya
    Incavo, Stephen J.
    [J]. JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2016, 98 (12) : e49
  • [4] Quantification of material loss from the neck piece taper junctions of a bimodular primary hip prosthesis. A retrieval study from 27 failed Rejuvenate bimodular hip arthroplasties
    Buente, D.
    Huber, G.
    Bishop, N.
    Morlock, M.
    [J]. BONE & JOINT JOURNAL, 2015, 97B (10) : 1350 - 1357
  • [5] The taper corrosion pattern observed for one bi-modular stem design is related to geometry-determined taper mechanics
    Buente, Dennis
    Bryant, Michael
    Ward, Michael
    Neville, Anne
    Morlock, Michael
    Huber, Gerd
    [J]. MEDICAL ENGINEERING & PHYSICS, 2017, 46 : 79 - 88
  • [6] Pseudotumour Formation Due to Tribocorrosion at the Taper Interface of Large Diameter Metal on Polymer Modular Total Hip Replacements
    Cook, Richard B.
    Bolland, Benjamin J. R. F.
    Wharton, Julian A.
    Tilley, Simon
    Latham, Jeremy M.
    Wood, Robert J. K.
    [J]. JOURNAL OF ARTHROPLASTY, 2013, 28 (08) : 1430 - 1436
  • [7] Friction in Total Hip Joint Prosthesis Measured In Vivo during Walking
    Damm, Philipp
    Dymke, Joern
    Ackermann, Robert
    Bender, Alwina
    Graichen, Friedmar
    Halder, Andreas
    Beier, Alexander
    Bergmann, Georg
    [J]. PLOS ONE, 2013, 8 (11):
  • [8] Corrosion and Fretting of a Modular Hip System: A Retrieval Analysis of 60 Rejuvenate Stems
    De Martino, Ivan
    Assini, Joseph B.
    Elpers, Marcella E.
    Wright, Timothy M.
    Westrich, Geoffrey H.
    [J]. JOURNAL OF ARTHROPLASTY, 2015, 30 (08) : 1470 - 1475
  • [9] Trunnionosis: Does Head Size Affect Fretting and Corrosion in Total Hip Arthroplasty?
    Del Balso, Christopher
    Teeter, Matthew G.
    Tan, Sok Chuen
    Howard, James L.
    Lanting, Brent A.
    [J]. JOURNAL OF ARTHROPLASTY, 2016, 31 (10) : 2332 - 2336
  • [10] Will New Metal Heads Restore Mechanical Integrity of Corroded Trunnions?
    Derasari, Aditya
    Gold, Jonathan E.
    Ismaily, Sabir
    Noble, Philip C.
    Incavo, Stephen J.
    [J]. JOURNAL OF ARTHROPLASTY, 2017, 32 (04) : 1356 - 1359