An improved sample preparation method for the sensitive detection of peptides by MALDI-MS

被引:6
作者
Hioki, Yusaku [1 ]
Kuyama, Hiroki [1 ]
Hamana, Chikako [1 ]
Takeyama, Kohei [1 ]
Tanaka, Koichi [1 ]
机构
[1] Shimadzu Co Ltd, KTLAST, Nakagyo Ku, Kyoto 6048511, Japan
来源
JOURNAL OF MASS SPECTROMETRY | 2013年 / 48卷 / 11期
基金
日本学术振兴会;
关键词
MALDI mass spectrometry; sensitivity; peptide analysis; matrix concentration; acetonitrile content; DESORPTION/IONIZATION MASS-SPECTROMETRY; MATRIX; IONIZATION; ADSORPTION; MIXTURES; SURFACE; MS/MS; TOF;
D O I
10.1002/jms.3283
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
We describe here an optimization study of the sample preparation conditions for sensitive detection of peptides by matrix-assisted laser desorption/ionization mass spectrometry (MALDI-MS). Among many factors in the conditions, we varied the percent acetonitrile in the peptide solution, the percent acetonitrile in the matrix solution and the -cyano-4-hydroxycinnamic acid (CHCA) concentration in the matrix solution. CHCA was chosen because it is the most frequently used matrix for analyzing peptides. The well-established dried-droplet method was employed for sample deposition. The examined range of the concentration of CHCA was from 0.01 to 10mg/ml, and the MeCN content of the solvent for matrix/analyte was 10% to 50%. The indicator for the detection sensitivity was the S/N ratio of the peaks of peptides used. Highly increased sensitivity (100- to 1000-fold) was observed for the optimal CHCA concentration of 0.1mg/ml in 20% MeCN/0.1% aq. trifluoroacetic acid (TFA), as compared with the conventional concentration (10mg/ml) in 50% MeCN/0.1% aq. TFA. For example, the limit of detection of human ACTH 18-39 was 10amol/well for the optimal condition but 10fmol/well for the conventional condition. The optimal condition (0.1mg/ml CHCA in 20% MeCN/0.1% aq. TFA) was verified with five model peptides and provided significant improvement in sensitivity (by two to three orders of magnitude) compared with the conventional conditions. Optimizing the CHCA concentration and solvent composition significantly improved the detection sensitivity in the analysis of peptides by MALDI-MS. Copyright (c) 2013 John Wiley & Sons, Ltd.
引用
收藏
页码:1217 / 1223
页数:7
相关论文
共 50 条
  • [31] Structure-performance relationships of phenyl cinnamic acid derivatives as MALDI-MS matrices for sulfatide detection
    Suparna Tambe
    Henning Blott
    Annabelle Fülöp
    Nils Spang
    Dirk Flottmann
    Stefan Bräse
    Carsten Hopf
    Hans-Dieter Junker
    Analytical and Bioanalytical Chemistry, 2017, 409 : 1569 - 1580
  • [32] Structure-performance relationships of phenyl cinnamic acid derivatives as MALDI-MS matrices for sulfatide detection
    Tambe, Suparna
    Blott, Henning
    Fueloep, Annabelle
    Spang, Nils
    Flottmann, Dirk
    Braese, Stefan
    Hopf, Carsten
    Junker, Hans-Dieter
    ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2017, 409 (06) : 1569 - 1580
  • [33] Fractional Factorial Design of MALDI-TOF-MS Sample Preparations for the Optimized Detection of Phospholipids and Acylglycerols
    AlMasoud, Najla
    Correa, Elon
    Trivedi, Drupad K.
    Goodacre, Royston
    ANALYTICAL CHEMISTRY, 2016, 88 (12) : 6301 - 6308
  • [34] Rapid detection of haloarchaeal carotenoids via liquid-liquid microextraction enabled direct TLC MALDI-MS
    Manikandan, Muthu
    Hasan, Nazim
    Wu, Hui-Fen
    TALANTA, 2013, 107 : 167 - 175
  • [35] MALDI-MS and HILIC ESI-MS/MS as Versatile Tools for Detection of Monoethanolamine Degradation Products in a Real Postcombustion Carbon Dioxide Capture Plant
    Cotugno, P.
    Monopoli, A.
    Nacci, A.
    Zambonin, C. G.
    Calvano, C. D.
    ENERGY & FUELS, 2014, 28 (02) : 1295 - 1303
  • [36] Compelling Advantages of Negative Ion Mode Detection in High-Mass MALDI-MS for Homomeric Protein Complexes
    Maedler, Stefanie
    Barylyuk, Konstantin
    Erba, Elisabetta Boeri
    Nieckarz, Robert J.
    Zenobi, Renato
    JOURNAL OF THE AMERICAN SOCIETY FOR MASS SPECTROMETRY, 2012, 23 (02) : 213 - 224
  • [37] Graphite Supported Preparation (GSP) of α-Cyano-4-Hydroxycinnamic Acid (CHCA) for Matrix-Assisted Laser Desorption/Ionization Mass Spectrometry (MALDI-MS) for Peptides and Proteins
    Gorka, Jan
    Bahr, Ute
    Karas, Michael
    JOURNAL OF THE AMERICAN SOCIETY FOR MASS SPECTROMETRY, 2012, 23 (11) : 1949 - 1954
  • [38] 2-Phenyl-3-(p-aminophenyl) Acrylonitrile: A Reactive Matrix for Sensitive and Selective Analysis of Glycans by MALDI-MS
    Ling, Ling
    Xiao, Chunsheng
    Ma, Yao
    Jiang, Liyan
    Wang, Sheng
    Guo, Liming
    Jiang, Shimei
    Guo, Xinhua
    ANALYTICAL CHEMISTRY, 2019, 91 (14) : 8801 - 8807
  • [39] Detection of Neutral Species in the MALDI Plume Using Femtosecond Laser Ionization: Quantitative Analysis of MALDI-MS Signals Based on a Semiequilibrium Proton Transfer Model
    Shirota, Tatsuro
    Tsuge, Masashi
    Hikosaka, Yasumasa
    Soejima, Koichi
    Hoshina, Kennosuke
    JOURNAL OF PHYSICAL CHEMISTRY A, 2017, 121 (01) : 31 - 39
  • [40] Re-exploring α-Cyano-4-Hydroxycinnamic Acid as a Reactive Matrix for Selective Detection of Glutathione via MALDI-MS
    Guo, Liming
    Xiao, Chunsheng
    Li, Jiarui
    Lin, Xi
    Zhao, Nan
    Wang, Xinyu
    Dong, Lejuan
    Guo, Xinhua
    JOURNAL OF THE AMERICAN SOCIETY FOR MASS SPECTROMETRY, 2021, 32 (12) : 2837 - 2841