Trace determination of perchlorate in drinking water by capillary zone electrophoresis with isotachophoresis sample cleanup and conductivity detection

被引:3
作者
Zabensky, Branislav [1 ]
Bodor, Robert [1 ]
Makata, David [1 ]
Szucs, Roman [1 ]
Masar, Marian [1 ]
机构
[1] Comenius Univ, Fac Nat Sci, Dept Analyt Chem, Mlynska Dolina CH2,Ilkovicova 6, Bratislava 84215, Slovakia
关键词
capillary zone electrophoresis; column coupling technology; isotachophoresis; perchlorate; polyvinylpyrrolidone; ION CHROMATOGRAPHY; ELECTROKINETIC INJECTION; POSTBLAST RESIDUES; MASS-SPECTROMETRY; INORGANIC ANIONS; IDENTIFICATION; CONTAMINATION; SEPARATIONS; ACIDS; CHIP;
D O I
10.1002/jssc.202200188
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
An analytical test procedure for the direct determination of trace levels of perchlorate in drinking water by isotachophoresis combined with capillary zone electrophoresis was developed. A capillary electrophoresis analyzer with column coupling technology, capable of combining capillaries with different internal diameters, was employed in combination with conductivity detection. This combination of the capillary electrophoresis techniques facilitated preconcentration of the trace analytes and elimination of potentially interfering macro-components. To eliminate the influence of weak and moderately strong acids on the migration of perchlorate, acidic leading electrolyte (pH 3.2) in the isotachophoresis step and acidic background electrolyte (pH 3.9) in the zone electrophoresis step were chosen. The addition of polyvinylpyrrolidone into the electrolytes enhanced the resolution of perchlorate from other anions, especially remaining anionic macro-components. The developed method is characterized by good repeatability of migration time (relative standard deviation less than 0.2%) as well as peak area (relative standard deviation less than 5.9%), linearity (R = 0.9996), recoveries (100%-112%), and sample throughput (90 samples/24 h). The limit of quantitation for perchlorate in drinking water was achieved at 12.5 nmol/L (1.25 mu g/L). This approach is more sensitive and more robust than transient isotachophoresis and offers advantages over some more established analytical techniques such as ion chromatography.
引用
收藏
页码:3339 / 3347
页数:10
相关论文
共 57 条
  • [51] US Environmental Protection Agency Method 314.1, an automated sample preconcentration/matrix elimination suppressed conductivity method for the analysis of trace levels (0.50 μg/L) of perchlorate in drinking water
    Wagner, Herbert P.
    Pepich, B. V.
    Pohl, C.
    Later, D.
    Joyce, R.
    Srinivasan, K.
    Thomas, D.
    Woodruff, A.
    DeBorba, B.
    Munch, D. J.
    [J]. JOURNAL OF CHROMATOGRAPHY A, 2006, 1118 (01) : 85 - 93
  • [52] Wendelken S., 2005, 3310 US EPA
  • [53] WHO, 2007, GUIDELINES DRINKING
  • [54] Selection and optimization of background electrolytes for simultaneous detection of small cations and organic acids by capillary electrophoresis with indirect photometry
    Xiong, X
    Li, SFY
    [J]. JOURNAL OF CHROMATOGRAPHY A, 1998, 822 (01) : 125 - 136
  • [55] Hyphenation of field-amplified sample injection and transient isotachophoresis in CE for the determination of sotalol and metoprolol in human urine samples
    Xu, Lei
    Meng, Weiqi
    Lu, Jing
    Cui, Fengjuan
    Gao, Lidi
    Chen, Lin
    Xin, Yang
    [J]. JOURNAL OF SEPARATION SCIENCE, 2020, 43 (11) : 2193 - 2200
  • [56] Simultaneous separation and indirect ultraviolet detection of chlorate and perchlorate by pyridinium ionic liquids in reversed-phase liquid chromatography
    Yin, Zhen-jie
    Zhang, Ya-nan
    Guan, Fu-jing
    Yu, Hong
    Ma, Ya-jie
    [J]. JOURNAL OF SEPARATION SCIENCE, 2020, 43 (20) : 3868 - 3875
  • [57] Perchlorate occurrence, sub-basin contribution and risk hotspots for drinking water sources in China based on industrial agglomeration method
    Zhang, Bin
    An, Wei
    Shi, Yali
    Yang, Min
    [J]. ENVIRONMENT INTERNATIONAL, 2022, 158