Arm cuff in the measurement of blood pressure

被引:10
|
作者
Mattoo, TK [1 ]
机构
[1] Wayne State Univ, Childrens Hosp Michigan, Sch Med, Detroit, MI 48201 USA
关键词
blood pressure measurement; bladder dimensions; arm cuff; cuff selection;
D O I
10.1016/S0895-7061(01)02301-9
中图分类号
R6 [外科学];
学科分类号
1002 ; 100210 ;
摘要
Selection of an arm cuff of the right size is necessary for accurate measurement of blood pressure (BP). Incorrect measurement can result either in unnecessary investigations, treatment, and follow-up for the mistaken diagnosis of hypertension, or no treatment for hypertension in individuals mistakenly thought to have normal BP. This is of particular importance in children, who have age-dependent variations in body size and BP. Studies have revealed that health care providers often do not follow published guidelines concerning cuff selection for children. Perceptions larger BP cuffs are required if upper arm length is used as a criterion for cuff selection. In contrast, smaller cuffs are required if 40% of the upper mid-arm circumference is used as a criterion for cuff selection. No single cuff, when used in accordance with current recommendations, yields accurate systolic as well as diastolic BP. The century-old issue of how to select a right size arm cuff for an accurate measurement of BP remains unresolved. Am J Hypertens 2002;15:67S-68S (C) 2002 American Journal of Hypertension, Ltd.
引用
收藏
页码:67S / 68S
页数:2
相关论文
共 50 条
  • [1] VALIDITY OF BLOOD PRESSURE MEASUREMENT WITH CUFF IN ARM AND FOREARM
    FORSBERG, SA
    DEGUZMAN, M
    BERLIND, S
    ACTA MEDICA SCANDINAVICA, 1970, 188 (05): : 389 - +
  • [2] Influence of Age on Upper Arm Cuff Blood Pressure Measurement
    Picone, Dean S.
    Schultz, Martin G.
    Otahal, Petr
    Black, J. Andrew
    Bos, Willem J.
    Chen, Chen-Huan
    Cheng, Hao-Min
    Cremer, Antoine
    Dwyer, Nathan
    Fonseca, Ricardo
    Hughes, Alun D.
    Kim, Hack-Lyoung
    Lacy, Peter S.
    Laugesen, Esben
    Ohte, Nobuyuki
    Omboni, Stefano
    Ott, Christian
    Pereira, Telmo
    Pucci, Giacomo
    Roberts-Thomson, Philip
    Rossen, Niklas B.
    Schmieder, Roland E.
    Sueta, Daisuke
    Takazawa, Kenji
    Wang, Jiguang
    Weber, Thomas
    Westerhof, Berend E.
    Williams, Bryan
    Yamada, Hirotsugu
    Yamamoto, Eiichiro
    Sharman, James E.
    HYPERTENSION, 2020, 75 (03) : 844 - 850
  • [3] SHOULD BLOOD PRESSURE MEASUREMENT BE PERFORMED WITH THE CUFF ON A BARE ARM?
    Keltjens, E. B. M.
    Deinum, J.
    Lenders, J. W. M.
    Thien, Th.
    JOURNAL OF HYPERTENSION, 2009, 27 : S3 - S4
  • [4] BLOOD PRESSURE MEASUREMENT: ARM CIRCUMFERENCE AND CUFF SIZE AVAILABILITY
    Veiga, Eugenia Velludo
    Moura Arcuri, Edna Aparecida
    Cloutier, Lyne
    Ferreira Santos, Jair Licio
    REVISTA LATINO-AMERICANA DE ENFERMAGEM, 2009, 17 (04): : 455 - 461
  • [5] THE EFFECT OF ARM CIRCUMFERENCE AND BLOOD PRESSURE CUFF SIZE ON BLOOD PRESSURE MEASUREMENT IN PREGNANCY
    Mangos, G. J.
    Ong, S.
    Kho, L.
    Brown, M.
    NEPHROLOGY, 2008, 13 : A129 - A129
  • [6] Clinical assessment of new technology for blood pressure measurement without arm cuff
    Watanabe, N.
    Bando, Y. Kureishi
    Kawachi, T.
    Futatsuyama, K.
    Yamakita, H.
    Nishimura, K.
    Okumura, T.
    Ishii, H.
    Murohara, T.
    EUROPEAN HEART JOURNAL, 2016, 37 : 401 - 401
  • [7] THE IMPACT OF RELATION BETWEEN CUFF SIZE AND ARM CIRCUMFERENCE ON BLOOD PRESSURE MEASUREMENT
    Xie Dongyang
    Cai Jiumei
    HEART, 2012, 98 : E254 - E254
  • [8] Blood pressure measurement in pregnancy: the effect of arm circumference and sphygmomanometer cuff size
    Kho, Chye L.
    Brown, Mark A.
    Ong, Sharon L. H.
    Mangos, George J.
    OBSTETRIC MEDICINE, 2009, 2 (03) : 116 - 120
  • [9] Blood Pressure Measurement in Pregnancy: The Effect of Arm Circumference and Sphygmomanometer Cuff Size
    Brown, Mark
    Kho, Lily
    Ong, S.
    Mangos, George
    HYPERTENSION IN PREGNANCY, 2008, 27 (04) : 595 - 595
  • [10] Influence of Age on Upper Arm Cuff Blood Pressure Measurement Across the Lifespan
    Bui, Tan V.
    Sharman, James E.
    Mynard, Jonathan P.
    Ayer, Julian G.
    Goupil, Remi
    Picone, Dean S.
    JOURNAL OF THE AMERICAN HEART ASSOCIATION, 2024, 13 (03):