Measurement and calculation of heterogeneity correction factors for an Ir-192 high dose-rate brachytherapy source behind tungsten alloy and steel shields

被引:27
作者
Kirov, AS
Williamson, JF
Meigooni, AS
Zhu, Y
机构
[1] Radiation Oncology Center, Mallinckrodt Institute of Radiology, Washington Univ. School of Medicine, St. Louis
关键词
heterogeneity corrections; Ir-192; HDR; TLD; Monte Carlo; detector artifacts;
D O I
10.1118/1.597733
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Shields made of high atomic number material are commonly used in vaginal applicators with high dose-rate (HDR) (192)IR remotely afterloaded brachytherapy sources. However little data is available for the dose distribution around such shields. Heterogeneity correction factors (HCFs) are defined as the ratio of the dose to a point with the heterogeneity (shield) in place, divided by the dose to the same point with no heterogeneity. Using thermoluminescent dosimeters (TLDs) in solid water phantom we have measured the HCFs behind 6 and 20 mm diam tungsten alloy disks, 4 and 2 mm thick and a 4 mm thick steel disk, positioned 15 mm from the source. For each measurement point, the heterogeneity correction factors were also inferred from Monte Carlo simulations, which accurately modeled the experimental geometry. The agreement between measured and calculated HCFs on the average was within 6%. Tungsten alloy disks resulted in about two times greater dose reduction in water (HCF approximate to 0.4, for 20X4 mm disk) than for a steel disk with the same dimensions (HCF approximate to 0.85). Reducing the disk diameter to 6 mm increased the dose transmission up to about 25%. Increasing the source-to-detector distance from 4 to 7 cm caused a change in HCF from 2% to more than 20%, depending on disk material and diameter. The detector artifact effects arising from the finite size and different composition of the TLD chips were determined. (C) 1996 American Association of Physicists in Medicine.
引用
收藏
页码:911 / 919
页数:9
相关论文
共 29 条
  • [1] AUTORADIOGRAPHY FOR I-125 SEEDS
    ALBERTI, W
    DIVOUX, S
    POTHMANN, B
    TABOR, P
    HERMANN, KP
    HARDER, D
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1993, 25 (05): : 881 - 884
  • [2] [Anonymous], ORNLRSIC46 RAD SHIEL
  • [3] Cameron JR., 1968, THERMOLUMINESCENT DO
  • [4] CERVICAL RADIUM APPLICATORS WITH SCREENING IN THE DIRECTION OF BLADDER AND RECTUM
    FLETCHER, GH
    [J]. RADIOLOGY, 1953, 60 (01) : 77 - 84
  • [5] GLASGOW G P, 1979, Medical Physics (Woodbury), V6, P49, DOI 10.1118/1.594551
  • [6] CALIBRATION OF IR-192 HIGH-DOSE-RATE AFTERLOADING SYSTEMS
    GOETSCH, SJ
    ATTIX, FH
    PEARSON, DW
    THOMADSEN, BR
    [J]. MEDICAL PHYSICS, 1991, 18 (03) : 462 - 467
  • [7] TLD, diode and monte carlo dosimetry of an Ir-192 source for high dose-rate brachytherapy
    Kirov, AS
    Williamson, JF
    Meigooni, AS
    Zhu, Y
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1995, 40 (12) : 2015 - 2036
  • [8] QUANTITATIVE VERIFICATION OF IR-192 PDR AND HDR SOURCE STRUCTURE BY PINHOLE AUTORADIOGRAPHY
    KIROV, AS
    MEIGOONI, AS
    ZHU, Y
    VALICENTI, RK
    WILLIAMSON, JF
    [J]. MEDICAL PHYSICS, 1995, 22 (11) : 1753 - 1757
  • [9] MONTE-CARLO CALCULATION OF KERMA TO A POINT IN THE VICINITY OF MEDIA INTERFACES
    LI, ZF
    WILLIAMSON, JF
    PERERA, H
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1993, 38 (12) : 1825 - 1840
  • [10] VOLUME-BASED GEOMETRIC MODELING FOR RADIATION TRANSPORT CALCULATIONS
    LI, ZF
    WILLIAMSON, JF
    [J]. MEDICAL PHYSICS, 1992, 19 (03) : 667 - 677