Design, optimization and numerical simulation of a MicroFlow sensor in the realistic model of human aorta

被引:6
作者
Mojarab, Amin [1 ]
Kamali, Reza [1 ]
机构
[1] Shiraz Univ, Sch Mech Engn, Shiraz, Iran
关键词
Atherosclerosis; Aneurysm; Microsensor; Aorta; CT Angiography; COMPUTATIONAL FLUID-DYNAMICS; WALL SHEAR-STRESS; BLOOD-FLOW; ABDOMINAL-AORTA; PULSATILE FLOW; ARTERIAL; ANEURYSM; SYSTEMS;
D O I
10.1016/j.flowmeasinst.2020.101791
中图分类号
TH [机械、仪表工业];
学科分类号
0802 ;
摘要
Among all arterial diseases, aneurysm and atherosclerosis are of great importance. In these diseases the cross-section area of the artery and therefore the blood flow velocity changes. Therefore, it is a good idea to use a micro flow sensor for measuring the blood flow velocity to diagnosing these diseases. In this study, design, geometric optimization and numerical simulation of a hotfilm microsensor in the realistic model of human's aorta are investigated. A normal aorta geometry is extracted from CT Angiography images, and after applying oscillating boundary conditions on the inlet and outlet arteries, blood flow parameters are investigated by computational fluid dynamics (CFD) simulation. After designing the microsensor, the effects of its location in different aorta regions on the blood flow characteristics are numerically investigated. The results show that in the presence of microsensor, streamlines patterns almost remain unchanged while the maximum blood flow velocity in the aorta cross section where the microsensor is located, increases up to 10%. It is also found that the secondary flow weakens when the microsensor enters the artery causing a reduction in velocity measurement error. Furthermore, because of the presence of a catheter, the pressure drop increases up to 768 Pa. Results show that less than 30% of the arterial cross-sectional area where the microsensor is located, senses an increase in temperature.
引用
收藏
页数:7
相关论文
共 32 条
[1]   Dynamical flow measurements in hydraulic systems using a polymer-based micro flow sensor [J].
Ahrens, Ralf ;
Festa, Manfred .
PROCEEDINGS OF THE EUROSENSORS XXIII CONFERENCE, 2009, 1 (01) :927-+
[2]   Numerical simulation of blood flow in a flexible stenosed abdominal real aorta [J].
Alishahi, M. ;
Alishahi, M. M. ;
Emdad, H. .
SCIENTIA IRANICA, 2011, 18 (06) :1297-1305
[3]  
[Anonymous], 2007, P 35 INT MAT C
[4]   Simulation of blood flow in human aorta with emphasis on outlet boundary conditions [J].
Benim, A. C. ;
Nahavandi, A. ;
Assmann, A. ;
Schubert, D. ;
Feindt, P. ;
Suh, S. H. .
APPLIED MATHEMATICAL MODELLING, 2011, 35 (07) :3175-3188
[5]  
Brida S., 1999, P 3 IT C SENS MICR R
[6]   A NUMERICAL STUDY OF FLOW IN CURVED TUBES SIMULATING CORONARY-ARTERIES [J].
CHANG, LJ ;
TARBELL, JM .
JOURNAL OF BIOMECHANICS, 1988, 21 (11) :927-937
[7]   A longitudinal study of Type-B aortic dissection and endovascular repair scenarios: Computational analyses [J].
Chen, Duanduan ;
Mueller-Eschner, Matthias ;
Kotelis, Drosos ;
Boeckler, Dittmar ;
Ventikos, Yiannis ;
von Tengg-Kobligk, Hendrik .
MEDICAL ENGINEERING & PHYSICS, 2013, 35 (09) :1321-1330
[8]  
CHO YI, 1991, BIORHEOLOGY, V28, P241
[9]   Computational fluid dynamics: Hemodynamic changes in abdominal aortic aneurysm after stent-graft implantation [J].
Frauenfelder, Thomas ;
Lotfey, Mourad ;
Boehm, Thomas ;
Wildermuth, Simon .
CARDIOVASCULAR AND INTERVENTIONAL RADIOLOGY, 2006, 29 (04) :613-623
[10]   Design considerations for silicon sensors for use in catheters and guide wires [J].
Goosen, JFL .
SMART MATERIALS & STRUCTURES, 2002, 11 (05) :804-812