Comparative study on contraceptive efficacy and clinical performance of the copper/low-density polyethylene nanocomposite IUD and the copper T220C IUD

被引:25
作者
Yu, Jing [1 ]
Li, Jun [1 ]
Li, Hong-gang [1 ]
Li, Jian-xiong [1 ]
Xie, Chang-sheng [2 ]
Zhu, Chang-hong [1 ]
机构
[1] Huazhong Univ Sci & Technol, Tongji Med Coll, Family Planning Res Inst, Wuhan 430030, Peoples R China
[2] Huazhong Univ Sci & Technol, Dept Mat Sci & Engn, Wuhan 430074, Peoples R China
关键词
copper/low-density polyethylene nanocomposite IUD; TCu220C IUD; antifertility effectiveness; side effects;
D O I
10.1016/j.contraception.2008.05.008
中图分类号
R71 [妇产科学];
学科分类号
100211 ;
摘要
Background: The study was conducted to compare the antifertility effectiveness and side effects of the copper/low-densily polyethylene nanocomposite IUD (experimental group) and the copper T220C IUD (control group). Study Design: One hundred females were randomly divided into two groups (experimental group and control group. n=50 in each group). Clinical observation and comparative study were performed oil the two groups for 12 months. Results: Follow-up rate was 100% at the 12th month. In the experimental group and control group, the cumulative continuation rates were both 92.0 per 100 women at the 12th month and there was no difference between them (p>.05). The pregnancy rate, removal rate and expulsion rate were low with the difference being not statistically significant (p>.05). The most common side effects were excessive menstrual bleeding. spotting and pain. The rates of side effects were lower ill the experimental group than ill control group, especially during the initial 3 months after insertion with the differences being statistically significant (p<.05). Conclusion: The new design of the copper/low-density polyethylene nanocomposite IUD showed low pregnancy rate, high contraceptive efficacy and satisfactory acceptability. The study suggested that the TCu220C IUD also had high contraceptive efficacy, but had relatively more side effects. Crown Copyright (C) 2008 Published by Elsevier Inc. All rights reserved.
引用
收藏
页码:319 / 323
页数:5
相关论文
共 25 条
[1]   Copper corrosion-simulated uterine solutions [J].
Bastidas, JM ;
Cano, E ;
Mora, N .
CONTRACEPTION, 2000, 61 (06) :395-399
[2]   Characterization of copper corrosion products originated in simulated uterine fluids and on packaged intrauterine devices [J].
Bastidas, JM ;
Mora, N ;
Cano, E ;
Polo, JL .
JOURNAL OF MATERIALS SCIENCE-MATERIALS IN MEDICINE, 2001, 12 (05) :391-397
[3]   Research on Cu2+ transformations of Cu and its oxides particles with different sizes in the simulated uterine solution [J].
Cai, SZ ;
Xia, XP ;
Xie, CS .
CORROSION SCIENCE, 2005, 47 (04) :1039-1047
[4]   Corrosion behavior of copper/LDPE nanocomposites in simulated uterine solution [J].
Cai, SZ ;
Xia, XP ;
Xie, CS .
BIOMATERIALS, 2005, 26 (15) :2671-2676
[5]  
[狄海红 Haihong DI], 2006, [生殖与避孕, Reproduction and Contraception], V26, P584
[6]   The evaluation of 318 intrauterine pregnancy cases with an intrauterine device [J].
Inal, MM ;
Ertopçu, K ;
Özelmas, I .
EUROPEAN JOURNAL OF CONTRACEPTION AND REPRODUCTIVE HEALTH CARE, 2005, 10 (04) :266-271
[7]  
KULIER R, 2007, COCHRANE DB SYST REV, V17
[8]  
LAI CH, 2006, CHIN J FAM PLANN, V2, P98
[9]   The antifertility effectiveness of copper/low-density polyethylene nanocomposite and its influence on the endometrial environment in rats [J].
Liu, He-fang ;
Liu, Zi-long ;
Xie, Chang-sheng ;
Yu, Jing ;
Zhu, Chang-hong .
CONTRACEPTION, 2007, 75 (02) :157-161
[10]   Intrauterine devices: Mechanisms of action, safety, and efficacy [J].
Mishell, DR .
CONTRACEPTION, 1998, 58 (03) :45S-53S