An Integrated Framework Based on Full Monte Carlo Simulations for Double-Scattering Proton Therapy

被引:0
作者
Yuan, Jiankui [1 ]
Mansur, David [1 ]
Yao, Min [1 ]
Biswas, Tithi [1 ]
Zheng, Yiran [1 ]
Jesseph, Rick [1 ]
Jin, Jian-Yue [1 ]
Machtay, Mitchell [1 ]
机构
[1] Univ Hosp Cleveland, Med Ctr, Dept Radiat Oncol, 11100 Euclid Ave, Cleveland, OH 44106 USA
来源
INTERNATIONAL JOURNAL OF PARTICLE THERAPY | 2019年 / 6卷 / 02期
关键词
double scattering; proton therapy; Monte Carlo; source model; PENCIL BEAM ALGORITHM; DOSE-CALCULATION; MODEL;
D O I
10.14338/IJPT-19-00063.1
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: We developed an integrated framework that employs a full Monte Carlo (MC) model for treatment-plan simulations of a passive double-scattering proton system. Materials and Methods: We have previously validated a virtual machine source model for full MC proton-dose calculations by comparing the percentage of depth-dose curves, spread-out Bragg peaks, and lateral profiles against measured commissioning data. This study further expanded our previous work by developing an integrate framework that facilitates its clinical use. Specifically, we have (1) constructed patient-specific applicator and compensator numerically from the plan data and incorporated them into the beamline, (2) created the patient anatomy from the computed tomography image and established the transformation between patient and machine coordinate systems, and (3) developed a graphical user interface to ease the whole process from importing the treatment plan in the Digital Imaging and Communications in Medicine format to parallelization of the MC calculations. End-to-end tests were performed to validate the functionality, and 3 clinical cases were used to demonstrate clinical utility of the framework. Results: The end-to-end tests demonstrated that the framework functioned correctly for all tested functionality. Comparisons between the treatment planning system calculations and MC results in 3 clinical cases revealed large dose difference up to 17%, especially in the beam penumbra and near the end of beam range. The discrepancy likely originates from a variety of sources, such as the dose algorithms, modeling of the beamline, and the dose metric. The agreement for other regions was acceptable. Conclusion: An integrated framework was developed for full MC simulations of double-scattering proton therapy. It can be a valuable tool for dose verification and plan evaluation.
引用
收藏
页码:31 / 41
页数:11
相关论文
共 23 条
[1]   Uncertainties and correction methods when modeling passive scattering proton therapy treatment heads with Monte Carlo [J].
Bednarz, Bryan ;
Lu, Hsiao-Ming ;
Engelsman, Martijn ;
Paganetti, Harald .
PHYSICS IN MEDICINE AND BIOLOGY, 2011, 56 (09) :2837-2854
[2]   An analytical approximation of the Bragg curve for therapeutic proton beams [J].
Bortfeld, T .
MEDICAL PHYSICS, 1997, 24 (12) :2024-2033
[3]   Validation of GEANT4, an object-oriented Monte Carlo toolkit, for simulations in medical physics [J].
Carrier, JF ;
Archambault, L ;
Beaulieu, L ;
Roy, R .
MEDICAL PHYSICS, 2004, 31 (03) :484-492
[4]   A pencil beam algorithm for proton dose calculations [J].
Hong, L ;
Goitein, M ;
Bucciolini, M ;
Comiskey, R ;
Gottschalk, B ;
Rosenthal, S ;
Serago, C ;
Urie, M .
PHYSICS IN MEDICINE AND BIOLOGY, 1996, 41 (08) :1305-1330
[5]   Oblique incidence for broad monoenergetic proton beams [J].
Jette, David ;
Yuan, Jiankui ;
Chen, Weimin .
MEDICAL PHYSICS, 2010, 37 (11) :5683-5690
[6]   Effects of Hounsfield number conversion on CT based proton Monte Carlo dose calculations [J].
Jiang, Hongyu ;
Seco, Joao ;
Paganetti, Harald .
MEDICAL PHYSICS, 2007, 34 (04) :1439-1449
[7]  
Kikinis R, 2014, INTRAOPERATIVE IMAGI, DOI [10.1007/978-1-4614-7657-3_19, DOI 10.1007/978-1-4614-7657-3_19]
[8]   Design and Verification of a Heterogeneous Proton Equivalent Thorax Phantom for Use in End-To-End Assessment of Pencil Beam Proton Therapy [J].
Neihart, J. ;
Sahoo, N. ;
Balter, P. ;
Summers, P. ;
Palmer, M. ;
Kerr, M. ;
Followill, D. .
MEDICAL PHYSICS, 2013, 40 (06)
[9]   Monte Carlo simulations for configuring and testing an analytical proton dose-calculation algorithm [J].
Newhauser, Wayne ;
Fontenot, Jonas ;
Zheng, Yuanshui ;
Polf, Jerimy ;
Titt, Uwe ;
Koch, Nicholas ;
Zhang, Xiaodong ;
Mohan, Radhe .
PHYSICS IN MEDICINE AND BIOLOGY, 2007, 52 (15) :4569-4584
[10]   Clinical implementation of full Monte Carlo dose calculation in proton beam therapy [J].
Paganetti, Harald ;
Jiang, Hongyu ;
Parodi, Katia ;
Slopsema, Roelf ;
Engelsman, Martijn .
PHYSICS IN MEDICINE AND BIOLOGY, 2008, 53 (17) :4825-4853