Heated Slurry Sampling for the Determination of Cadmium in Food by Electrothermal Atomic Absorption Spectrometry
被引:10
作者:
Peng, Yue'e
论文数: 0引用数: 0
h-index: 0
机构:
China Univ Geosci, State Key Lab Biogeol & Environm Geol, Wuhan 430074, Peoples R China
China Univ Geosci, Fac Mat Sci & Chem, Wuhan 430074, Peoples R ChinaChina Univ Geosci, State Key Lab Biogeol & Environm Geol, Wuhan 430074, Peoples R China
Peng, Yue'e
[1
,2
]
Guo, Wei
论文数: 0引用数: 0
h-index: 0
机构:
China Univ Geosci, State Key Lab Biogeol & Environm Geol, Wuhan 430074, Peoples R ChinaChina Univ Geosci, State Key Lab Biogeol & Environm Geol, Wuhan 430074, Peoples R China
Guo, Wei
[1
]
Zhang, Ping
论文数: 0引用数: 0
h-index: 0
机构:
PerkinElmer Inc, Beijing CKC, Beijing, Peoples R ChinaChina Univ Geosci, State Key Lab Biogeol & Environm Geol, Wuhan 430074, Peoples R China
Zhang, Ping
[3
]
Jin, Lanlan
论文数: 0引用数: 0
h-index: 0
机构:
China Univ Geosci, State Key Lab Biogeol & Environm Geol, Wuhan 430074, Peoples R ChinaChina Univ Geosci, State Key Lab Biogeol & Environm Geol, Wuhan 430074, Peoples R China
Jin, Lanlan
[1
]
Hu, Shenghong
论文数: 0引用数: 0
h-index: 0
机构:
China Univ Geosci, State Key Lab Biogeol & Environm Geol, Wuhan 430074, Peoples R ChinaChina Univ Geosci, State Key Lab Biogeol & Environm Geol, Wuhan 430074, Peoples R China
Hu, Shenghong
[1
]
机构:
[1] China Univ Geosci, State Key Lab Biogeol & Environm Geol, Wuhan 430074, Peoples R China
[2] China Univ Geosci, Fac Mat Sci & Chem, Wuhan 430074, Peoples R China
[3] PerkinElmer Inc, Beijing CKC, Beijing, Peoples R China
A heating procedure is reported with slurry sampling electrothermal atomic absorption spectrometry to improve the accuracy of cadmium determination in food. In comparison to conventional slurry sampling, the heating significantly increased cadmium recovery and improved the precision. For the optimized procedure, 25-250mg of food were treated with 2% HNO3 and 1% H2O2 with heating at 120 degrees C for 20min, followed by the addition of 50 mu L of 10% Triton X-100, and homogenization in an ultrasonic bath prior to analysis. Tungsten and rhodium were employed as a permanent modifier with optimum pyrolysis and atomization temperatures of 500 degrees C and 1500 degrees C. Calibration with aqueous standards resulted in good agreement between certified or information values and measured results at the 95% confidence level. A characteristic mass of 0.8 +/- 0.1 pg and a detection limit of 0.7ngg(-1) for a 2% slurry were obtained. The method was employed for the direct determination of cadmium in food certified reference materials.
机构:
Sao Paulo State Univ Unesp, Dept Analyt Chem, POB 355, BR-14801970 Araraquara, SP, BrazilSao Paulo State Univ Unesp, Dept Analyt Chem, POB 355, BR-14801970 Araraquara, SP, Brazil
Fortunato, Felipe Manfroi
Gomes Neto, Jose Anchieta
论文数: 0引用数: 0
h-index: 0
机构:
Sao Paulo State Univ Unesp, Dept Analyt Chem, POB 355, BR-14801970 Araraquara, SP, BrazilSao Paulo State Univ Unesp, Dept Analyt Chem, POB 355, BR-14801970 Araraquara, SP, Brazil
Gomes Neto, Jose Anchieta
Giovanni Freschi, Gian Paulo
论文数: 0引用数: 0
h-index: 0
机构:
Alfenas Fed Univ UNIFAL MG, Inst Sci & Technol, BR-37715400 Pocos De Caldas, MG, BrazilSao Paulo State Univ Unesp, Dept Analyt Chem, POB 355, BR-14801970 Araraquara, SP, Brazil
机构:
Sao Paulo State Univ Unesp, Dept Analyt Chem, POB 355, BR-14801970 Araraquara, SP, BrazilSao Paulo State Univ Unesp, Dept Analyt Chem, POB 355, BR-14801970 Araraquara, SP, Brazil
Fortunato, Felipe Manfroi
Gomes Neto, Jose Anchieta
论文数: 0引用数: 0
h-index: 0
机构:
Sao Paulo State Univ Unesp, Dept Analyt Chem, POB 355, BR-14801970 Araraquara, SP, BrazilSao Paulo State Univ Unesp, Dept Analyt Chem, POB 355, BR-14801970 Araraquara, SP, Brazil
Gomes Neto, Jose Anchieta
Giovanni Freschi, Gian Paulo
论文数: 0引用数: 0
h-index: 0
机构:
Alfenas Fed Univ UNIFAL MG, Inst Sci & Technol, BR-37715400 Pocos De Caldas, MG, BrazilSao Paulo State Univ Unesp, Dept Analyt Chem, POB 355, BR-14801970 Araraquara, SP, Brazil