Longitudinal miniaturization of fiber-packed capillary column in high temperature gas chromatography

被引:25
作者
Ogawa, M.
Saito, Y.
Imaizumi, M.
Wada, H.
Jinno, K. [1 ]
机构
[1] Toyohashi Univ Technol, Sch Mat Sci, Toyohashi, Aichi 4418580, Japan
[2] Shinwa Chem Ind Ltd, Fushimi Ku, Kyoto 6128307, Japan
基金
日本学术振兴会;
关键词
gas chromatography; miniaturization; fiber-packed capillary columns;
D O I
10.1365/s10337-006-0771-4
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Miniaturization of fiber-packed capillary column in gas chromatography was studied. Packed with a bundle of polymer-coated filaments into a short metal capillary, longitudinal miniaturization was demonstrated without any significant disadvantages for practical retentivity and sample loading capacity. With a metal capillary of 0.3 mm i.d., the typical column length could be reduced to be several centimeters. The retentivity for alkanes on the fiber-packed capillary column was compared to that obtained by a conventional open-tubular capillary column of the same length. Taking advantage of the heat-resistant property of both the fiber and the metal capillary, the separation of a polystyrene mixture on the polymer-coated fiber-packed short metal capillary column with a temperature-programmed run up to 400 degrees C was successfully carried out. As the typical application, rapid GC separations were also carried out with the short column of 0.05 m length.
引用
收藏
页码:459 / 463
页数:5
相关论文
共 36 条
[1]   Surface derivatization of poly(p-phenylene terephthalamide) fiber designed for novel separation and extraction media [J].
Abe, A ;
Saito, Y ;
Imaizumi, M ;
Ogawa, M ;
Takeichi, T ;
Jinno, K .
JOURNAL OF SEPARATION SCIENCE, 2005, 28 (17) :2413-2418
[2]   High-speed gas chromatography with direct resistively-heated column (ultra fast module-GC) separation measure (S) and other chromatographic parameters under different analysis conditions for samples of different complexities and volatilities [J].
Bicchi, C ;
Brunelli, C ;
Cordero, C ;
Rubiolo, P ;
Galli, M ;
Sironi, A .
JOURNAL OF CHROMATOGRAPHY A, 2005, 1071 (1-2) :3-12
[3]   Conventional inner diameter short capillary columns: an approach to speeding up gas chromatographic analysis of medium complexity samples [J].
Bicchi, C ;
Brunelli, C ;
Galli, M ;
Sironi, A .
JOURNAL OF CHROMATOGRAPHY A, 2001, 931 (1-2) :129-140
[4]   High-speed gas chromatography: an overview of various concepts [J].
Cramers, CA ;
Janssen, HG ;
van Deursen, MM ;
Leclercq, PA .
JOURNAL OF CHROMATOGRAPHY A, 1999, 856 (1-2) :315-329
[5]   Gas chromatography [J].
Eiceman, GA ;
Gardea-Torresdey, J ;
Overton, E ;
Carney, K ;
Dorman, F .
ANALYTICAL CHEMISTRY, 2004, 76 (12) :3387-3394
[6]   In-valve sample preparation cartridge designed for microcolumn liquid chromatography [J].
Imaizumi, M ;
Saito, Y ;
Ban, K ;
Wada, H ;
Hayashida, M ;
Jinno, K .
CHROMATOGRAPHIA, 2004, 60 (11-12) :619-623
[7]   Polymer-coated fibrous extraction medium for sample preparation coupled to microcolumn liquid-phase separations [J].
Imaizumi, M ;
Saito, Y ;
Hayashida, M ;
Takeichi, T ;
Wada, H ;
Jinno, K .
JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2003, 30 (06) :1801-1808
[8]  
Jinno K, 2001, ELECTROPHORESIS, V22, P3785, DOI 10.1002/1522-2683(200109)22:17<3785::AID-ELPS3785>3.0.CO
[9]  
2-U
[10]  
Jinno K, 2001, ELECTROPHORESIS, V22, P3371, DOI 10.1002/1522-2683(200109)22:16<3371::AID-ELPS3371>3.3.CO