Validation of a NaI(Tl) detector's model developed with MCNP-X code

被引:89
作者
Salgado, C. M. [1 ]
Brandao, L. E. B. [1 ]
Schirru, R. [2 ]
Pereira, C. M. N. A. [1 ]
Conti, C. C. [3 ]
机构
[1] CNEN IEN, Inst Engn Nucl, BR-21945970 Rio De Janeiro, Brazil
[2] Univ Fed Rio de Janeiro, COPPE PEN, BR-21941972 Rio De Janeiro, Brazil
[3] CNEN IRD, Inst Radioprotecao & Dosimetria, BR-22780160 Rio De Janeiro, Brazil
关键词
MCNP-X simulation; NaI (Tl) detector; Gamma spectrometry; MONTE-CARLO CALCULATION; RESPONSE FUNCTIONS; GAMMA-RAYS; SCINTILLATION DETECTORS; SIMULATION; EFFICIENCIES; CALIBRATION;
D O I
10.1016/j.pnucene.2012.03.006
中图分类号
TL [原子能技术]; O571 [原子核物理学];
学科分类号
0827 ; 082701 ;
摘要
The Monte Carlo method was used to calculate the photon detection efficiency and energy resolution curves for a 1.5 '' x 1 '' NaI(Tl) scintillator detector (crystal + housing + photomultiplier tube material equivalent) exposed to gamma rays in the energy range from 20 key to 662 keV. This work aims to design a precise computational model, based in Monte Carlo simulation, which can be used in practical application. The energy resolution curve was used to improve the response of the mathematical simulation of the detector. The detector was modeled with the MCNP-X code and the results were compared to experimental photopeak efficiency measurements of radiation sources. The results showed good agreement with the experimental data. (C) 2012 Elsevier Ltd. All rights reserved.
引用
收藏
页码:19 / 25
页数:7
相关论文
共 27 条
[1]   RESPONSE FUNCTIONS FOR SODIUM IODIDE SCINTILLATION DETECTORS [J].
BERGER, MJ ;
SELTZER, SM .
NUCLEAR INSTRUMENTS & METHODS, 1972, 104 (02) :317-&
[2]  
Conti C.C., 1999, THESIS COPPE UFRJ RI
[3]  
Conti CC., 1999, 2 INT S TECHN ENH NA
[4]  
Debertin K., 1988, Gamma- and X-Ray spectrometry with semiconductor detectors
[5]  
Ewa I.O., 2001, NUCL INSTRUM METH A, V479, P618
[6]   MONTE-CARLO CALCULATION OF ENERGY LOSS SPECTRA FOR GAMMA-RAYS IN CYLINDRICAL NAI(TI) CRYSTALS [J].
GIANNINI, M ;
OLIVA, PR ;
RAMORINO, MC .
NUCLEAR INSTRUMENTS & METHODS, 1970, 81 (01) :104-&
[7]  
Hashem M., 2007, ASIAN J EXP SCI, V21, P1
[8]  
HEATH RL, 1964, SCINTILLATION SPECTR, V1
[9]   Study of the NaI(Tl) efficiency by Monte Carlo method [J].
Jehouani, A ;
Ichaoui, R ;
Boulkheir, M .
APPLIED RADIATION AND ISOTOPES, 2000, 53 (4-5) :887-891
[10]   Monte Carlo simulations of response functions for gas filled and scintillator detectors with MCNPX code [J].
Kovaltchouk, V. ;
Machrafi, R. .