Study on the response of indigenously developed CaSO4:Dy phosphor-based neutron dosemeter

被引:4
作者
Bakshi, A. K. [1 ]
Pradhan, A. S. [1 ]
Kher, R. K. [1 ]
Srivastava, K. [1 ]
Varadharajan, G. [1 ]
Chatterjee, S. [1 ]
Sathian, V. [2 ]
机构
[1] Bhabha Atom Res Ctr, Radiol Phys & Advisory Div, Mumbai 400085, Maharashtra, India
[2] Bhabha Atom Res Ctr, Radiat Safety Syst Div, Mumbai 400085, Maharashtra, India
关键词
MIXED FIELD DOSIMETRY; DY;
D O I
10.1093/rpd/ncp019
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
In the present paper, we report the indigenous development of a neutron-sensitive thermoluminescent dosemeter (TLD) based on CaSO4:Dy Teflon TL disc. The study includes indigenous development of a neutron dosemeter and its response in terms of operational quantity to different energies of neutrons under various irradiation conditions. It was found that the thermal neutron sensitivity of the CaSO4:Dy Teflon neutron disc is about one-third of TLD-600. However, the thermal neutron sensitivity with respect to CaSO4:Dy Teflon gamma disc is about 42 times for in-air irradiation and about 84 times for on-phantom irradiation condition. This newly developed neutron dosemeter can be used as a routine TLD with a slight change in the design of the TLD badge system in the mixed fields of gamma and neutrons of energy up to 500 keV for radiation workers engaged in nuclear fuel cycle operation and other applications.
引用
收藏
页码:73 / 80
页数:8
相关论文
共 17 条
[1]   MIXED FIELD DOSIMETRY WITH CASO4 - DY [J].
AYYANGAR, K ;
CHANDRA, B ;
LAKSHMANAN, AR .
PHYSICS IN MEDICINE AND BIOLOGY, 1974, 19 (05) :656-664
[2]   Development of an algorithm for TLD badge system for dosimetry in the field of X and gamma radiation in terms of Hp(10) [J].
Bakshi, A. K. ;
Srivastava, K. ;
Varadharajan, G. ;
Pradhan, A. S. ;
Kher, R. K. .
RADIATION PROTECTION DOSIMETRY, 2007, 123 (02) :148-155
[3]  
BARTLETT DT, 2005, 385 HLTH SAF EX
[4]  
BEACH JL, 1976, HEALTH PHYS, V31, P452
[5]  
FEHRENBACHER G, 2004, P INT RAD PROT ASS I
[6]  
HARVEY JR, 1973, P IAEA C NEUTR MON R, P199
[7]   THERMOLUMINESCENT RESPONSE OF MIXTURE OF CASO4(MN) AND LI2SO4 TO THERMAL NEUTRON AND GAMMA-RAY FIELDS [J].
IKEYA, M ;
ISHIBASHI, M ;
ITOH, N .
HEALTH PHYSICS, 1971, 21 (03) :429-+
[8]  
*INT EL COMM, 1991, 45B INT EL COMM TECH
[9]  
*INT ORG STAND, 1998, 85293 ISO
[10]  
International Organization for Standardization, 2001, 85291 ISO