Large Axial Range Frequency-Domain Optical Low Coherence Interferometry

被引:4
作者
Cerrotta, Santiago [1 ]
Morel, Eneas N. [1 ,2 ]
Alvira, Fernando C. [3 ,4 ]
Torga, Jorge R. [1 ,2 ]
机构
[1] Natl Technol Univ, Fac Reg Delta, C1041AAJ, Campana, Argentina
[2] Consejo Nacl Invest Cient & Tecn, Natl Sci & Tech Res Council, C1425FQB, Buenos Aires, DF, Argentina
[3] Natl Univ Quilmes, Biomembranes Lab, B1876BXD, Bernal, Argentina
[4] Natl Univ Quilmes, Control Automat & Instrumentat Lab, B1876BXD, Bernal, Argentina
关键词
Interferometry; FD-LCI; FD-OCT; optical coherence tomography; large axial range; SPECTROMETER;
D O I
10.1109/LPT.2018.2883207
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
In this letter, we propose a new experimental setup to extend the measurement range of frequency-domain low coherence interferometry (FD-LCI) from 3 mm to 2 cm. The FD-LCI can be extended to the technique of frequency domain optical coherence tomography by performing additional lateral scanning. This new and simple arrangement just attaches an extra interferometer before the spectrometer in a typical FD-LCI setup. With this configuration, it is possible to overcome the limitation of the dynamic range of the conventional technique. This allows us to measure the optical path difference of the sample of interest, which is outside the detection range of the spectrometer, using an extra interferometer as a tuner system. In addition, this tuning allows significant minimization of the effects produced by the fall-off. The accuracy of the new system is obtained by adding the FD-LCI and the tuning micrometric translator accuracies. In our experimental setup, this value is 20 mu m.
引用
收藏
页码:125 / 128
页数:4
相关论文
共 16 条
[1]   Scanning optical coherence tomography applied to the characterization of surfaces and coatings [J].
Cerrotta, S. ;
Morel, E. N. ;
Torga, J. R. .
INTERNATIONAL CONGRESS OF SCIENCE AND TECHNOLOGY OF METALLURGY AND MATERIALS, SAM - CONAMET 2014, 2015, 9 :142-149
[2]   Wide band interferometry for thickness measurement [J].
Costantino, S ;
Martínez, OE ;
Torga, JR .
OPTICS EXPRESS, 2003, 11 (08) :952-957
[3]  
Fujimoto J, 2008, BIOL MED PHYS BIOMED, P1, DOI 10.1007/978-3-540-77550-8_1
[4]   Optical Coherence Tomography: History, Current Status, and Laboratory Work [J].
Gabriele, Michelle L. ;
Wollstein, Gadi ;
Ishikawa, Hiroshi ;
Kagemann, Larry ;
Xu, Juan ;
Folio, Lindsey S. ;
Schuman, Joel S. .
INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2011, 52 (05) :2425-2436
[5]  
Hirai A., 2003, CLEO/Pacific Rim 2003. The 5th Pacific Rim Conference on Lasers and Electro-Optics (IEEE Cat. No.03TH8671), DOI 10.1109/CLEOPR.2003.1277183
[6]   Analytical model of spectrometer-based two-beam spectral interferometry [J].
Hu, Zhilin ;
Pan, Yinsheng ;
Rollins, Andrew M. .
APPLIED OPTICS, 2007, 46 (35) :8499-8505
[7]   OPTICAL COHERENCE TOMOGRAPHY [J].
HUANG, D ;
SWANSON, EA ;
LIN, CP ;
SCHUMAN, JS ;
STINSON, WG ;
CHANG, W ;
HEE, MR ;
FLOTTE, T ;
GREGORY, K ;
PULIAFITO, CA ;
FUJIMOTO, JG .
SCIENCE, 1991, 254 (5035) :1178-1181
[8]   Design of a k-space spectrometer for ultra-broad waveband spectral domain optical coherence tomography [J].
Lan, Gongpu ;
Li, Guoqiang .
SCIENTIFIC REPORTS, 2017, 7
[9]   Performance of fourier domain vs. time domain optical coherence tomography [J].
Leitgeb, R ;
Hitzenberger, CK ;
Fercher, AF .
OPTICS EXPRESS, 2003, 11 (08) :889-894
[10]   Measuring large step heights by variable synthetic wavelength interferometry [J].
Lu, SH ;
Lee, CC .
MEASUREMENT SCIENCE AND TECHNOLOGY, 2002, 13 (09) :1382-1387