Comparison of Two Spectral Domain Optical Coherence Tomography Devices for Angle-Closure Assessment

被引:28
作者
Quek, Desmond T. [2 ]
Narayanaswamy, Arun K. [2 ]
Tun, Tin A. [2 ]
Htoon, Hla M. [2 ]
Baskaran, Mani [2 ]
Perera, Shamira A. [2 ]
Aung, Tin [1 ,2 ,3 ]
机构
[1] Singapore Natl Eye Ctr, Glaucoma Serv, Singapore 168751, Singapore
[2] Singapore Eye Res Inst, Singapore, Singapore
[3] Natl Univ Singapore, Yong Loo Lin Sch Med, Singapore 117595, Singapore
基金
英国医学研究理事会;
关键词
ANTERIOR-CHAMBER ANGLE; LIWAN EYE; GONIOSCOPY; GLAUCOMA; CHINESE;
D O I
10.1167/iovs.12-10132
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
PURPOSE. To compare two spectral domain optical coherence tomography (SD-OCT) devices for the identification of angle structures and the presence of angle closure. METHODS. This was a prospective comparative study. Consecutive patients underwent gonioscopy and anterior segment imaging using two SD-OCT devices (iVue and Cirrus). Images were evaluated for the ability to detect angle structures such as Schwalbe's line (SL), trabecular meshwork (TM), Schlemm's canal (SC), and scleral spur (SS), and the presence of angle closure. Angle closure was defined as iris contact with the angle wall anterior to the SS on SD-OCT, and nonvisibility of the posterior TM on gonioscopy. Angle closure in an eye was defined as >= two quadrants of closed angles. AC1 statistic was used to assess the agreement between devices. RESULTS. Of the 69 subjects studied (46.4% male, 84.1% Chinese, mean age 64.0 +/- 10.5 years), 40 subjects (40 eyes, 58.0%) had angle closure on gonioscopy. The most identifiable structure on Cirrus SD-OCT was the SS (82.2%) and SL on iVue SD-OCT (74.5%). Angle closure was indeterminable in 14.5% and 50.7% of Cirrus and iVue scans (P < 0.001), respectively. Interdevice agreement for angle closure was moderately strong (AC1 = 0.67), but agreement with gonioscopy was only fair (AC1 = 0.35 and 0.50 for Cirrus and iVue, respectively). CONCLUSIONS. It was more difficult to determine angle closure status with iVue compared with Cirrus SD-OCT. There was fair agreement between both devices with gonioscopy for identifying angle closure. (Invest Ophthalmol Vis Sci. 2012;53:5131-5136) DOI:10.1167/iovs.12-10132
引用
收藏
页码:5131 / 5136
页数:6
相关论文
共 17 条
[1]   Detailed visualization of the anterior segment using Fourier-domain optical coherence tomography [J].
Asrani, Sanjay ;
Sarunic, Marinko ;
Santiago, Cecilia ;
Izatt, Joseph .
ARCHIVES OF OPHTHALMOLOGY, 2008, 126 (06) :765-771
[2]   Configuration of the drainage angle, intraocular pressure, and optic disc cupping in subjects with chronic angle-closure glaucoma [J].
Aung, T ;
Lim, MCC ;
Chan, YH ;
Rojanapongpun, P ;
Chew, PTK .
OPHTHALMOLOGY, 2005, 112 (01) :28-32
[3]   Novel anterior-chamber angle measurements by high-definition optical coherence tomography using the Schwalbe line as the landmark [J].
Cheung, Carol Y. ;
Zheng, Ce ;
Ho, Ching-Lin ;
Tun, Tin A. ;
Kumar, Rajesh S. ;
El Sayyad, Fouad ;
Wong, Tien Y. ;
Aung, Tin .
BRITISH JOURNAL OF OPHTHALMOLOGY, 2011, 95 (07) :955-959
[5]   Glaucoma in Mongolia - A population-based survey in Hovsgol Province, Northern Mongolia [J].
Foster, PJ ;
Baasanhu, J ;
Alsbirk, PH ;
Munkhbayar, D ;
Uranchimeg, D ;
Johnson, GJ .
ARCHIVES OF OPHTHALMOLOGY, 1996, 114 (10) :1235-1241
[6]   Detection of gonioscopically occludable angles and primary angle closure glaucoma by estimation of limbal chamber depth in Asians: modified grading scheme [J].
Foster, PJ ;
Devereux, JG ;
Alsbirk, PH ;
Lee, PS ;
Uranchimeg, D ;
Machin, D ;
Johnson, GJ ;
Baasanhu, J .
BRITISH JOURNAL OF OPHTHALMOLOGY, 2000, 84 (02) :186-192
[7]   The prevalence of glaucoma in Chinese residents of Singapore - A cross-sectional population survey of the Tanjong Pagar District [J].
Foster, PJ ;
Oen, FTS ;
Machin, D ;
Ng, TP ;
Devereux, JG ;
Johnson, GJ ;
Khaw, PT ;
Seah, SKL .
ARCHIVES OF OPHTHALMOLOGY, 2000, 118 (08) :1105-1111
[8]   Laser peripheral iridotomy in primary angle-closure suspects: Biometric and gonioscopic outcomes - The Liwan Eye Study [J].
He, Mingguang ;
Friedman, David S. ;
Ge, Jian ;
Huang, Wenyong ;
Jin, Chenjin ;
Lee, Pak Sang ;
Khaw, Peng T. ;
Foster, Paul J. .
OPHTHALMOLOGY, 2007, 114 (03) :494-500
[9]   Gonioscopy in adult Chinese: The Liwan Eye Study [J].
He, Mingguang ;
Foster, Paul J. ;
Ge, Jian ;
Huang, Wenyong ;
Wang, Dandan ;
Friedman, David S. ;
Khaw, Peng T. .
INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2006, 47 (11) :4772-4779
[10]  