Hydration measurements of the stratum corneum: comparison between the capacitance method (digital version of the Corneometer CM 825®) and the impedance method (Skicon-200EX®)

被引:111
作者
Clarys, Peter [1 ]
Clijsen, Ron [2 ]
Taeymans, Jan [3 ]
Barel, Andre O. [1 ]
机构
[1] Vrije Univ Brussel, Dept Human Biometry & Biomech, Fac Phys Educ & Physiotherapy, B-1050 Brussels, Belgium
[2] Univ Coll Physiotherapy Thim van der Laan, Landquart, Switzerland
[3] Berner Fachhsch, Bern, Switzerland
关键词
calibration; capacitance; conductance;
D O I
10.1111/j.1600-0846.2011.00573.x
中图分类号
R75 [皮肤病学与性病学];
学科分类号
100206 ;
摘要
Background Measurement of stratum corneum (SC) hydration often involves the use of commercial instruments. The aim of this study was to compare and validate two recent instruments: the Corneometer 825 (R) (digital probe) and the Skicon-200 EX (R). Methods In vitro calibration was carried out on filter pads using different solvents, measurements over different layers of plastic foils, and evaluation of desorption kinetics. In vivo measurements were carried out on skin sites covering a range of very dry to well-hydrated skin areas. Results Conductance measurements are influenced by electrolytes while capacitance measurements are not. Dielectric constant of the solvents influences the values of both instruments (r respectively 0.92 and 0.99). The capacitance method carries information from deeper layers (up to 45 mu m) compared with the conductance instrument (up to 15 mu m). Desorption experiments show a strong relation between the amount of water and respectively the capacitance and the conductance values. The in vivo experiments revealed a strong relation between the two methods (r = 0.97). Sensitivity of the capacitance method is limited for the highest hydration values. Conclusions Both instruments allow a certain calibration and both give good estimates of the SC hydration.
引用
收藏
页码:316 / 323
页数:8
相关论文
共 20 条
[1]  
[Anonymous], VALUES DIELECTRIC CO
[2]  
[Anonymous], HDB NONINVASIVE METH
[3]  
Barel A., 2006, Handbook of Non-Invasive Methods and the Skin, V2, P337
[4]  
Barel A O, 1997, Skin Res Technol, V3, P107, DOI 10.1111/j.1600-0846.1997.tb00171.x
[5]  
Barel A.O., 1995, HDB NONINVASIVE METH, P165
[6]  
Bielfeldt S, 2010, CAPACITANCE MEASUREM
[7]  
Clarys P., 1999, Skin Research and Technology, V5, P14, DOI DOI 10.1111/j.1600-0846.1999.tb00198.x
[8]  
Clarys P, INFLUENCE PROBE APPL, DOI [10. 1111/j. 1600-0846. 2011. 00516. x, DOI 10.1111/J.1600-0846.2011.00516.X]
[9]   Comparative study of five instruments measuring stratum corneurn hydration (Comeometer CM 820 and CM 825, Skicon 200, Nova DPM 9003, DermaLab). Part I. In vitro [J].
Fluhr, Joachim W. ;
Gloor, Max ;
Lazzerini, Sabrina ;
Kleesz, Peter ;
Grieshaber, Romano ;
Berardesca, Enzo .
SKIN RESEARCH AND TECHNOLOGY, 1999, 5 (03) :161-170
[10]   Comparative study of five instruments measuring stratum corneum hydration (Corneometer CM 820 and CM 825, Skicon 200, Nova DPM 9003, DermaLab). Part II. In vivo [J].
Fluhr, Joachim W. ;
Gloor, Max ;
Lazzerini, Sabrina ;
Kleesz, Peter ;
Grieshaber, Romano ;
Berardesca, Enzo .
SKIN RESEARCH AND TECHNOLOGY, 1999, 5 (03) :171-178