Fiber-optic dosimeters for radiation therapy

被引:0
作者
Li, Enbang [1 ]
Archer, James [1 ]
机构
[1] Univ Wollongong, Ctr Med Radiat Phys, Wollongong, NSW 2522, Australia
来源
AOPC 2017: FIBER OPTIC SENSING AND OPTICAL COMMUNICATIONS | 2017年 / 10464卷
关键词
Fiber-optic dosimetry; Radiation detection; Scintillators; Cancer therapy; VIVO NEUROPHARMACOLOGY INVESTIGATIONS; INTRACEREBRAL RADIOSENSITIVE PROBE; ORGANIC SCINTILLATOR DOSIMETERS; SMALL LABORATORY-ANIMALS; ENERGY BEAM DOSIMETRY; MICROBEAM IRRADIATION; BRAIN-TUMOR; LIGHT; QUANTIFICATION; DETECTORS;
D O I
10.1117/12.2285522
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
According to the figures provided by the World Health Organization, cancer is a leading cause of death worldwide, accounting for 8.8 million deaths in 2015. Radiation therapy, which uses x-rays to destroy or injure cancer cells, has become one of the most important modalities to treat the primary cancer or advanced cancer. The newly developed microbeam radiation therapy (MRT), which uses highly collimated, quasi-parallel arrays of x-ray microbeams (typically 50 mu m wide and separated by 400 mu m) produced by synchrotron sources, represents a new paradigm in radiotherapy and has shown great promise in pre-clinical studies on different animal models. Measurements of the absorbed dose distribution of microbeams are vitally important for clinical acceptance of MRT and for developing quality assurance systems for MRT, hence are a challenging and important task for radiation dosimetry. On the other hand, during the traditional LINAC based radiotherapy and breast cancer brachytherapy, skin dose measurements and treatment planning also require a high spatial resolution, tissue equivalent, on-line dosimeter that is both economical and highly reliable. Such a dosimeter currently does not exist and remains a challenge in the development of radiation dosimetry. High resolution, water equivalent, optical and passive x-ray dosimeters have been developed and constructed by using plastic scintillators and optical fibers. The dosimeters have peak edge-on spatial resolutions ranging from 50 to 500 microns in one dimension, with a 10 micron resolution dosimeter under development. The developed fiber-optic dosimeters have been test with both LINAC and synchrotron x-ray beams. This work demonstrates that water-equivalent and high spatial resolution radiation detection can be achieved with scintillators and optical fiber systems. Among other advantages, the developed fiber-optic probes are also passive, energy independent, and radiation hard.
引用
收藏
页数:10
相关论文
共 51 条
  • [1] Measuring directionality in double-beta decay and neutrino interactions with kiloton-scale scintillation detectors
    Aberle, C.
    Elagin, A.
    Frisch, H. J.
    Wetstein, M.
    Winslow, L.
    [J]. JOURNAL OF INSTRUMENTATION, 2014, 9
  • [2] [Anonymous], 1937, Dokl. Akad. Nauk SSSR
  • [3] [Anonymous], 1964, THEORY PRACTICE SCIN
  • [4] A mathematical formalism for hyperspectral, multipoint plastic scintillation detectors
    Archambault, Louis
    Therriault-Proulx, Francois
    Beddar, Sam
    Beaulieu, Luc
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2012, 57 (21) : 7133 - 7145
  • [5] Archer J., 2017, SCI REPORTS UNPUB
  • [6] High-resolution fiber-optic dosimeters for microbeam radiation therapy
    Archer, James
    Li, Enbang
    Petasecca, Marco
    Lerch, Michael
    Rosenfeld, Anatoly
    Carolan, Martin
    [J]. MEDICAL PHYSICS, 2017, 44 (05) : 1965 - 1968
  • [7] Micrometer-resolved film dosimetry using a microscope in microbeam radiation therapy
    Bartzsch, Stefan
    Lott, Johanna
    Welsch, Katrin
    Braeuer-Krisch, Elke
    Oelfke, Uwe
    [J]. MEDICAL PHYSICS, 2015, 42 (07) : 4069 - 4079
  • [8] Review of plastic and liquid scintillation dosimetry for photon, electron, and proton therapy
    Beaulieu, Luc
    Beddar, Sam
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2016, 61 (20) : R305 - R343
  • [9] WATER-EQUIVALENT PLASTIC SCINTILLATION DETECTORS FOR HIGH-ENERGY BEAM DOSIMETRY .2. PROPERTIES AND MEASUREMENTS
    BEDDAR, AS
    MACKIE, TR
    ATTIX, FH
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1992, 37 (10) : 1901 - 1913
  • [10] WATER-EQUIVALENT PLASTIC SCINTILLATION DETECTORS FOR HIGH-ENERGY BEAM DOSIMETRY .1. PHYSICAL CHARACTERISTICS AND THEORETICAL CONSIDERATIONS
    BEDDAR, AS
    MACKIE, TR
    ATTIX, FH
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1992, 37 (10) : 1883 - 1900