Objectives:Clinical evidence suggests that the use of Y-shaped locking plates with anterior and lateral locking screws provide improved primary fixation stability over the use of straight locking plates in the treatment of greater trochanter (GT) fractures. However, it remains unclear how the use of cable cerclages, in replacement or addition to the locking screws, would impact the primary fixation stability.Methods:Nine surgical fixations were tested on a dynamic test bench through 27 sawbones experiments. These fixations on the GT and bone shaft (BS) included locking screws alone, cable cerclages alone, or 7 combinations of both. Migrations, defined as the remaining movements of the unloaded GT fragment, were measured for all 9 fixations submitted to dynamic stair climbing cycling after an experimental plan.Results:Under dynamic stair climbing conditions: (1) Y-shaped locking plates fixed by locking screws alone (vs cable cerclages alone) significantly reduced GT migrations, (2) adding cable cerclages to screws provided no significant reduction of migrations, and (3) cable cerclages alone can be used on the BS but is not recommended on the GT.Conclusions:Y-shaped locking plates with anterior and lateral locking screws improved primary stability of GT fractures over fixation with cable cerclages. Cable cerclages provided acceptable primary stability when screw fixations were unachievable but on the BS only.
引用
收藏
页码:702 / 707
页数:6
相关论文
共 26 条
  • [1] COMPLICATIONS OF TROCHANTERIC OSTEOTOMY IN TOTAL HIP-REPLACEMENT
    AMSTUTZ, HC
    MAKI, S
    [J]. JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1978, 60 (02) : 214 - 216
  • [2] Testing System for the Comparative Evaluation of Greater Trochanter Re-attachment Devices
    Baril, Y.
    Bourgeois, Y.
    Brailovski, V.
    Duke, K.
    Laflamme, G. Y.
    Petit, Y.
    [J]. EXPERIMENTAL TECHNIQUES, 2012, 36 (04) : 74 - 82
  • [3] Improving greater trochanteric reattachment with a novel cable plate system
    Baril, Yannick
    Bourgeois, Yan
    Brailovski, Vladimir
    Duke, Kajsa
    Laflamme, G. Yves
    Petit, Yvan
    [J]. MEDICAL ENGINEERING & PHYSICS, 2013, 35 (03) : 383 - 391
  • [4] Current status of trochanteric reattachment in complex total hip arthroplasty
    Barrack, RL
    Butler, RA
    [J]. CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2005, (441) : 237 - 242
  • [5] Hip contact forces and gait patterns from routine activities
    Bergmann, G
    Deuretzbacher, G
    Heller, M
    Graichen, F
    Rohlmann, A
    Strauss, J
    Duda, GN
    [J]. JOURNAL OF BIOMECHANICS, 2001, 34 (07) : 859 - 871
  • [6] IS STAIRCASE WALKING A RISK FOR THE FIXATION OF HIP IMPLANTS
    BERGMANN, G
    GRAICHEN, F
    ROHLMANN, A
    [J]. JOURNAL OF BIOMECHANICS, 1995, 28 (05) : 535 - 553
  • [7] Subtrochanteric fixation stability depends on discrete fracture surface points
    Bredbenner, TL
    Snyder, SA
    Mazloomi, FR
    Le, T
    Wilber, RG
    [J]. CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2005, (432) : 217 - 225
  • [8] Charnley SJ., 1979, Low friction arthroplasty of the hip
  • [9] Fractures of the greater trochanter induced by osteolysis with the anatomic medullary locking prosthesis
    Claus, AM
    Hopper, RH
    Engh, CA
    [J]. JOURNAL OF ARTHROPLASTY, 2002, 17 (06) : 706 - 712
  • [10] RE-ATTACHMENT OF THE GREATER TROCHANTER - THE USE OF THE TROCHANTER CABLE-GRIP SYSTEM
    DALL, DM
    MILES, AW
    [J]. JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1983, 65 (01): : 55 - 59