Preparation and Characterization of Porous Titanium Alloy Scaffold Coated with Tantalum

被引:0
作者
Li Xiang [1 ]
Yu Xiaoming [2 ]
Wang Chengtao
Tan Lili [2 ]
Yang Ke [2 ]
Luo Yun
Zhang Wenguang
机构
[1] Shanghai Jiao Tong Univ, Sch Mech Engn, State Key Lab Mech Syst & Vibrat, Shanghai 200240, Peoples R China
[2] Chinese Acad Sci, Inst Met Res, Shenyang 110016, Peoples R China
关键词
electron beam melting; scaffold; tantalum coating; titanium alloy; MECHANICAL-PROPERTIES; POROSITY; INGROWTH;
D O I
暂无
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
Tantalum is an ideal medical metal material, which can integrate with human soft/hard tissue. Chemical vapor deposition method was used to deposit tantalum film on the surface of Ti6A14V scaffolds with controlled porous structure. The Ti6A14V scaffolds could have ideal three-dimensional pore structure and mechanical compatibility, but also have excellent biological properties of tantalum. The investigation results show that the surface color changes of porous scaffolds before and after coating are obvious. After coating, the scaffolds show Ta metal color. The scanning electron microscopy (SEM) and X-ray diffraction (XRD) tests demonstrate that the deposition on the surface of porous titanium alloy scaffolds is the tantalum metal. Compared to the Zimmer trabecular metal, the elastic modulus and compressive strength of the porous titanium alloy scaffold coated with tantalum are more similar to that of human cortical bone, which could be an ideal substitute for bone repair.
引用
收藏
页码:2049 / 2053
页数:5
相关论文
共 14 条
[1]  
Hacking SA, 2000, J BIOMED MATER RES, V52, P631
[2]   Cellular Ti-6Al-4V structures with interconnected macro porosity for bone implants fabricated by selective electron beam melting [J].
Heinl, Peter ;
Mueller, Lenka ;
Koerner, Carolin ;
Singer, Robert F. ;
Mueller, Frank A. .
ACTA BIOMATERIALIA, 2008, 4 (05) :1536-1544
[3]   Structural, mechanical and in vitro characterization of individually structured Ti-6Al-4V produced by direct laser forming [J].
Hollander, DA ;
von Walter, M ;
Wirtz, T ;
Sellei, R ;
Schmidt-Rohlfing, B ;
Paar, O ;
Erli, HJ .
BIOMATERIALS, 2006, 27 (07) :955-963
[4]   Hydroxyapatite coating on titanium substrate with titania buffer layer processed by sol-gel method [J].
Kim, HW ;
Koh, YH ;
Li, LH ;
Lee, S ;
Kim, HE .
BIOMATERIALS, 2004, 25 (13) :2533-2538
[5]  
Levine Brett, 2007, J Knee Surg, V20, P185
[6]   Experimental and clinical performance of porous tantalum in orthopedic surgery [J].
Levine, Brett Russell ;
Sporer, Scott ;
Poggie, Robert A. ;
Della Valle, Craig J. ;
Jacobs, Joshua J. .
BIOMATERIALS, 2006, 27 (27) :4671-4681
[7]   Fabrication and characterization of porous Ti6Al4V parts for biomedical applications using electron beam melting process [J].
Li, Xiang ;
Wang, Chengtao ;
Zhang, Wenguang ;
Li, Yuanchao .
MATERIALS LETTERS, 2009, 63 (3-4) :403-405
[8]   Comparison of the mechanical properties between tantalum and nickel-titanium foams implant materials for bone ingrowth applications [J].
Sevilla, P. ;
Aparicio, C. ;
Planell, J. A. ;
Gil, F. J. .
JOURNAL OF ALLOYS AND COMPOUNDS, 2007, 439 (1-2) :67-73
[9]   Effect of porosity on the fluid flow characteristics and mechanical properties of tantalum scaffolds [J].
Shimko, DA ;
Shimko, VF ;
Sander, EA ;
Dickson, KF ;
Nauman, EA .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH PART B-APPLIED BIOMATERIALS, 2005, 73B (02) :315-324
[10]   Effects of titanium surface topography on bone integration: a systematic review [J].
Wennerberg, Ann ;
Albrektsson, Tomas .
CLINICAL ORAL IMPLANTS RESEARCH, 2009, 20 :172-184