An Experimental Study to Evaluate the Fluobeam 800 Imaging System for Fluorescence-Guided Lymphatic Imaging and Sentinel Node Biopsy

被引:45
作者
Hirche, Christoph [1 ,2 ]
Engel, Holger [1 ,2 ]
Kolios, Leila [1 ,2 ]
Cognie, Juliette [3 ]
Huenerbein, Michael [4 ]
Lehnhardt, Marcus [1 ,2 ]
Kremer, Thomas [1 ,2 ]
机构
[1] Trauma Ctr Ludwigshafen, D-67071 Ludwigshafen, Germany
[2] Heidelberg Univ, Heidelberg, Germany
[3] Univ Tours, Tours, France
[4] Helios Hosp Berlin Buch, Berlin, Germany
关键词
lymphatic imaging; fluorescence; lymph node; solid tumors; sentinel lymph node biopsy; indocyanine green; ICG; lymphedema; INDOCYANINE GREEN FLUORESCENCE; BREAST-CANCER PATIENTS; EXTREMITY LYMPHEDEMA; GASTRIC-CANCER; SURGERY; IDENTIFICATION; LYMPHOGRAPHY; DYE; LYMPHOSCINTIGRAPHY; ANASTOMOSIS;
D O I
10.1177/1553350612468962
中图分类号
R61 [外科手术学];
学科分类号
摘要
Background. Visualization of the lymphatic system is challenging. Lymphatic imaging is a crucial diagnostic tool for benign and malignant lymphatic pathologies. Fluorescence-guided imaging allows selective lymphatic mapping and sentinel lymph node (SLN) identification. There are a few fluorescence systems, but some drawbacks remain due to technical and ergonomic aspects. The aim of this study was to evaluate the feasibility of the new Fluobeam 800 imaging system. Methods. After approval by the ethics committee, the system was evaluated for lymphography and SLN biopsy in an animal model. Five pigs each with 4 lymph node (LN) stations (n = 20 LN stations) were subjected to lymphatic imaging using indocyanine green (ICG). Additionally, the use of ICG was compared with ICG adsorbed to human serum albumin (ICG-HSA). Lymphatic vessels and SLN identification rates were measured. Results. After injection, a clear fluorescence signal of the lymphatic vessels was visualized leading to the LN station. Overall, ICG fluorescence imaging identified a mean of 2.0 lymphatic vessels and 1.1 (range = 1-2) SLN in 20 of 20 LN stations. Reverse lymphography was feasible. A clinical difference in resolution was not detected between use of ICG-HSA and ICG. Conclusion. This is the first study analyzing the feasibility of the Fluobeam 800 imaging system allowing transcutaneous real-time imaging. It enables detection of the SLN by fluorescence retention with increased detection depth and resolution. After fixation to the ceiling, the ergonomics advanced for simultaneous field navigation and dissection. The new system can be applied for lymphatic imaging for lympatico-reconstructive surgery and SLN biopsy.
引用
收藏
页码:516 / 523
页数:8
相关论文
共 40 条
[11]   High rate of solitary sentinel node metastases identification by fluorescence-guided lymphatic imaging in breast cancer [J].
Hirche, Christoph ;
Mohr, Zarah ;
Kneif, Soeren ;
Murawa, Dawid ;
Huenerbein, Michael .
JOURNAL OF SURGICAL ONCOLOGY, 2012, 105 (02) :162-166
[12]   ICG fluorescence-guided sentinel node biopsy for axillary nodal staging in breast cancer [J].
Hirche, Christoph ;
Murawa, Dawid ;
Mohr, Zarah ;
Kneif, Soeren ;
Huenerbein, Michael .
BREAST CANCER RESEARCH AND TREATMENT, 2010, 121 (02) :373-378
[13]   Optimization of near-infrared fluorescent sentinel lymph node mapping for vulvar cancer [J].
Hutteman, Merlijn ;
van der Vorst, Joost R. ;
Gaarenstroom, Katja N. ;
Peters, Alexander A. W. ;
Mieog, J. Sven D. ;
Schaafsma, Boudewijn E. ;
Lowik, Clemens W. G. M. ;
Frangioni, John V. ;
van de Velde, Cornelis J. H. ;
Vahrmeijer, Alexander L. .
AMERICAN JOURNAL OF OBSTETRICS AND GYNECOLOGY, 2012, 206 (01) :89.e1-89.e5
[14]   Randomized, double-blind comparison of indocyanine green with or without albumin premixing for near-infrared fluorescence imaging of sentinel lymph nodes in breast cancer patients [J].
Hutteman, Merlijn ;
Mieog, J. Sven D. ;
van der Vorst, Joost R. ;
Liefers, Gerrit Jan ;
Putter, Hein ;
Lowik, Clemens W. G. M. ;
Frangioni, John V. ;
van de Velde, Cornelis J. H. ;
Vahrmeijer, Alexander L. .
BREAST CANCER RESEARCH AND TREATMENT, 2011, 127 (01) :163-170
[15]   Clinical Translation of Ex Vivo Sentinel Lymph Node Mapping for Colorectal Cancer Using Invisible Near-Infrared Fluorescence Light [J].
Hutteman, Merlijn ;
Choi, Hak Soo ;
Mieog, J. Sven D. ;
van der Vorst, Joost R. ;
Ashitate, Yoshitomo ;
Kuppen, Peter J. K. ;
van Groningen, Marian C. ;
Lowik, Clemens W. G. M. ;
Smit, Vincent T. H. B. M. ;
van de Velde, Cornelis J. H. ;
Frangioni, John V. ;
Vahrmeijer, Alexander L. .
ANNALS OF SURGICAL ONCOLOGY, 2011, 18 (04) :1006-1014
[16]   Sentinel node navigation surgery using indocyanine green in patients with lung cancer [J].
Ito, N ;
Fukuta, M ;
Tokushima, T ;
Nakai, K ;
Ohgi, S .
SURGERY TODAY, 2004, 34 (07) :581-585
[17]   Sentinel node mapping with indocyanine green (ICG) and infrared ray detection in early gastric cancer: An accurate method that enables a limited lymphadenectomy [J].
Kelder, W. ;
Nimura, H. ;
Takahashi, N. ;
Mitsumori, N. ;
van Dam, G. M. ;
Yanaga, K. .
EJSO, 2010, 36 (06) :552-558
[18]   The sentinel node procedure in colon carcinoma: a multi-centre study in The Netherlands [J].
Kelder, Wendy ;
Braat, Andries E. ;
Karrenbeld, Arend ;
Grond, Joris A. K. ;
De Vries, Johannes E. ;
Oosterhuis, J. Wolter A. ;
Baas, Peter C. ;
Plukker, John T. M. .
INTERNATIONAL JOURNAL OF COLORECTAL DISEASE, 2007, 22 (12) :1509-1514
[19]   Fluorescence navigation with indocyanine green for detecting sentinel lymph nodes in breast cancer [J].
Kitai T. ;
Inomoto T. ;
Miwa M. ;
Shikayama T. .
Breast Cancer, 2005, 12 (3) :211-215
[20]   Intraoperative Near-infrared Fluorescence Imaging in Perforator Flap Reconstruction: Current Research and Early Clinical Experience [J].
Lee, Bernard T. ;
Matsui, Aya ;
Hutteman, Merlijn ;
Lin, Samuel J. ;
Winer, Joshua H. ;
Laurence, Rita G. ;
Frangioni, John V. .
JOURNAL OF RECONSTRUCTIVE MICROSURGERY, 2010, 26 (01) :59-65