The Insertion of Intramedullary Nail Locking Screws Without Fluoroscopy: A Faster and Safer Technique

被引:24
作者
Chan, Daniel S. [1 ]
Burris, Richard B. [2 ]
Erdogan, Murat [3 ]
Sagi, H. Claude [1 ]
机构
[1] Florida Orthopaed Inst, Orthopaed Trauma Serv, Tampa, FL 33606 USA
[2] Univ S Florida, Tampa, FL USA
[3] Ondokuz Mayis Univ, Samsun, Turkey
关键词
IMN; fluoroscopy; radiation; locking screw; electromagnetic; SureShot; INTERLOCKING NAILS; RADIATION-EXPOSURE; DISTAL LOCKING; FRACTURES; SURGEON; DEVICE;
D O I
10.1097/BOT.0b013e3182828e10
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Objective: This study was designed to compare the accuracy, time, and radiation exposure during the insertion of intramedullary nail locking screws using either standard fluoroscopic assistance or an electromagnetic (EM)-based navigational system without fluoroscopy.Design: Prospective.Setting: Level I academic trauma center.Methods: Patients were divided into 2 groups: group 1 (fluoroscopic assistance), consisted of standard freehand fluoroscopically assisted insertion of locking screws (OEC 9900; G.E. HealthCare, Waukesha, WI), whereas group 2 (EM), consisted of EM navigationally assisted insertion without fluoroscopy (SureShot; Smith & Nephew, Memphis, TN). Technician arrival time, setup (SU) time, screw insertion (SI) time (seconds), fluoroscopy time (seconds), radiation exposure (mrads), and accuracy (hit or miss) were recorded for each screw. For group 1, the SU time was recorded as the time and radiation required to obtain perfect circles before insertion, and for group 2, the SU time was recorded as the time required to set up the navigational EM unit. Data collected regarding SI were then compared using standard analysis of variance.Results: Forty-one locking screws were inserted in group 1, whereas 60 screws were inserted in group 2. Accuracy was 100% for both groups. For group 1, mean technician wait time was 77 seconds plus a mean perfect circle SU time of 105 seconds (9.2 mrads and 10 seconds of fluoroscopy). Mean SU time for group 2 was 94 seconds (no fluoroscopy). Mean insertion time was 342 seconds per screw for group 1 (32.9 mrads and 18 seconds of fluoroscopy) compared with 234 seconds per screw for group 2 (no fluoroscopy). These differences were statistically significant (P = 0.006).Conclusions: The use of EM navigation (SureShot; Smith & Nephew) for the insertion of intramedullary nail locking screws demonstrated accuracy similar to conventional fluoroscopic-guided insertion. However, EM-guided locking SI resulted in a significantly shorter total procedural time and completely eliminated radiation exposure.
引用
收藏
页码:363 / 366
页数:4
相关论文
共 16 条
[1]   Distal locking of tibial nails - A new device to reduce radiation exposure [J].
Anastopoulos, George ;
Ntagiopoulos, Panagiotis G. ;
Chissas, Dionisios ;
Papaeliou, Athanasios ;
Asimakopoulos, Antonios .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2008, 466 (01) :216-220
[2]  
Center for Devices and Radiological Health U.S. Food and Drug Administration, 2010, IN RED UNN RAD EXP M
[3]  
COETZEE JC, 1992, S AFR MED J, V81, P312
[4]  
Gugala G, 2001, INJURY S4, V32, pSD21
[5]   Achieving interlocking nails without using an image intensifier [J].
Ikem, Innocent C. ;
Ogunlusi, Johnson D. ;
Ine, Henry R. .
INTERNATIONAL ORTHOPAEDICS, 2007, 31 (04) :487-490
[6]   External jig-aided intramedullary interlocking nailing of diaphyseal fractures: experience from a tropical developing centre [J].
Ikpeme, Ikpeme ;
Ngim, Ngim ;
Udosen, Anthony ;
Onuba, Obiora ;
Enembe, Okokon ;
Bello, Segun .
INTERNATIONAL ORTHOPAEDICS, 2011, 35 (01) :107-111
[7]   A SIMPLE TECHNIQUE FOR INSERTION OF DISTAL SCREWS INTO INTERLOCKING NAILS [J].
KELLEY, SS ;
BONAR, S ;
HUSSAMY, OD ;
MORRISON, JA .
JOURNAL OF ORTHOPAEDIC TRAUMA, 1995, 9 (03) :227-230
[8]   CLOSED LOCKED INTRAMEDULLARY NAILING - ITS APPLICATION TO COMMINUTED FRACTURES OF THE FEMUR [J].
KEMPF, I ;
GROSSE, A ;
BECK, G .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1985, 67A (05) :709-720
[9]   A new mechanical aiming device for the placement of distal interlocking screws in femoral nails [J].
Krettek, C ;
Konemann, B ;
Miclau, T ;
Kolbli, R ;
Machreich, T ;
Kromm, A ;
Tscherne, H .
ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 1998, 117 (03) :147-152
[10]   RADIATION EXPOSURE TO THE SURGEON DURING CLOSED INTERLOCKING INTRAMEDULLARY NAILING [J].
LEVIN, PE ;
SCHOEN, RW ;
BROWNER, BD .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1987, 69A (05) :761-766