Automated Quantitative Analysis of Anterior Segment Inflammation Using Swept-Source Anterior Segment Optical Coherence Tomography: A Pilot Study

被引:4
|
作者
Keino, Hiroshi [1 ]
Aman, Takuto [2 ]
Furuya, Ryota [3 ]
Nakayama, Makiko [1 ]
Okada, Annabelle A. [1 ]
Sunayama, Wataru [3 ]
Hatanaka, Yuji [2 ]
机构
[1] Kyorin Univ, Dept Ophthalmol, Sch Med, 6-20-2 Shinkawa, Tokyo 1818611, Japan
[2] Oita Univ, Fac Sci & Engn, 700 Dannoharu, Oita 8701192, Japan
[3] Univ Shiga Prefecture, Fac Engn, 2500 Hassaka, Hikone 5228533, Japan
关键词
uveitis; anterior chamber inflammation; optical coherence tomography (OCT); image analysis; Standardization of Uveitis Nomenclature (SUN); CHAMBER INFLAMMATION; DIAGNOSTIC-CRITERIA; UVEITIS; FREQUENCY; DISEASE; SYSTEM;
D O I
10.3390/diagnostics12112703
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: The aim of this study is to develop an automated evaluation of anterior chamber (AC) cells in uveitis using anterior segment (AS) optical coherence tomography (OCT) images. Methods: We analyzed AS swept-source (SS)-OCT (CASIA 2) images of 31 patients (51 eyes) with uveitis using image analysis software (Python). An automated algorithm was developed to detect cellular spots corresponding to hyper-reflective spots in the AC, and the correlation with Standardization of Uveitis Nomenclature (SUN) grading AC cells score was evaluated. The approximated AC grading value was calculated based on the logarithmic approximation curve between the number of cellular spots and the SUN grading score. Results: Among 51 eyes, cellular spots were automatically segmented in 48 eyes, whereas three eyes (all SUN grading AC cells score: 4+) with severe fibrin formation in the AC were removed by the automated algorithm. The AC cellular spots increased with an increasing SUN grading score (p < 0.001). The 48 eyes were split into training data (26 eyes) and test data (22 eyes). There was a significant correlation between the SUN grading score and the number of cellular spots in 26 eyes (rho: 0.843, p < 0.001). There was a significant correlation between the SUN grading score and the approximated grading value of 22 eyes based on the logarithmic approximation curve (rho: 0.774, p < 0.001). Leave-one-out cross-validation analysis demonstrated a significant correlation between the SUN grading score and the approximated grading value of 48 eyes (rho: 0.748, p < 0.001). Conclusions: This automated anterior AC cell analysis using AS SS-OCT showed a significant correlation with clinical SUN grading scores and provided SUN AC grading values as a continuous variable. Our findings suggest that automated grading of AC cells could improve the accuracy of a quantitative assessment of AC inflammation using AS-OCT images and allow the objective and rapid evaluation of anterior segment inflammation in uveitis. Further investigations on a large scale are required to validate this quantitative measurement of anterior segment inflammation in uveitic eyes.
引用
收藏
页数:11
相关论文
共 50 条
  • [1] Automated quantification of anterior chamber cells using swept-source anterior segment optical coherence tomography
    Pillar, Shani
    Kadomoto, Shin
    Chen, Keren
    Gonzalez, Saitiel Sandoval
    Cherian, Nina
    Privratsky, Joseph K.
    Zargari, Nicolette
    Jackson, Nicholas J.
    Corradetti, Giulia
    Chen, Judy L.
    Sadda, SriniVas R.
    Holland, Gary N.
    Tsui, Edmund
    JOURNAL OF OPHTHALMIC INFLAMMATION AND INFECTION, 2025, 15 (01):
  • [2] Anterior Vitreous Objective Assessment in Uveitis: An Anterior Segment Swept Source Optical Coherence Tomography Study
    Invernizzi, Alessandro
    Zaffalon, Chiara
    Manni, Priscilla
    Zicarelli, Federico
    Chisari, Domenico
    Adani, Chantal
    Mastrofilippo, Valentina
    Bolletta, Elena
    Gozzi, Fabrizio
    De Simone, Luca
    Staurenghi, Giovanni
    Cimino, Luca
    OCULAR IMMUNOLOGY AND INFLAMMATION, 2025, 33 (01) : 24 - 32
  • [3] Role of swept-source anterior segment optical coherence tomography in imaging pediatric cataract
    Bhardwaj, Anand
    Kaur, Savleen
    Sukhija, Jaspreet
    Gupta, Parul Chawla
    Ram, Jagat
    INDIAN JOURNAL OF OPHTHALMOLOGY, 2023, 71 (05) : 2132 - 2138
  • [4] Analysis of ocular inflammation in anterior chamber—involving uveitis using swept-source anterior segment OCT
    Elmira Baghdasaryan
    Tudor C. Tepelus
    Kenneth M. Marion
    Jianyan Huang
    Ping Huang
    SriniVas R. Sadda
    Olivia L. Lee
    International Ophthalmology, 2019, 39 : 1793 - 1801
  • [5] Screening for Subclinical Keratoconus Using Swept-Source Fourier Domain Anterior Segment Optical Coherence Tomography
    Steinberg, Johannes
    Casagrande, Maria K.
    Frings, Andreas
    Katz, Toam
    Druchkiv, Vasyl
    Richard, Gisbert
    Linke, Stephan J.
    CORNEA, 2015, 34 (11) : 1413 - 1419
  • [6] Three-Dimensional Morphometric Analysis of the Iris by Swept-Source Anterior Segment Optical Coherence Tomography in a Caucasian Population
    Invernizzi, Alessandro
    Giardini, Piero
    Cigada, Mario
    Viola, Francesco
    Staurenghi, Giovanni
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2015, 56 (08) : 4796 - 4801
  • [7] Analysis of ocular inflammation in anterior chamber-involving uveitis using swept-source anterior segment OCT
    Baghdasaryan, Elmira
    Tepelus, Tudor C.
    Marion, Kenneth M.
    Huang, Jianyan
    Huang, Ping
    Sadda, SriniVas R.
    Lee, Olivia L.
    INTERNATIONAL OPHTHALMOLOGY, 2019, 39 (08) : 1793 - 1801
  • [8] Comparison of a New Swept-Source Anterior Segment Optical Coherence Tomography and a Scheimpflug Camera for Measurement of Corneal Curvature
    Zhang, Ting
    Zhou, Yijing
    Young, Charlotte Aimee
    Chen, Aiming
    Jin, Guangming
    Zheng, Danying
    CORNEA, 2020, 39 (07) : 818 - 822
  • [9] Anterior segment optical coherence tomography in acute anterior uveitis
    Agra, Cristiana
    Agra, Lydianne
    Dantas, Jeanine
    Faria e Arantes, Tiago Eugenio
    de Andrade Neto, Joao Lins
    ARQUIVOS BRASILEIROS DE OFTALMOLOGIA, 2014, 77 (01) : 1 - 3
  • [10] Automated measurement of iris surface smoothness using anterior segment optical coherence tomography
    Zarei, Mohammad
    Mahmoudi, Tahereh
    Riazi-Esfahani, Hamid
    Mousavi, Behnam
    Ebrahimiadib, Nazanin
    Yaseri, Mehdi
    Pour, Elias Khalili
    Arabalibeik, Hossein
    SCIENTIFIC REPORTS, 2021, 11 (01)