In vivo comparison of key quantitative parameters measured with 3D peripheral angiography, 2D peripheral quantitative angiography and intravascular ultrasound

被引:4
作者
Zasada, Wojciech [1 ,2 ]
Slezak, Magdalena [2 ]
Pociask, Elzbieta [2 ]
Malinowski, Krzysztof Piotr [2 ,3 ]
Proniewska, Klaudia [2 ,7 ]
Buszman, Piotr [4 ,5 ]
Milewski, Krzysztof [4 ,8 ]
Granada, Juan F. [6 ]
Kaluza, Grzegorz L. [6 ]
机构
[1] Univ Hosp Krakow, Dept Cardiol 2, Krakow, Poland
[2] KCRI, Miechowska 5b, PL-30055 Krakow, Poland
[3] Jagiellonian Univ, Inst Publ Hlth, Fac Hlth Sci, Med Coll, Krakow, Poland
[4] Ctr Cardiovasc Res & Dev Amer Heart Poland, Kostkowice, Poland
[5] Andrzej Frycz Modrzewski Krakow Univ, Krakow, Poland
[6] CRF, Skirball Ctr Innovat, New York, NY USA
[7] Jagiellonian Univ, Dept Bioinformat & Telemed, Med Coll, Krakow, Poland
[8] Jerzy Kukuczka Acad Phys Educ, Katowice, Poland
关键词
QCA; IVUS; IAFimaging; angiographic; fluoroscopic; Peripheral angiography; CORONARY-ANGIOGRAPHY; ELUTING STENTS; RECONSTRUCTION; VALIDATION; LENGTHS; LESIONS; IMPACT;
D O I
10.1007/s10554-019-01529-5
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
The aim of this study was to compare the measures of luminal stenosis between the two-dimensional (2D) and three-dimensional (3D) Quantitative Vessel Analysis (QVA) generated by CAAS QVA software and intravascular ultrasound (IVUS). Invasive contrast angiography is considered gold standard for diagnostic imaging and intervention in both coronary and peripheral arterial disease. However, it is based on 2D images depicting complicated 3D arterial anatomy. To overcome these limitations, 3D QVA has been developed to bridge the gap between 2D QVA and endovascular imaging. Thirty porcine femoral angiograms (common, profunda and superficial) with matching intravascular ultrasound (IVUS) pullbacks featuring variable degree of stenosis were analysed by 2D QVA, 3D QVA and quantitative IVUS. All 3 modalities provided similar data regarding the length of the investigated segment. Median lumen diameter was nearly identical in IVUS (4.69mm) and in 3D QVA (4.76mm) but quite a bit lower in 2D QVA (4.47mm, Kruskal-Wallis test p=0.1648). Lumen area measured in 2D QVA was lower than in IVUS and in 3D QVA. Lumen areas rendered by IVUS and 3D QVA were similar. Bland-Altman plots showed that the lowest differences were observed between IVUS and 3D QVA. IVUS and 3D QVA results were consistently higher than 2D QVA. 3D QVA is a useful surrogate of IVUS for precise luminal morphology measurements of peripheral arteries, rendering results that are much closer to IVUS than 2D QVA can provide.
引用
收藏
页码:215 / 223
页数:9
相关论文
共 23 条
[1]   Evaluation of peripheral atherosclerosis: A comparative analysis of angiography and intravascular ultrasound imaging [J].
Arthurs, Zachary M. ;
Bishop, Paul D. ;
Feiten, Lindsay E. ;
Eagleton, Matthew J. ;
Clair, Daniel G. ;
Kashyap, Vikram S. .
JOURNAL OF VASCULAR SURGERY, 2010, 51 (04) :933-939
[2]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[3]   Impact of stent mis-sizing and mis-positioning on coronary fluid wall shear and intramural stress [J].
Chen, Henry Y. ;
Koo, Bon-Kwon ;
Bhatt, Deepak L. ;
Kassab, Ghassan S. .
JOURNAL OF APPLIED PHYSIOLOGY, 2013, 115 (02) :285-292
[4]   Three Dimensional Quantitative Coronary Angiography Can Detect Reliably Ischemic Coronary Lesions Based on Fractional Flow Reserve [J].
Chung, Woo-Young ;
Choi, Byoung-Joo ;
Lim, Seong-Hoon ;
Matsuo, Yoshiki ;
Lennon, Ryan J. ;
Gulati, Rajiv ;
Sandhu, Gurpreet S. ;
Holmes, David R., Jr. ;
Rihal, Charanjit S. ;
Lerman, Amir .
JOURNAL OF KOREAN MEDICAL SCIENCE, 2015, 30 (06) :716-724
[5]   Impact of stent deployment procedural factors on long-term effectiveness and safety of sirolimus-eluting stents (final results of the multicenter prospective STLLR trial) [J].
Costa, Marco A. ;
Angiolillo, Dominick J. ;
Tannenbaum, Mark ;
Driesman, Mitchell ;
Chu, Alan ;
Patterson, John ;
Kuehl, William ;
Battaglia, Joseph ;
Dabbons, Samir ;
Shamoon, Fayez ;
Flieshman, Bruce ;
Niederman, Alan ;
Bass, Theodore A. .
AMERICAN JOURNAL OF CARDIOLOGY, 2008, 101 (12) :1704-1711
[6]  
Gagne PJ, 2017, J VASC SURG-VENOUS L, V5, P678, DOI 10.1016/j.jvsv.2017.04.007
[7]   Validity and variability in visual assessment of stenosis severity in phantom bifurcation lesions: A survey in experts during the fifth meeting of the european bifurcation club [J].
Girasis, Chrysafios ;
Onuma, Yoshinobu ;
Schuurbiers, Johan C. H. ;
Morel, Marie-angele ;
van Es, Gerrit-Anne ;
van Geuns, Robert-Jan ;
Wentzel, Jolanda J. ;
Serruys, Patrick W. .
CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2012, 79 (03) :361-368
[8]   Angiographic views used for percutaneous coronary interventions: A three-dimensional analysis of physician-determined vs. computer-generated views [J].
Green, NE ;
Chen, SYJ ;
Hansgen, AR ;
Messenger, JC ;
Groves, BM ;
Carroll, JD .
CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2005, 64 (04) :451-459
[9]   Three-dimensional vascular angiography [J].
Green, NE ;
Chen, SYJ ;
Messenger, JC ;
Groves, BM ;
Carroll, JD .
CURRENT PROBLEMS IN CARDIOLOGY, 2004, 29 (03) :104-142
[10]   QUANTITATIVE AND QUALITATIVE CORONARY ANGIOGRAPHIC ANALYSIS - REVIEW OF METHODS, UTILITY, AND LIMITATIONS [J].
HERMILLER, JB ;
CUSMA, JT ;
SPERO, LA ;
FORTIN, DF ;
HARDING, MB ;
BASHORE, TM .
CATHETERIZATION AND CARDIOVASCULAR DIAGNOSIS, 1992, 25 (02) :110-131